{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78219",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide within U.S.A and Puerto Rico; International: Australia, Brazil, Canada, China, Germany, India, Japan, Netherlands & Singapore",
      "recall_number": "Z-0547-2018",
      "product_description": "M/DN Intramedullary Fixation Humeral Guide Wire - Smooth 2.4 mm Diameter 70 cm Length Single Use Only, Item Number 00-2255-025-00    Product Usage:  The Ball Tip and Tear Drop Guide Wires are used during the initial reaming of the intramedullary canal. They also assists in guiding the nail during implantation. The guide wire incorporates either a ball or tapered tip to aid in removal of the intramedullary reamer if it becomes lodged or fails.",
      "product_quantity": "6,290 units",
      "reason_for_recall": "The design verification for the previous packaging configuration G928 does not cover the 70cm wires. A design verification has been completed to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed.",
      "recall_initiation_date": "20170605",
      "center_classification_date": "20180206",
      "termination_date": "20190417",
      "report_date": "20180214",
      "code_info": "All Lot Numbers   Expiry Date Before April 2022"
    }
  ]
}