  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72684",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Australia, Austria, Bangladesh, Belgium, Brazil, Cambodia, Canada, China, Egypt, France, Germany, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Lebanon, Libya, Lithuania, Malaysia, Martinique, Myanmar, Mexico, Mozambique, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Vietnam.",
      "recall_number": "Z-0547-2016",
      "product_description": "Ingenuity Core Computed Tomography X-ray system",
      "product_quantity": "135 Units",
      "reason_for_recall": "Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.",
      "recall_initiation_date": "20151029",
      "center_classification_date": "20151224",
      "termination_date": "20180626",
      "report_date": "20151230",
      "code_info": "Model No. 728321; S/N: 52007, 52008, 52010, 52013, 52014, 52015, 52018, 52019, 52020, 52021, 52024, 52027, 52032, 52033, 52034, 52035, 52036, 52037, 52038, 52040, 52041, 52042, 52045, 52046, 52047, 52048, 52051, 52053, 52055, 52057, 52058, 52060, 52064, 52065, 52067, 52068, 52069, 52070, 52075, 52078, 310017, 310026, 310039, 310042, 310052, 310055, 310059, 310064, 310069, 310071, 310081, 310088, 310098, 310111, 310113, 310119, 310121, 310133, 310143, 310144, 310149, 310151, 310157, 310159, 310160, 310204, 310205, 310206, 310208, 310209, 310210, 310211, 310212, 310213, 310215, 310216, 310217, 310219, 310220, 310223, 310224, 310225, 310226, 310228, 310231, 333002, 333009, 333014, 333015, 333030, 333031, 333037, 333042, 333043, 333044, 333046, 333048, 333049, 333050, 333052, 333053, 333054, 333055, 333056, 333058, 333059, 333060, 333061, 333062, 333063, 333065, 333066, 333067, 333069, 333070, 333072, 333073, 333074, 333075, 333076, 333077, 333078, 333079, 333080, 333081, 333082, 333086, 333087, 333090, 333091, 333093, 333101, 333110 & 333112."
    }
  ]
}