{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.",
      "address_2": "",
      "product_quantity": "178886",
      "code_info": "Lots 089610  241030  089600  241020  238180  416260  507280  450740  314580  736790  779690  889140  836030  970200  444550  984630  316530  198990  419500  431591  444655  459030  633280  921380  148200  441530  718410  884690  856150  444530  066610  352090  553250  HA109-05  148190  666010  825420  989490R  989490  444510  444510R  984610  241410  390410  152390  316520  390410R  432706  484710  484720  238170  427280  450730  323210  507270  748450  789870  877610  002130  444500  984600  363080  670430  136700  198960  198970  230290  230300  413443  413444  432705  984600R  148180  441520  456560  732840  825410  889130  970190  444420  984580  921390  530950  120230  198910  198930  230280  413441  803810  803810R  086800  357830  740750  921550  450720  323200  640280  718400  856140  946100  444470  057550  386080  303010  633210  403720  516120  418870  824120  592130  148930  017430  802520  274630  966250  966260  047790  351330  822190  846490  131420  131430  912600  965370  277220  314570  450710  771080  856130R  856130  988850  251590  444450  621070  147070  530940  120220  198850  341593  419499  422450  450710R  633250  831320  089680  340980  241400  118720  441510  516910  771070  836020  835980  444140  066600  120210  198820  436780  697210  722780  238160  375410  375410R  344370  441500  456550  825400  831310  884680  924720  946090  989480  242060  444120  984570  524230  524230R  559330  628180  017420  110510  139030  198790  230260  230270  349160  428342  428608  432704  553220  553230  553240  671740  671750  803800  237530  541190  427240  740740  064900  003150  314520  148170  456540  338850  507260  547070  659870  630650  775690  889120  779680  804350  856120  946080  970180  002120  004740  251580  444100  007640  333720  449860  102550  194590  194600  237250  418830  774130  110710  273080  638390  141600  950170  950130  484730  484680  524220  621770  785560  210030  714850  717010  966230  717130  152420  002370  949680  949690  068340  120400  131400  131410  230310  230320  2375300  341591  349159  418581  432702  439398  619620  799540  921370  237520  427220  826470  118710  328020  405050  450700  483680  547060  640270  713900  856110  789860  889110  889100  989470  004730  242050  251570  444090  984560  621060  843670  956170  956030  956130  877470  723170  152380  921380  139020  230200  341590  349158  435195  619610  637560  671570  768780  799530  912580  140590  425870  826460  198330  323190  450690  465770  483660  483670  604030  836010  889090  877600  779670  946070  979720  033500  444070  444070R  984550  374440  867190  516050  921370  126090  732750  029130  771190  360100  392660  598350  152370  002340  966220  970150  821360  949660  949670  139010  198740  230190  410785  419769  440610  637540  671720  089670  241390  671460  740730  064890  921440  921540  198320  328010  284830  298720  231090  428460  395330  421200  441490  456530  544180  547050  465750  465760  500530  513090  604020  500540  601650  607620  624020  649280  630640  684560R  684560  718390  856100  889080  979710  930710  002110  320780  251560  444060  984540  057530  092160  241380  590720  241370  492090  403710  670420  732710  536860  580250  592100  592090  459910  154350  017410  877450  360080  717000  966210  316480  092160R  341588  386580  388258  418573  435193  435194  634920  634930  634970  634990  635030  722770  817050  100430  541100  712020  148160  003020  416250  314500  338840  284820  375400  441480  465740  513080  596390  684550  630630  718380  659860  771060  979700  856090  002100  946060  070080  251550  444010  444050  984530  057520  102540  473360  516040  921360  931580  459900  950060  524210  459500  134030  766440  559120  672750  970140  316460  821350  949650  103150  139000  198690  230140  230150  321830  350841  352080  414693  440550  513080R  671700  737280  799520  866015  921430  921420  064880  231050  198300  630610  700100  624010  649270  666000  630620  718370  736780  771040  771030  804340R  804340  946050  979690  771050  048480  251520  444020  444040  645360  621050  269140  066590  899060  066580  241350  241360  990740  147330  374430  424630  424640  399880  876280  772560  367210  516030  616950  735990  732700  732690  732680  735980  148470  931570  141580  459830  141540  592050  592070  459820  764570  340000  459460  770890  770890R  134020  017240  017220  181620  017230  225590  283370  722870  037930  714840  152360  274610  716980  966200  002320  716970  316440  716990  966180  970110  065580  316450  949630  949640  057520  403449  435191  458990  493800  493810  493820  493830  493850  493860  553200  553210  322090  540880  230980  314490  416240  483640R  450680  465730  483650  483640  500520  547040  736770  889070  946040R  946040  070070  989450  251500  444000  984520  066570  340700  590200  968750  213450  256630  765760  824090  273070  931560  484600  524200  360070  628160  530930  127040  127060  198640  198670  341585  388257  403451  553170  619550  619560  671640  671650  765760R  767020  855710  089660  241340  553160  829260  089650  241330  418150  241320  120200  295870  241310  671630  829250  089640  401020  241300  553150  212700  829240  070060  340690  066560  074170  391466  566260  416230  416220  465720  554000  500510  529010  529020  592840  592820  610810  610800  640250  640260  380990  251490  443980  984510  057500  241280  333580  241290  633190  670410  566280  566270  276250  276270  276230  276210  732660  141530  591950  459810  591930  484580  950050  804000  484560  229800  621950  621960  016850  133940  134010  134000  558870  866260  186130  186140  186120  152350  233930  427190  751860  019760  019750  966170  966160  002300  970000  970010  019740  990700  425090  664410  118700  338830  465710  523590  624000  684540  831300  889060  251470  443950  621040  984500  241260  131380  230110  364827  418854  435189  458970  553140  601620  684540R  825470  086780  640240R  540830  851890  118690  305580  483630  314480  421190  441470  416210  450670  456520  516900  553990  601640  544160  630600  640240  659850  677530  718360  825390  825380  884670  877580  930700  924710  924700  979680  989440  004710  443940  443940R  251480  621030  057490  396180  649850  555250  876180  516010  616940  276200  950040  016820  133990  256190  558820  558850  628130  714830  152340  186100  186110  966110  966150  966080  966090  966140  002280  966130  969970  969950  522860  131360  131370  230090  230100  336811  403448  419768  698900  794200  912570  540810  425230  921410  328000  456510  640230  684530  924690  970170  070050  242040  443930  251440  443930R  621020  066550  241250  445640  340650  340650R  399860  276170  276190  770840  923930  628110  152330  316400  131340  131350  230070  230080  357718  372684  419672  436750  527900  527930  553110  553130  737270  779270  794190  921410R  164340  322080  740720  425110  540780  851880  002810  284810  198290  323180  421180  344350  405040  327990  456500  553970  553980  500490  483620  500500  547020  640220  592810  665990  601630  596380  601630R  607610  700090  736760  732830  930690  979670  002090  320770  443880  443890  443900  898440  984490  066540  092150  066530  241220  340640  241240  445630  363040  363060  403700  566250  599370  363020  450900  399770  424610  478730  503850  374410  740340  793330  876210  876220  367200  367190  516000  616920  213440  276110  276180  732650  732640  732630  824070  824060  612680  459790  459800  591830  591780  591860  591800  591810  949950  484540  484520  484530  804040  484510  785530  973730  621760  099330  770740  877440  133980  133950  133960  133970  628090  558800  941270  310500  588250  186080  186080R  152320  252460  186090  233920  571910  571920  202720  199210  002140  716940  716930  717110  717120  002210  716950  716960  316350  002260  002230  316330  316390  316340  969940  202740  316300  522850  949610  949620  129880  198560  375846  435186  443900R  527850  527860  553090  628090R  700090R  766950  HA063-04  425100  921400  064870  198280  314470  284740  338820  450660  465700  483610  500480  513070  601620  592800  700080  630590  718350  684520  804330  659840  884660  924680  946030R  946030  979650  979660  004700  242030  380970  251430  443870  380980  903320  590710  896870  621010  092140  066510  066490  241200  241210  990690  633180  233990  424600  374400  399850  180690  843580  515960  798930  921350  968760  070840  089790  611580  765740  765750  129930  148460  989920  931540  733210  141520  949920  484500  804010  484490  524190  621940  133930  133910  597140  855270  948470  861810  364660  558780  866240  162650  530920  817170  002200  002150  966060  966050  250690  522840  821340  853150  853210  949590  949600  969840  230000  432695  435183  524190R  527830  698880  817010  861810R  086790  826450  118680R  118680  450650  665980  771020  889050R  835970  889050  004690  250400  443860  251420  984480  241190  966040  352070  357715  388259  410780  414689  432693  436740  527780  089630  241180  089620  241170  298710  825370  340970  390360  966020  553080  999570  241160  877570  118030  241150  177140  125040  664400  118670  921390  002800  284730  456490  483590  500470  547010  677520  771010  877560  889040  251410  443840  984470  241140  068320  074200  357717  422227  435181  435182  527730  527740  671580  794180  826440  002790  064780  064840  198270  298700  338810  344340  421170  456480  395320  428450  516890  483570  483580  500460  544130  544140  586790  553950  553960  586780  596350  604010  601610  623990  665970  684510  718340  771000  816120  924660  884650  924670  877550  930680  979640  004680  946010  320750  443810R  320760  443810  443830  443790  984450  007630  057480  353090  459830  649840  450880  374390  473340  493990  384720  924340  515890  451250  692180  692180R  276080  276100  276070  612670  931530  141490  141510  141500  591680  591690  591740  779490  779500  779480  459430  016540  227400  723150  907320  254000  628070  162630  162640  714820  152310  299060  186060  427180  694920  751850  274590  002120  717070  717080  717090  717100  002040  002070  002040R  002070R  152450  716920  065530  671270  671260  065510  969830  522810  129830  129840  229970  372010  403444  409810  409820  410778  410779  414688  428616  443830R  521810  521820  521830  521840  540760  553050  582610  601610R  671510  671530  671540  671560  782108  794170  834370  984450R  HA052-03  212690  212680  740710  424520  664390  851870  921380  064750  298690  230970  284800  327980  428440  344330  405030  465690  456470  483550  507250  500450  553940  544110  586770  607600  596330  630580  665950  684500  700070  718330  732820  884640  946000  979630  004670  261750  320740  443730  443750  251380  443770  984430  621000  092130  066260  348710  066290  241100  241110  445610  353080  689630  459750  403640  384590  403650  403660  384580  403670  374370  839580  503790  503780  374360  493640  566240  450870  876240  409590  515830  616850  671250  798910  750090  921340  276060  276050  276040  732600  795650  774120  148410  273060  588290  931520  141470  141420  459740  141410  141480  459730  459770  589690  589700  459760  459780  589680  949850  050790  141390  949860  484430  484470  484450  484440  484460  778910  778920  484410  484480  778900  133890  181600  016230  019570  736140  225130  365790  627760  627820  123440  144140  162620  123400  752490  674960  714810  152300  252450  392950  392960  597010  694910  751840  873310  M431520  431520  202670  202690  202680  001880  001990  001930  001970  001900  716880  001920  202700  716890  716900  002020  152460  716910  671170  065490  047770  047780  294890  351320  522790  671090  821330  876230  928980  949570  949580  229940  388255  414687  432692  433033  436730  440520  484440R  521760  621000R  671490  921350  829230  089590  241090  424500  740700  540760  002780  198240  338800  441460  630570  649260  829220  970150  836000  877540  002080  889030  930670  242020  251360  443710  443710R  065150  530910  966010  124990  242020R  272330  349154  359548  388254  422225  432691  436710  436720  521710  698870  740700R  799510  921340  851860  740690  456460  825300  825350  889020  831290  984420  241080  390390  327250  068310  074160  124970  385586  409780  511600  730230  HA101-05  HA109-05  140580  540740  851850  740680  375390  284790  314460  416200  456450  825260  884630  004660  930660  979620  443700  984410  027512  116870  124960  127650  127660  129100  129110  129120  129130  229920  385592  403443  419767  432690  440500  511560  511580  511590  521660  541470  063470  063640  064480  063710  063790  327970  284780  298670  298680  344320  421150  441450  523550  421160  456440  450640  465680  507240  500440  523560  523570  500430  553930  564580  596280  607590  630560  665940  677510  700060  732810  779660  856080  884620  930650  970140  979600  979610  004640  033520  070030  930640  070040  070020  204370  251350  443620  443650  898430  984390  057470  057450  092120  241060  241040  241070  492070  407230  358530  358540  384560  384570  358560  385150  403590  566220  503730  403600  566230  399740  399730  399780  493620  473160  555240  924330  515820  616830  692170  276000  276010  276030  276020  732550  758800  758820  758780  758810  823750  823700  823740  050630  050680  141250  453470  459710  141270  141260  459720  453370  141280  459700  589670  948660  589620  589620R  589630  948580  948650  484390  484400  484370  484360  778890  778880  778770  099270  229750  330280  016170  133870  016150  165130  016210  256030  133860  016200  019540  097760  097740  097710  097750  097720  097730  097700  530900  714800  714790  817160  674950  752480  152290  392940  392930  694900  737410  737400  737430  737420  019720  019700  019690  019710  431510  202650  716870  717020  716850  001780  001820  001850  316140  717030  316210  316190  316180  716860  717060  001800  001830  001870  717050  047750  047760  316150  316170  316220  316230  522700  949550  949560  033569  048028  122670  129080  198280  341576  341578  349153  403441  403442  419669  425735  435175  435176  440470  458910  504000  511500  566230R  671470  737250  978468  340620  441440  298660  465670  825110  825160  930630  831220  141510  443600  064370  519060  257710  503960  651070  775700  639190  298750  434200  450810  465810  547100  856290  979900  550460  550200  019590  175930  204880  444661  826920  322110  118760  344380  450800  465800  547090  732870  856280  979890  444720  550450  550180  175900  352130  357720  459150  649290  633370  089690  202790  164360  065110  619020  284770  323230  416270  441560  465790  601680  713910  732860  856270  879370  979880  879350  879360  444710  550170  591610  288130  413445  413446  414698  424890  444665  826900  866016  921560  118750  441550  456580  732850  856260  929510  444700  241430  603610  142270  309770  336105  386590  419677  422510  440491  633330  633350  826890  322100  619010  118740  284750  328030  450790  456570  554010  718420  856250  929500  979860  979870  444690  550440  550150  136740  142220  175880  341596  390405  414696  422500  633320  767080  826880  865225  618970  005020  231180  428470  450780  465780  507300  547080  775710  666020  856240  979850  550780  444680  550140  970190  175810  175840  204870  336104  348055  357719  385566  436800  439399  444658  459120  553290  633300  803830  003160  148230  298740  450770  601670  856230  879320  879330  879340  444660  550130  550770  136720  142210  364828  377416  381912  422490  431656  459080  803820  826850  739760  618930  618940  323220  450760  856220  961010  444640  550120  970180  149270  149280  212710  385630  553280  826840  483690  411800  913830  879180  212710  428120  005010  298730  450750  601660  704560  856210  979840  979830  550760  444630  550430  136710  175710  175720  348054  385628  422480  432707  553270  826820  440550  428100  826480  231170  856200  979820  550420  550100  165140  452130  520980  142200  210556  257730  385625  444668  698850  826810  238210  005000  148220  375420  856190  960980  979810  979800  444610  451080  550090  550410  550410R  038800  175690  175700  375848  385561  431592  436790  441460  553260  741170  826800  238200  118730  441540  507290  856180  856180R  979790  550750  550080  444600  352110  385624  431663  722790  795652  826520  618910  338860  856170  808390  550740  550070  444580  202760  142190  331526  385558  414695  419676  428618  435197  444657  459060  826430  877620  550400  550060  065060  148210  856160  550050  444570  142170  175630  431654  826420  784390  658260  911110  044380  487720  487740  139190  366770  482470  373650  622290  622460  070360  121516  152831  287951  345809  A66AA7  AE9AB7000  B1AAA7  B5KAB7  BY3AA7  C1VAA7  C3KAA7  C57AA7  C5XAA7  C74AE7  CA9AA7  CH7AF7  D5LAB7  D96AB7  DE4AA7  DJ4AA7  DX6AA7  E11AH7  E27AH7  E5VAB7  FA4AB7  FD4AA7  Z28AA7000      UDI:    (01)00888480055181(17)180801(10)416260  (01)00888480055181(17)190201(10)736790  (01)00888480055181(17)190401(10)889140  (01)00888480055181(17)280523(10)316530  (01)00887868104176(17)180701(10)441530  (01)00887868104176(17)190501(10)884690  (01)00887868104176(17)200301(10)444530  (01)00887868104152(17)181201(10)666010  (01)00887868104152(17)190801(10)989490R  (01)00887868104152(17)200301(10)444510  (01)00887868104152(17)200401(10)444510R  (01)00887868104152(17)201101(10)984610  (01)00887868104152(17)260602(10)152390  (01)00887868104152(17)280505(10)316520  (01)00887868104169(17)171201(10)238170  (01)00887868104169(17)180801(10)450730  (01)00887868104169(17)190101(10)748450  (01)00887868104169(17)190401(10)789870  (01)00887868104169(17)201201(10)984600  (01)00887868104145(17)180801(10)441520  (01)00887868104145(17)190401(10)889130  (01)00887868104145(17)200301(10)444420  (01)00887868104145(17)210226(10)984580  (01)00887868104145(17)230401(10)921390  (01)00887868104145(17)260501(10)530950  (01)00887868104114(17)180801(10)450720  (01)00887868104114(17)200301(10)444470  (01)00887868104114(17)230226(10)516120  (01)00887868104114(17)230621(10)418870  (01)00887868104114(17)231214(10)824120  (01)00887868104114(17)240519(10)592130  (01)00887868104114(17)240910(10)148930  (01)00887868104114(17)250324(10)017430  (01)00887868104114(17)260523(10)802520  (01)00887868104114(17)260901(10)274630  (01)00887868104114(17)280321(10)966250  (01)00887868104114(17)280326(10)966260  (01)00887868104138(17)180728(10)314570  (01)00887868104138(17)190515(10)856130R  (01)00887868104138(17)190601(10)988850  (01)00887868104138(17)250930(10)147070  (01)00887868104138(17)260419(10)530940  (01)00887868104107(17)190201(10)771070  (01)00887868104091(17)171201(10)238160  (01)00887868104091(17)180701(10)375410  (01)00887868104091(17)180801(10)375410R  (01)00887868104091(17)190501(10)884680  (01)00887868104091(17)210101(10)984570  (01)00887868104091(17)241119(10)524230  (01)00887868104091(17)241119(10)524230R  (01)00887868104091(17)251207(10)559330  (01)00887868104091(17)251216(10)628180  (01)00887868104084(17)180201(10)541190  (01)00887868104084(17)180301(10)740740  (01)00887868104084(17)180601(10)064900  (01)00887868104084(17)180701(10)314520  (01)00887868104084(17)180801(10)456540  (01)00887868104084(17)190301(10)775690  (01)00887868104084(17)190401(10)889120  (01)00887868104084(17)200201(10)251580  (01)00887868104084(17)200201(10)444100  (01)00887868104084(17)210601(10)333720  (01)00887868104084(17)230514(10)194600  (01)00887868104084(17)230515(10)237250  (01)00887868104084(17)230619(10)418830  (01)00887868104084(17)230723(10)774130  (01)00887868104084(17)231120(10)110710  (01)00887868104084(17)240211(10)273080  (01)00887868104084(17)240211(10)638390  (01)00887868104084(17)240408(10)141600  (01)00887868104084(17)240623(10)950170  (01)00887868104084(17)240624(10)950130  (01)00887868104084(17)240817(10)484730  (01)00887868104084(17)240825(10)484680  (01)00887868104084(17)241015(10)524220  (01)00887868104084(17)241020(10)621770  (01)00887868104084(17)241112(10)785560  (01)00887868104084(17)260119(10)210030  (01)00887868104084(17)260502(10)714850  (01)00887868104084(17)280308(10)717010  (01)00887868104084(17)280309(10)966230  (01)00887868104084(17)280313(10)717130  (01)00887868104084(17)280325(10)152420  (01)00887868104084(17)280327(10)002370  (01)00887868104053(17)180101(10)237520  (01)00887868104053(17)190731(10)989470  (01)00887868104053(17)200201(10)251570  (01)00887868104053(17)230404(10)956170  (01)00887868104053(17)230405(10)956030  (01)00887868104053(17)230405(10)956130  (01)00887868104053(17)250218(10)877470  (01)00887868104053(17)250807(10)723170  (01)00887868104053(17)260601(10)152380  (01)00887868104077(17)181101(10)604030  (01)00887868104077(17)190401(10)836010  (01)00887868104077(17)190401(10)889090  (01)00887868104077(17)190501(10)877600  (01)00887868104077(17)200301(10)444070  (01)00887868104077(17)200401(10)444070R  (01)00887868104077(17)230226(10)516050  (01)00887868104077(17)230404(10)921370  (01)00887868104077(17)230515(10)126090  (01)00887868104077(17)230710(10)732750  (01)00887868104077(17)240906(10)029130  (01)00887868104077(17)250212(10)771190  (01)00887868104077(17)250414(10)360100  (01)00887868104077(17)260424(10)392660  (01)00887868104077(17)260502(10)598350  (01)00887868104077(17)260623(10)152370  (01)00887868104077(17)280309(10)002340  (01)00887868104077(17)280309(10)966220  (01)00887868104077(17)290214(10)970150  (01)00887868104022(17)180301(10)740730  (01)00887868104022(17)180601(10)064890  (01)00887868104022(17)181001(10)544180  (01)00887868104022(17)181001(10)547050  (01)00887868104022(17)181101(10)604020  (01)00887868104022(17)190201(10)684560R  (01)00887868104022(17)190615(10)979710  (01)00887868104022(17)211001(10)241380  (01)00887868104022(17)211001(10)590720  (01)00887868104022(17)220901(10)403710  (01)00887868104022(17)230710(10)732710  (01)00887868104022(17)231120(10)536860  (01)00887868104022(17)240212(10)580250  (01)00887868104022(17)240515(10)592100  (01)00887868104022(17)240520(10)592090  (01)00887868104022(17)240602(10)459910  (01)00887868104022(17)240919(10)154350  (01)00887868104022(17)250319(10)017410  (01)00887868104022(17)250320(10)877450  (01)00887868104022(17)250410(10)360080  (01)00887868104022(17)280311(10)717000  (01)00887868104022(17)280329(10)966210  (01)00887868104046(17)171201(10)100430  (01)00887868104046(17)180501(10)148160  (01)00887868104046(17)180501(10)003020  (01)00887868104046(17)180701(10)416250  (01)00887868104046(17)190101(10)684550  (01)00887868104046(17)190501(10)979700  (01)00887868104046(17)190601(10)002100  (01)00887868104046(17)200331(10)444010  (01)00887868104046(17)201201(10)984530  (01)00887868104046(17)210501(10)057520  (01)00887868104046(17)230319(10)516040  (01)00887868104046(17)230329(10)921360  (01)00887868104046(17)240304(10)931580  (01)00887868104046(17)240508(10)459900  (01)00887868104046(17)240625(10)950060  (01)00887868104046(17)241022(10)524210  (01)00887868104046(17)250117(10)459500  (01)00887868104046(17)250324(10)134030  (01)00887868104046(17)250608(10)766440  (01)00887868104046(17)251210(10)559120  (01)00887868104046(17)260208(10)672750  (01)00887868104046(17)290108(10)970140  (01)00887868104039(17)190101(10)700100  (01)00887868104039(17)190515(10)804340R  (01)00887868104039(17)210501(10)066590  (01)00887868104039(17)220901(10)374430  (01)00887868104039(17)230212(10)367210  (01)00887868104039(17)230201(10)516030  (01)00887868104039(17)230318(10)616950  (01)00887868104039(17)230710(10)735990  (01)00887868104039(17)230710(10)732700  (01)00887868104039(17)230710(10)732690  (01)00887868104039(17)230711(10)732680  (01)00887868104039(17)230701(10)735980  (01)00887868104039(17)240110(10)148470  (01)00887868104039(17)240226(10)931570  (01)00887868104039(17)240415(10)141580  (01)00887868104039(17)240417(10)459830  (01)00887868104039(17)240420(10)141540  (01)00887868104039(17)240505(10)592050  (01)00887868104039(17)240520(10)592070  (01)00887868104039(17)240602(10)459820  (01)00887868104039(17)240819(10)764570  (01)00887868104039(17)240918(10)340000  (01)00887868104039(17)250113(10)459460  (01)00887868104039(17)250212(10)770890  (01)00887868104039(17)250212(10)770890R  (01)00887868104039(17)250324(10)134020  (01)00887868104039(17)250324(10)017240  (01)00887868104039(17)250324(10)017220  (01)00887868104039(17)250401(10)181620  (01)00887868104039(17)250421(10)017230  (01)00887868104039(17)251002(10)225590  (01)00887868104039(17)251104(10)283370  (01)00887868104039(17)251209(10)722870  (01)00887868104039(17)260119(10)037930  (01)00887868104039(17)260504(10)714840  (01)00887868104039(17)260608(10)152360  (01)00887868104039(17)260916(10)274610  (01)00887868104039(17)280308(10)716980  (01)00887868104039(17)280309(10)966200  (01)00887868104039(17)280311(10)002320  (01)00887868104039(17)280402(10)716970  (01)00887868104039(17)280403(10)316440  (01)00887868104039(17)280403(10)716990  (01)00887868104039(17)280410(10)966180  (01)00887868104039(17)281110(10)970110  (01)00887868104039(17)281210(10)065580  (01)00887868103940(17)180915(10)483640R  (01)00887868103940(17)190615(10)946040R  (01)00887868103940(17)210401(10)066570  (01)00887868103940(17)211116(10)590200  (01)00887868103940(17)230405(10)968750  (01)00887868103940(17)230509(10)213450  (01)00887868103940(17)230501(10)256630  (01)00887868103940(17)230814(10)765760  (01)00887868103940(17)231218(10)824090  (01)00887868103940(17)240212(10)273070  (01)00887868103940(17)240226(10)931560  (01)00887868103940(17)240803(10)484600  (01)00887868103940(17)241014(10)524200  (01)00887868103940(17)250411(10)360070  (01)00887868103940(17)251207(10)628160  (01)00887868103940(17)260425(10)530930  (01)00887868104008(17)190701(10)089650  (01)00887868103957(17)210501(10)066560  (01)00887868103896(17)181001(10)554000  (01)00887868103896(17)181115(10)592840  (01)00887868103896(17)210701(10)241280  (01)00887868103896(17)230522(10)276250  (01)00887868103896(17)230523(10)276270  (01)00887868103896(17)230529(10)276230  (01)00887868103896(17)230530(10)276210  (01)00887868103896(17)230710(10)732660  (01)00887868103896(17)240414(10)141530  (01)00887868103896(17)240505(10)591950  (01)00887868103896(17)240506(10)459810  (01)00887868103896(17)240506(10)591930  (01)00887868103896(17)240814(10)484580  (01)00887868103896(17)240818(10)950050  (01)00887868103896(17)240819(10)804000  (01)00887868103896(17)240819(10)484560  (01)00887868103896(17)241203(10)229800  (01)00887868103896(17)250212(10)621950  (01)00887868103896(17)250212(10)621960  (01)00887868103896(17)250318(10)016850  (01)00887868103896(17)250320(10)133940  (01)00887868103896(17)250324(10)134010  (01)00887868103896(17)250401(10)134000  (01)00887868103896(17)251117(10)558870  (01)00887868103896(17)260115(10)866260  (01)00887868103896(17)260526(10)186130  (01)00887868103896(17)260526(10)186140  (01)00887868103896(17)260527(10)186120  (01)00887868103896(17)260602(10)152350  (01)00887868103896(17)260621(10)233930  (01)00887868103896(17)260628(10)427190  (01)00887868103896(17)260805(10)751860  (01)00887868103896(17)260829(10)019760  (01)00887868103896(17)260901(10)019750  (01)00887868103896(17)280309(10)966170  (01)00887868103896(17)280311(10)966160  (01)00887868103896(17)280329(10)002300  (01)00887868103896(17)281126(10)970000  (01)00887868103896(17)281126(10)970010  (01)00887868103933(17)180301(10)664410  (01)00887868103933(17)180501(10)118700  (01)00887868103933(17)181201(10)624000  (01)00887868103933(17)190101(10)684540  (01)00887868103933(17)190301(10)831300  (01)00887868103933(17)200201(10)251470  (01)00887868103933(17)200331(10)443950  (01)00887868103926(17)171101(10)086780  (01)00887868103926(17)190115(10)640240R  (01)00887868103926(17)180701(10)305580  (01)00887868103926(17)180801(10)483630  (01)00887868103926(17)181001(10)516900  (01)00887868103926(17)181001(10)553990  (01)00887868103926(17)181101(10)601640  (01)00887868103926(17)190501(10)884670  (01)00887868103926(17)200201(10)443940  (01)00887868103926(17)200301(10)443940R  (01)00887868103926(17)230219(10)516010  (01)00887868103926(17)230226(10)616940  (01)00887868103926(17)230522(10)276200  (01)00887868103926(17)240618(10)950040  (01)00887868103926(17)250319(10)016820  (01)00887868103926(17)250325(10)133990  (01)00887868103926(17)250417(10)256190  (01)00887868103926(17)251207(10)558820  (01)00887868103926(17)251207(10)558850  (01)00887868103926(17)251207(10)628130  (01)00887868103926(17)260502(10)714830  (01)00887868103926(17)260613(10)152340  (01)00887868103926(17)260618(10)186100  (01)00887868103926(17)260620(10)186110  (01)00887868103926(17)280303(10)966110  (01)00887868103926(17)280308(10)966150  (01)00887868103926(17)280309(10)966080  (01)00887868103926(17)280309(10)966090  (01)00887868103926(17)280309(10)966140  (01)00887868103926(17)280323(10)002280  (01)00887868103926(17)280323(10)966130  (01)00887868103926(17)281125(10)969970  (01)00887868103926(17)281127(10)969950  (01)00887868103919(17)180201(10)540810  (01)00887868103919(17)181201(10)640230  (01)00887868103919(17)190501(10)924690  (01)00887868103919(17)200301(10)443930  (01)00887868103919(17)200401(10)443930R  (01)00887868103919(17)201101(10)621020  (01)00887868103919(17)210401(10)066550  (01)00887868103919(17)211001(10)241250  (01)00887868103919(17)211201(10)445640  (01)00887868103919(17)220228(10)340650R  (01)00887868103919(17)230522(10)276170  (01)00887868103919(17)230524(10)276190  (01)00887868103919(17)250211(10)770840  (01)00887868103919(17)250612(10)923930  (01)00887868103919(17)251207(10)628110  (01)00887868103919(17)260609(10)152330  (01)00887868103902(17)180301(10)740720  (01)00887868103902(17)181001(10)553970  (01)00887868103902(17)181001(10)553980  (01)00887868103902(17)190101(10)640220  (01)00887868103902(17)190101(10)665990  (01)00887868103902(17)190115(10)601630R  (01)00887868103902(17)190301(10)736760  (01)00887868103902(17)201101(10)898440  (01)00887868103902(17)210401(10)066540  (01)00887868103902(17)221001(10)399770  (01)00887868103902(17)230219(10)516000  (01)00887868103902(17)230319(10)616920  (01)00887868103902(17)230509(10)213440  (01)00887868103902(17)230521(10)276110  (01)00887868103902(17)230528(10)276180  (01)00887868103902(17)230711(10)732650  (01)00887868103902(17)230711(10)732640  (01)00887868103902(17)230807(10)732630  (01)00887868103902(17)231215(10)824070  (01)00887868103902(17)231215(10)824060  (01)00887868103902(17)240209(10)612680  (01)00887868103902(17)240414(10)459790  (01)00887868103902(17)240415(10)459800  (01)00887868103902(17)240602(10)591830  (01)00887868103902(17)240602(10)591780  (01)00887868103902(17)240610(10)591860  (01)00887868103902(17)240611(10)591800  (01)00887868103902(17)240611(10)591810  (01)00887868103902(17)240623(10)949950  (01)00887868103902(17)240805(10)484540  (01)00887868103902(17)240807(10)484520  (01)00887868103902(17)240812(10)484530  (01)00887868103902(17)240817(10)804040  (01)00887868103902(17)240819(10)484510  (01)00887868103902(17)241111(10)785530  (01)00887868103902(17)241112(10)973730  (01)00887868103902(17)241202(10)621760  (01)00887868103902(17)250116(10)099330  (01)00887868103902(17)250213(10)770740  (01)00887868103902(17)250218(10)877440  (01)00887868103902(17)250319(10)133980  (01)00887868103902(17)250324(10)133950  (01)00887868103902(17)250325(10)133960  (01)00887868103902(17)250331(10)133970  (01)00887868103902(17)251207(10)628090  (01)00887868103902(17)251209(10)558800  (01)00887868103902(17)251217(10)941270  (01)00887868103902(17)260119(10)310500  (01)00887868103902(17)260204(10)588250  (01)00887868103902(17)260527(10)186080  (01)00887868103902(17)260527(10)186080R  (01)00887868103902(17)260602(10)152320  (01)00887868103902(17)260616(10)252460  (01)00887868103902(17)260617(10)186090  (01)00887868103902(17)260621(10)233920  (01)00887868103902(17)260711(10)571910  (01)00887868103902(17)260714(10)571920  (01)00887868103902(17)270908(10)202720  (01)00887868103902(17)271012(10)199210  (01)00887868103902(17)280302(10)002140  (01)00887868103902(17)280307(10)716940  (01)00887868103902(17)280308(10)716930  (01)00887868103902(17)280308(10)717110  (01)00887868103902(17)280313(10)717120  (01)00887868103902(17)280320(10)002210  (01)00887868103902(17)280320(10)716950  (01)00887868103902(17)280322(10)716960  (01)00887868103902(17)2803",
      "center_classification_date": "20191122",
      "distribution_pattern": "US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.",
      "state": "IN",
      "product_description": "HA Pins (DFS HA Cort Screw), Hoffman/Colles Pins (Colles CSER Sterile w/ Pin Kit)    Item Nos: HAA60-09030  HAA60-09040  HAA60-10030  HAA60-10040  HAA60-11030  HAA60-11040  HAA60-11050  HAA60-12040  HAA60-12060  HAA60-13030  HAA60-13040  HAA60-13050  HAA60-13060  HAA60-13070  HAA60-14050  HAA60-14080  HAA60-14090  HAA60-15020  HAA60-15030  HAA60-15040  HAA60-15050  HAA60-15060  HAA60-16030  HAA60-16040  HAA60-16050  HAA60-16070  HAA60-16090  HAA60-17040  HAA60-17060  HAA60-17080  HAA60-18010  HAA60-18030  HAA60-18040  HAA60-18050  HAA60-18060  HAA60-18090  HAA60-20030  HAA60-20040  HAA60-20050  HAA60-20060  HAA60-20080  HAA60-20090  HAA60-22050  HAA60-22060  HAA60-25030  HAA60-25040  HAA60-25050  HAA60-25060  HAA60-30060  HAB60-09030  HAB60-10030  HAB60-10040  HAB60-11040  HAB60-11050  HAB60-12040  HAB60-12060  HAB60-13040  HAB60-13050  HAB60-13060  HAB60-14050  HAB60-15060  HAB60-16070  HAB60-16090  HAB60-17080  HAB60-18010  HAB60-18090  HAB60-20080  HAB60-20090  808200000  Product Usage: These devices are intended for osteosynthesis, external fixation of fractures or bone lengthening procedures.",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0528-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83783",
      "termination_date": "20210721",
      "more_code_info": "28(10)316350  (01)00887868103902(17)280329(10)002260  (01)00887868103902(17)280330(10)002230  (01)00887868103902(17)280402(10)316330  (01)00887868103902(17)280403(10)316390  (01)00887868103902(17)280405(10)316340  (01)00887868103902(17)281110(10)969940  (01)00887868103834(17)180101(10)425100  (01)00887868103834(17)181201(10)601620  (01)00887868103834(17)190101(10)700080  (01)00887868103834(17)190201(10)718350  (01)00887868103834(17)190501(10)884660  (01)00887868103834(17)190615(10)946030R  (01)00887868103834(17)210601(10)066510  (01)00887868103834(17)211001(10)241200  (01)00887868103834(17)230228(10)515960  (01)00887868103834(17)230401(10)798930  (01)00887868103834(17)230402(10)921350  (01)00887868103834(17)230405(10)968760  (01)00887868103834(17)230418(10)070840  (01)00887868103834(17)230510(10)089790  (01)00887868103834(17)230711(10)611580  (01)00887868103834(17)230725(10)765740  (01)00887868103834(17)230725(10)765750  (01)00887868103834(17)231120(10)129930  (01)00887868103834(17)240110(10)148460  (01)00887868103834(17)240211(10)989920  (01)00887868103834(17)240226(10)931540  (01)00887868103834(17)240228(10)733210  (01)00887868103834(17)240423(10)141520  (01)00887868103834(17)240624(10)949920  (01)00887868103834(17)240728(10)484500  (01)00887868103834(17)240819(10)804010  (01)00887868103834(17)240821(10)484490  (01)00887868103834(17)241118(10)524190  (01)00887868103834(17)250212(10)621940  (01)00887868103834(17)250318(10)133930  (01)00887868103834(17)250322(10)133910  (01)00887868103834(17)250416(10)597140  (01)00887868103834(17)250609(10)855270  (01)00887868103834(17)250609(10)948470  (01)00887868103834(17)250701(10)861810  (01)00887868103834(17)250709(10)364660  (01)00887868103834(17)251209(10)558780  (01)00887868103834(17)260115(10)866240  (01)00887868103834(17)260418(10)162650  (01)00887868103834(17)260424(10)530920  (01)00887868103834(17)260509(10)817170  (01)00887868103834(17)280308(10)002200  (01)00887868103834(17)280309(10)002150  (01)00887868103834(17)280321(10)966060  (01)00887868103834(17)280402(10)966050  (01)00887868103889(17)180615(10)118680R  (01)00887868103889(17)180801(10)450650  (01)00887868103889(17)190101(10)665980  (01)00887868103889(17)190201(10)771020  (01)00887868103889(17)190515(10)889050R  (01)00887868103889(17)190701(10)004690  (01)00887868103889(17)191201(10)250400  (01)00887868103889(17)200201(10)443860  (01)00887868103889(17)200201(10)251420  (01)00887868103889(17)210601(10)241190  (01)00887868103889(17)280309(10)966040  (01)00887868103858(17)280403(10)966020  (01)00887868103803(17)240618(10)177140  (01)00887868103827(17)180301(10)664400  (01)00887868103827(17)180501(10)118670  (01)00887868103827(17)181101(10)547010  (01)00887868103827(17)190101(10)677520  (01)00887868103827(17)210601(10)241140  (01)00887868103810(17)181001(10)516890  (01)00887868103810(17)190401(10)816120  (01)00887868103810(17)190501(10)924660  (01)00887868103810(17)190501(10)884650  (01)00887868103810(17)200301(10)443810R  (01)00887868103810(17)201201(10)984450  (01)00887868103810(17)210301(10)057480  (01)00887868103810(17)220401(10)353090  (01)00887868103810(17)230201(10)515890  (01)00887868103810(17)230218(10)451250  (01)00887868103810(17)230318(10)692180  (01)00887868103810(17)230318(10)692180R  (01)00887868103810(17)230520(10)276080  (01)00887868103810(17)230523(10)276100  (01)00887868103810(17)230529(10)276070  (01)00887868103810(17)240212(10)612670  (01)00887868103810(17)240303(10)931530  (01)00887868103810(17)240408(10)141490  (01)00887868103810(17)240414(10)141510  (01)00887868103810(17)240420(10)141500  (01)00887868103810(17)240511(10)591680  (01)00887868103810(17)240520(10)591690  (01)00887868103810(17)240602(10)591740  (01)00887868103810(17)240818(10)779490  (01)00887868103810(17)240819(10)779500  (01)00887868103810(17)240820(10)779480  (01)00887868103810(17)250115(10)459430  (01)00887868103810(17)250316(10)016540  (01)00887868103810(17)250420(10)227400  (01)00887868103810(17)250806(10)723150  (01)00887868103810(17)250918(10)907320  (01)00887868103810(17)251014(10)254000  (01)00887868103810(17)251208(10)628070  (01)00887868103810(17)260407(10)162630  (01)00887868103810(17)260418(10)162640  (01)00887868103810(17)260509(10)714820  (01)00887868103810(17)260601(10)152310  (01)00887868103810(17)260609(10)299060  (01)00887868103810(17)260616(10)186060  (01)00887868103810(17)260630(10)427180  (01)00887868103810(17)260803(10)694920  (01)00887868103810(17)260803(10)751850  (01)00887868103810(17)260916(10)274590  (01)00887868103810(17)280308(10)002120  (01)00887868103810(17)280308(10)717070  (01)00887868103810(17)280308(10)717080  (01)00887868103810(17)280308(10)717090  (01)00887868103810(17)280309(10)717100  (01)00887868103810(17)280326(10)002040  (01)00887868103810(17)280326(10)002070  (01)00887868103810(17)280326(10)002040R  (01)00887868103810(17)280326(10)002070R  (01)00887868103810(17)280329(10)152450  (01)00887868103810(17)280329(10)716920  (01)00887868103810(17)281110(10)065530  (01)00887868103810(17)281110(10)671270  (01)00887868103810(17)281124(10)671260  (01)00887868103810(17)281220(10)065510  (01)00887868103810(17)281227(10)969830  (01)00887868103797(17)180301(10)740710  (01)00887868103797(17)190501(10)884640  (01)00887868103797(17)201201(10)984430  (01)00887868103797(17)210401(10)092130  (01)00887868103797(17)210501(10)066260  (01)00887868103797(17)221101(10)839580  (01)00887868103797(17)230219(10)515830  (01)00887868103797(17)230226(10)616850  (01)00887868103797(17)230301(10)671250  (01)00887868103797(17)230401(10)798910  (01)00887868103797(17)230401(10)750090  (01)00887868103797(17)230405(10)921340  (01)00887868103797(17)230521(10)276060  (01)00887868103797(17)230528(10)276050  (01)00887868103797(17)230619(10)276040  (01)00887868103797(17)230711(10)732600  (01)00887868103797(17)230807(10)795650  (01)00887868103797(17)230807(10)774120  (01)00887868103797(17)240113(10)148410  (01)00887868103797(17)240212(10)273060  (01)00887868103797(17)240217(10)588290  (01)00887868103797(17)240303(10)931520  (01)00887868103797(17)240324(10)141470  (01)00887868103797(17)240331(10)141420  (01)00887868103797(17)240414(10)459740  (01)00887868103797(17)240415(10)141410  (01)00887868103797(17)240422(10)141480  (01)00887868103797(17)240506(10)459730  (01)00887868103797(17)240514(10)459770  (01)00887868103797(17)240521(10)589690  (01)00887868103797(17)240526(10)589700  (01)00887868103797(17)240602(10)459760  (01)00887868103797(17)240610(10)459780  (01)00887868103797(17)240611(10)589680  (01)00887868103797(17)240616(10)949850  (01)00887868103797(17)240617(10)050790  (01)00887868103797(17)240617(10)141390  (01)00887868103797(17)240624(10)949860  (01)00887868103797(17)240727(10)484430  (01)00887868103797(17)240730(10)484470  (01)00887868103797(17)240830(10)484450  (01)00887868103797(17)240817(10)484440  (01)00887868103797(17)240818(10)484460  (01)00887868103797(17)240819(10)778910  (01)00887868103797(17)240819(10)778920  (01)00887868103797(17)240821(10)484410  (01)00887868103797(17)240821(10)484480  (01)00887868103797(17)240821(10)778900  (01)00887868103797(17)250331(10)133890  (01)00887868103797(17)250401(10)181600  (01)00887868103797(17)250401(10)016230  (01)00887868103797(17)250622(10)019570  (01)00887868103797(17)250807(10)736140  (01)00887868103797(17)251002(10)225130  (01)00887868103797(17)260202(10)365790  (01)00887868103797(17)260203(10)627760  (01)00887868103797(17)260203(10)627820  (01)00887868103797(17)260407(10)123440  (01)00887868103797(17)260407(10)144140  (01)00887868103797(17)260414(10)162620  (01)00887868103797(17)260422(10)123400  (01)00887868103797(17)260502(10)752490  (01)00887868103797(17)260502(10)674960  (01)00887868103797(17)260503(10)714810  (01)00887868103797(17)260608(10)152300  (01)00887868103797(17)260616(10)252450  (01)00887868103797(17)260624(10)392950  (01)00887868103797(17)260630(10)392960  (01)00887868103797(17)260705(10)597010  (01)00887868103797(17)260730(10)694910  (01)00887868103797(17)260816(10)751840  (01)00887868103797(17)260816(10)873310  (01)00887868103797(17)270907(10)431520  (01)00887868103797(17)270907(10)431520  (01)00887868103797(17)271002(10)202670  (01)00887868103797(17)271013(10)202690  (01)00887868103797(17)280121(10)202680  (01)00887868103797(17)280301(10)001880  (01)00887868103797(17)280307(10)001990  (01)00887868103797(17)280308(10)001930  (01)00887868103797(17)280308(10)001970  (01)00887868103797(17)280309(10)001900  (01)00887868103797(17)280311(10)716880  (01)00887868103797(17)280320(10)001920  (01)00887868103797(17)280323(10)202700  (01)00887868103797(17)280326(10)716890  (01)00887868103797(17)280327(10)716900  (01)00887868103797(17)280327(10)002020  (01)00887868103797(17)280327(10)152460  (01)00887868103797(17)280402(10)716910  (01)00887868103797(17)281110(10)671170  (01)00887868103797(17)281227(10)065490  (01)00887868103780(17)180101(10)424500  (01)00887868103780(17)180301(10)740700  (01)00887868103780(17)181101(10)630570  (01)00887868103780(17)190301(10)829220  (01)00887868103780(17)190501(10)970150  (01)00887868103780(17)190601(10)002080  (01)00887868103780(17)200201(10)251360  (01)00887868103780(17)200201(10)443710  (01)00887868103780(17)200301(10)443710R  (01)00887868103780(17)260425(10)530910  (01)00887868103780(17)280306(10)966010  (01)00887868103766(17)180401(10)851860  (01)00887868103766(17)180801(10)456460  (01)00887868103766(17)190401(10)825300  (01)00887868103766(17)190401(10)825350  (01)00887868103766(17)190501(10)889020  (01)00887868103759(17)171201(10)140580  (01)00887868103759(17)180401(10)851850  (01)00887868103759(17)180701(10)375390  (01)00887868103759(17)190601(10)004660  (01)00887868103759(17)201201(10)984410  (01)00887868103230(17)180901(10)523550  (01)00887868103230(17)200201(10)251350  (01)00887868103230(17)210901(10)241060  (01)00887868103230(17)220901(10)503730  (01)00887868103230(17)230218(10)515820  (01)00887868103230(17)230226(10)616830  (01)00887868103230(17)230318(10)692170  (01)00887868103230(17)230520(10)276000  (01)00887868103230(17)230520(10)276010  (01)00887868103230(17)230520(10)276030  (01)00887868103230(17)230524(10)276020  (01)00887868103230(17)230711(10)732550  (01)00887868103230(17)231108(10)758800  (01)00887868103230(17)231118(10)758820  (01)00887868103230(17)231120(10)758780  (01)00887868103230(17)231120(10)758810  (01)00887868103230(17)231214(10)823750  (01)00887868103230(17)231218(10)823700  (01)00887868103230(17)231218(10)823740  (01)00887868103230(17)240327(10)050630  (01)00887868103230(17)240331(10)050680  (01)00887868103230(17)240406(10)141250  (01)00887868103230(17)240414(10)453470  (01)00887868103230(17)240416(10)459710  (01)00887868103230(17)240425(10)141270  (01)00887868103230(17)240427(10)141260  (01)00887868103230(17)240427(10)459720  (01)00887868103230(17)240507(10)453370  (01)00887868103230(17)240511(10)141280  (01)00887868103230(17)240521(10)459700  (01)00887868103230(17)240602(10)589670  (01)00887868103230(17)240616(10)948660  (01)00887868103230(17)240617(10)589620  (01)00887868103230(17)240618(10)589620R  (01)00887868103230(17)240619(10)589630  (01)00887868103230(17)240625(10)948580  (01)00887868103230(17)240625(10)948650  (01)00887868103230(17)240720(10)484390  (01)00887868103230(17)240806(10)484400  (01)00887868103230(17)240810(10)484370  (01)00887868103230(17)240818(10)484360  (01)00887868103230(17)240820(10)778890  (01)00887868103230(17)240820(10)778880  (01)00887868103230(17)240904(10)778770  (01)00887868103230(17)241114(10)099270  (01)00887868103230(17)241203(10)229750  (01)00887868103230(17)250126(10)330280  (01)00887868103230(17)250321(10)016170  (01)00887868103230(17)250323(10)133870  (01)00887868103230(17)250327(10)016150  (01)00887868103230(17)250401(10)165130  (01)00887868103230(17)250401(10)016210  (01)00887868103230(17)250402(10)256030  (01)00887868103230(17)250407(10)133860  (01)00887868103230(17)250415(10)016200  (01)00887868103230(17)250615(10)019540  (01)00887868103230(17)250925(10)097760  (01)00887868103230(17)251001(10)097740  (01)00887868103230(17)251002(10)097710  (01)00887868103230(17)251002(10)097750  (01)00887868103230(17)251005(10)097720  (01)00887868103230(17)251019(10)097730  (01)00887868103230(17)251117(10)097700  (01)00887868103230(17)260426(10)530900  (01)00887868103230(17)260501(10)714800  (01)00887868103230(17)260505(10)714790  (01)00887868103230(17)260505(10)817160  (01)00887868103230(17)260506(10)674950  (01)00887868103230(17)260506(10)752480  (01)00887868103230(17)260608(10)152290  (01)00887868103230(17)260701(10)392940  (01)00887868103230(17)260707(10)392930  (01)00887868103230(17)260731(10)694900  (01)00887868103230(17)260801(10)737410  (01)00887868103230(17)260802(10)737400  (01)00887868103230(17)260802(10)737430  (01)00887868103230(17)260803(10)737420  (01)00887868103230(17)260822(10)019720  (01)00887868103230(17)260826(10)019700  (01)00887868103230(17)260902(10)019690  (01)00887868103230(17)260902(10)019710  (01)00887868103230(17)261018(10)431510  (01)00887868103230(17)270912(10)202650  (01)00887868103230(17)280222(10)716870  (01)00887868103230(17)280228(10)717020  (01)00887868103230(17)280304(10)716850  (01)00887868103230(17)280309(10)001780  (01)00887868103230(17)280315(10)001820  (01)00887868103230(17)280316(10)001850  (01)00887868103230(17)280316(10)316140  (01)00887868103230(17)280316(10)717030  (01)00887868103230(17)280324(10)316210  (01)00887868103230(17)280402(10)316190  (01)00887868103230(17)280403(10)316180  (01)00887868103230(17)280409(10)716860  (01)00887868103230(17)280412(10)717060  (01)00887868103230(17)280529(10)001800  (01)00887868103230(17)280821(10)001830  (01)00887868103230(17)280822(10)001870  (01)00887868103230(17)280822(10)717050  (01)00887868103735(17)180701(10)441440  (01)00887868103735(17)190301(10)825110  (01)00887868103735(17)191101(10)141510  (01)00887868103735(17)200304(10)443600  (01)00887868103735(17)260426(10)519060  (01)00887868104206(17)190201(10)775700  (01)00887868104374(17)180601(10)298750  (01)00887868104374(17)250713(10)019590  (01)00887868104350(17)180101(10)322110  (01)00887868104350(17)180702(10)344380  (01)00887868104350(17)190401(10)856280  (01)00887868104350(17)190601(10)979890  (01)00887868104367(17)181101(10)649290  (01)00887868104336(17)180401(10)065110  (01)00887868104336(17)190601(10)979880  (01)00887868104336(17)201201(10)444710  (01)00887868104336(17)210601(10)550170  (01)00887868104336(17)251210(10)591610  (01)00887868104299(17)180501(10)118750  (01)00887868104299(17)180801(10)456580  (01)00887868104299(17)190701(10)929510  (01)00887868104299(17)211101(10)603610  (01)00887868104329(17)180101(10)322100  (01)00887868104329(17)180801(10)450790  (01)00887868104329(17)190601(10)979860  (01)00887868104329(17)190601(10)979870  (01)00887868104312(17)180501(10)005020  (01)00887868104312(17)190201(10)775710  (01)00887868104312(17)190401(10)856240  (01)00887868104312(17)190603(10)979850  (01)00887868104305(17)180601(10)298740  (01)00887868104305(17)190501(10)879330  (01)00887868104305(17)190506(10)879340  (01)00887868104282(17)190401(10)856220  (01)00887868104282(17)190501(10)961010  (01)00887868104251(17)231113(10)913830  (01)00887868104275(17)171201(10)212710  (01)00887868104275(17)181201(10)601660  (01)00887868104275(17)190201(10)704560  (01)00887868104275(17)190501(10)979840  (01)00887868104275(17)190604(10)979830  (01)00887868104275(17)210906(10)550430  (01)00887868104268(17)201001(10)550420  (01)00887868104268(17)250418(10)165140  (01)00887868104244(17)180101(10)238210  (01)00887868104244(17)180401(10)005000  (01)00887868104244(17)180806(10)375420  (01)00887868104244(17)190401(10)856190  (01)00887868104244(17)210401(10)444610  (01)00887868104244(17)211201(10)550410R  (01)00887868104237(17)180501(10)118730  (01)00887868104237(17)190501(10)856180R  (01)00887868104237(17)190701(10)979790  (01)00887868104183(17)180701(10)338860  (01)00887868104183(17)190301(10)856170  (01)00887868104213(17)190401(10)856160    (01)00887868034657(17)191022(10)658260  (01)00887868034657(17)191103(10)911110  (01)00887868034657(17)200515(10)044380  (01)00887868034657(17)200806(10)487720  (01)00887868034657(17)210315(10)487740  (01)00887868034657(17)210310(10)139190  (01)00887868034657(17)210315(10)366770  (01)00887868034657(17)210317(10)482470  (01)00887868034657(17)210331(10)373650  (01)00887868034657(17)210713(10)622290  (01)00887868034657(17)210713(10)622460",
      "recall_initiation_date": "20190821",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}