  "results": [
    {
      "status": "Terminated",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75491",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0541-2017",
      "product_description": "Right Heart Cath Lab Kit w/2 Port \"OFF\" Manifold (600 psi), 72\" Spike Set, Transpac¿ IV and 48\" HP Tubing, Item No. 46100-87  The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.",
      "product_quantity": "40 units",
      "reason_for_recall": "ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.",
      "recall_initiation_date": "20160927",
      "center_classification_date": "20161115",
      "termination_date": "20170711",
      "report_date": "20161123",
      "code_info": "3251051"
    }
  ]
}