{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "345 E Main St",
      "reason_for_recall": "All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening",
      "address_2": "",
      "product_quantity": "591",
      "code_info": "Part Number 42-5320-071-01 TIBIA SIZE E, LEFT,   Lot 11005332, 11005333, 11005334, 11005359, 11005360, 11005361, 11006133, 11006134, 11006135, 11006136, 11006137, 11006138, 11006712, 11007124, 61956760, 61956761, 61956762, 61978561, 61978562, 61978563, 62022179, 62030574, 62033389, 62079745, 62030575, 11007596, 62091364, 62103722, 62060744, 62068418, 62079746, 62103723, 62110198, 62117658, 62076580, 62128899, 11007595, 11007597, 62144992, 62150529, 62150530 ,   Part Number 42-5320-071-02 TIBIA SIZE E RIGHT, Lot 11006515, 11007238, 11007574, 11007575, 11007576, 62018345, 62018346, 62022180, 62022181, 62033391, 62051023, 62051801, 62068419, 62076582, 62079747, 62079761, 62103725, 62110199, 62117659, 62128904, 11007641, 11007640, 11007642,",
      "center_classification_date": "20121218",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.",
      "state": "IN",
      "product_description": "Persona The Personalized Knee System, Natural Tibia,   Cemented, 5 Degree Stemmed Tivanium Sterile  TIBIA SIZE E, LEFT  TIBIA SIZE E RIGHT    This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.",
      "report_date": "20121226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-0541-2013",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "63690",
      "termination_date": "20140416",
      "more_code_info": "",
      "recall_initiation_date": "20121114",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}