  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Parsippany",
      "address_1": "100 Interpace Pkwy",
      "reason_for_recall": "Broken screw insertor hardness was measured and found to be below the acceptable limits.",
      "address_2": "",
      "product_quantity": "15 units",
      "code_info": "Lot Numbers 540199, 540199-1",
      "center_classification_date": "20121217",
      "distribution_pattern": "Nationwide Distribution, including the states of MO, TX, and IN.",
      "state": "NJ",
      "product_description": "Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178",
      "report_date": "20121226",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ebi, Llc",
      "recall_number": "Z-0536-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63653",
      "termination_date": "20151207",
      "more_code_info": "",
      "recall_initiation_date": "20120929",
      "postal_code": "07054-1149",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}