  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86650",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "FAX",
      "distribution_pattern": "US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.",
      "recall_number": "Z-0532-2021",
      "product_description": "CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, Model numbers: Infinix-8000C and Infinix-8000H - Product Usage: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.",
      "product_quantity": "14 units",
      "reason_for_recall": "There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in  a delay with procedure  and/or  potential damage to the patients blood vessels due to removal without visibility.",
      "recall_initiation_date": "20201008",
      "center_classification_date": "20201203",
      "termination_date": "20221206",
      "report_date": "20201209",
      "code_info": "Affected Serial Numbers:    1 AFA1662001,   2 AFB1872005,  3 AFB1832003,   4 AFB1872004,   5 AFB1892006,   6 BFA18Y7001,   7 AEA17X2001,   8 AFA1692002,   9 AEB1812002,   10 BFA1932001,   11 BFB1952003,   12 BFB1942002,   13 BFB1972004,   14 BFB1972005,   15 AFB1892007,   16 BEC2032001,   17 BFC2062007,   18 BFC2062008,   19 BHA1992001,   20 BFC2012006",
      "more_code_info": ""
    }
  ]
}