{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Golden",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86047",
      "recalling_firm": "CME America, LLC",
      "address_1": "14998 W 6th Ave Ste 830",
      "address_2": "N/A",
      "postal_code": "80401-5025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.",
      "recall_number": "Z-0527-2021",
      "product_description": "CME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).",
      "product_quantity": "Total pumps = 4,182 (all models)",
      "reason_for_recall": "Multiple reasons:  1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated.  2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.",
      "recall_initiation_date": "20200730",
      "center_classification_date": "20201203",
      "termination_date": "20220308",
      "report_date": "20201209",
      "code_info": "Model T34L PCA - Part Number(s) (UDI Codes)  100-106PSL  (UDI Code: B101100106PSL0)  Serial Number:  26486  26485  26484  S93716    100-106PSLM (retired prior to UDI)",
      "more_code_info": ""
    }
  ]
}