  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.",
      "address_2": "",
      "product_quantity": "219988 units (total)",
      "code_info": "All lots manufactured between January 2008 - May 2019:    197000  838570  047410  869160  047400  047390  203030  203040  224470  482850  052950  076580  917420  326610  362350  565380  374150  420170  374160  882210  882090  746150  691040  969190  942350  726180  048880  385080  111310  013150  868010  638680  745730  780270  780240  013140  834290  623340  817490  028380  186120  409420  847030  679990  755600  409330  706270  169490  936000  400990  343410  235880  917410  872790  715870  179470  313330  324140  362650  305750  326590  920210  564070  305740  632120  684640  536380  564770  014640R  745910  629320  629350  014640  807570  946770  111280  946800  735450  108390  370610  385070  884110  013160  884080  621660  919220  836010  645380  706600  899340  712500  838730  963630  021010  896300  673910  669190  146510  613110  678830  335900  684340  106100  105780  140000  447810  154700  911280  363670  102500  931980  519820  114830  762310  076570  296510  917390  096130  059170  362810  552990  684260  587040  565220  484880  634740  485340  414520  906620  808780  678310  969290  013240  013250  014670  014850  384890  108310  808790  561410  808770  336710  018410  617860  313520  378380  579070  566190  927200  297800  673900  348320  702090  939400  736230  268220  055290  347210  347230  342820  725450  817310  153990  725430  698880  698890  823840  188130  037810  705530  602470  301330  463540  940980  028450  304010  491810  596490  684830  771630  731090  384570  343400  049790  057690  766800  513670  114820  915280  408980  715850  917370  917380  296500  076560  326600  510940  568130  363140  420030R  420030  808760  099520  917350  414510  629400  808720  014610  813330  044040  484870  621620  694280  694250  739520  991720  563690  671840  983580  250250  190920  231850  390870  757150  799950  517900  204080  764010  114810  249030  915250  061900  917340  076550  917360  131600  320520  348900  326030  565370  788320  610590  200890  462920  540650R  584540  540650  540610  691160  634350  485330  808710  085020  108330  906800  042230  367330  013270  100260  584530  759100  621970  808700  781920  834590  645360  654830  307730  580280  789060  295500  110720  296810  218340  971510  290820  984540  715090  130360  280430  896280  701290  322680  582620  912810  911710  178270  591690  191690  367880  867300  112070  266870  798940  568280  613100  028440  620410  620420  784160  046210  712950  443870  642180  503770  420330  134180  294890  628730  763910  251590  096000  381290  086080  208040  304890  314410  498860  498870  546180  831840  419740  828320  767110  917310  991930  771690  296490  260290  235790  917330  076520  076540  427590  326690  326700  484860  674000  540570  584510  584520  808670  553580  484850  746060  683370  808660  946820  049430  125700  385030  677300  917320  645350  371650  371910  919370  808680  089930  089920  485520  606500  208850  502490  298670  656410  045750  085450  313510  745020  721210  144450  378370  712490  431600  433430  297790  760970  842280  174340  250930  297880  934950  206630  912780  434850  201690  915720  182160  627310  411050  450540  644770  191740  763920  944360  613090  168190  716270  824710  565110  759500  279840  751390  313950  297070  102460  399850  728220  741280  062670  599590  070530  958220  079180  917290  917300  052930  132080  132090  261190  193570  227680  378250  326580  363150  305470  553570  787630  484840  634340  606650  227710  674890  654520  866510  399620  485320  808640  055710  942310  795900  969300  808650  055720  744940  561400  056140  809740  477760  354200  896230  322660  483840  930010  135000  329910  991820  449680  192060  479170  738870  626460  715840  917270  320620  881600  540140  808620  553610  808630  485310  606660  553611  808510  055700  758650  143790  055690  552750  422710  751380  973120  763930  233200  503760  305760  211440  715830  791670  035100  457730  209330  965510  324130  052920  911290  339120  586290  540090  364180  586300  276570  162030  162040  553940  785680  852640  054840  382030  374030  805470  134580  790650  302140  547610R  297990  547610  896180  356960  771130  372160  747300  787100  688710  343770  823820  352910  611740  475630  768550  768550R  465490  919980  485950  139990  284880  342280  266180  728880  875320  541510  652040  747580  331370  327140  169040  866470  715820  261020  245080  275200  308520  275190  296480  250150  245050  374480  563990  563970  558620  558630  586280  561200  561210  558610  786920  865980  865990  385950  196500  044100  973720  373980  301080  301090  373990  374000  772460  697320  021410  748760  987390  165460  986940  383960  245320  429590  358790  095990  018850  293160  293170  886190  855020  941050  965160  855030  815540  992450  171270  172200  934030  934020  299320  337530  971820  179730  299340  482040  499120  701660  374100  646370  373540  374110  684180  705520  646320  944070  944380  906950  802610  802590  057210  969620  144420  296660  144410  144430  429290  712460  183950  680430  760910  715080  917470  760980  242270  968580  132950  297960  297810  297850  399380  297820  673960  673920  896130  936160  939390  364060  770560  415910  322670  471660  740640  687610  947490  025190  574890  851130  627290  787090  787080  531340  135140  309650  191670  191680  547500  367870  450150  725280  688670  992130  787730  093270  797850  602450  602460  134990  686500  768520  824690  809340  991800  336150  946080  335880  201960  684430  505740  490180  759490  117200  024480  252230  406320  132610  143430  409410  925400  682980  435670  561500  342270  271940  827340  139770  161980  453360  542830  522760  875310  039180  473210  781530  610870  294890  947310  070400  069880  026340  372930  430830  819690  399840  496920  886220  980410  416120  363660  164590  821940  115710  503750  212310  462060  120690  881240  881180  079230  169030  079220  531660  327130  090650  211420  845510  866440  090660  045460  819530  700180  972280  500720  209350  081430  304580  304620  304590  304610  232410  304600  304630  304640  209360  304650  308090  304670  209340  304680  363390  304660  439320  308270  552440  439330  586260  395930  564370  395960  586270  865920  865910  081420  865900  014180  276590  188720  188710  188700  289960  308150  305600  305610  314160  742650  874490  822830  990810  484530  933900  043010  336660  366540  624460  292740  426590  021300  608740  757630  125560  125550  125570  256220  256230  256250  256260  256270  256240  301800  301810  256280  346750  398940  533890  533900  322450  463080  322460  606490  671690  398950  671680  016951  859320  242750  031320  031310  136950  601010  295710  580240  580260  580270  483470  580250  656390  603690  603710  603700  912390  912440  912410  912360  912460  108920  044790  045110  912350  912370  912400  912430  971030  170520  117340  117380  254630  316930  330330  117230  117240  179720  117250  218290  491020  490960  381410  381160  898760  144330  144320  302070  997400  997020  712510  406860  874810  547650  117620  712480  227950  258090  442140  441380  535900  651460  578700  487770  650360  715530  680500  995880  650300  968570  968560  715070  229180  760730  964980  399420  297940  297950  399410  570680  399430  297930  673950  954480  896270  939380  033680  983850  985750  109670  251080  251090  322640  519470  687600  693640  749050  915700  955480  142250  021050  419750  273850  531290  604870  627280  787070  627270  706110  240900  084560  367830  367860  367820  367840  669170  725250  367850  669140  688700  669130  669160  698670  688690  669180  458280  669120  883600  787720  125880  622120  602440  602430  134980  910760  205920  416030  329900  611720  686490  898590  037300  037310  063740  946070  965570  824680  991790  069560  809330  946060  113230  201950  335870  671830  006130  085250  160010  359270  422910  423070  670240  970390  057550  906250  231030  455030  455360  266250  406310  162310  649440  649430  828560  766990  870250  945370  143410  143420  410710  421170  880810  506030  682970  919970  785240  551280  119550  751370  561490  485930  003950  409830  243880  649340  271930  139980  522000  453340  542810  542820  668240  453350  126380  589770  313940  313930  555550  712940  875300  039170  214050  154780  673520  685010  935690  294870  935680  411780  452970  580230  940620  299340  584000  819680  053300  819670  289370  496800  496770  496780  399830  095080  164510  095090  744250  652030  685720  095070  115700  115670  171340  115690  503740  115680  893580  747570  212300  436640  841670  705920  866330  866320  169020  247060  547110  595380  595410  233730  531650  169000  595390  845490  090630  845500  090620  881170  866410  683040  090640  626430  211410  866400  626390  560760  035110  709380  052900  457800  132070  540070  539990  361570  586240  586250  605810  052880  886710  967150  886550  013320  013370  278090  668560  834550  656110  790070  656100  174700  933990  992440  701930  685180  374040  749140  237030  144310  277060  277070  555150  302060  594510  962630  442090  058150  297830  896150  673940  297870  487570  566520  566540  191660  658790  688660  893550  372670  824670  335860  024470  682960  421160  485920  729180  925510  485940  424440  353450  000130  522010  542800  927390  268300  555540  039150  552140  164570  685710  989050  533590  531640  866390  700170  915670  762330  971710  917650  052850  052860  132050  276630  132060  361560  416000  439280  416010  439310  439300  439260  361540  439270  439290  609990  876830  832940  876840  876850  013400  013380  144040  771280  935470  935480  339050  373950  301070  373970  483060  435000  813570  715820  715810  199740  199750  199760  898100  919050  391980  383940  545820  383950  018840  870250  095970  846240  431610  148760  431620  693580  341560  886180  698720  932880  012030  616130  616150  729610  398630  162020  162000  189050  279040  162010  289360  392050  600250  392060  580320  631730  710790  891500  553930  852630  934010  010230  972020  155640  226070  117570  179670  179680  505250  570330  373510  373500  374050  373520  373490  639310  566150  786760  786190  786780  841980  898030  898050  898040  898020  134530  282140  898750  452970  431680  277050  302050  549460  302040  408370  406850  710750  927560  968540  227940  258110  113160  673420  441370  552900  680450  650350  715060  014620  916220  650310  940490  048900  968550  283510  089290  715160  148040  148050  229070  121320  534520  986580  986570  033670  251070  322610  394060  322540  332100  394080  325680  553970  554000  687580  867010  603250  711220  762890  820860  819490  553980  693630  749040  687590  746390  714720  831850  021040  178260  296580  695730  574880  787060  627260  123210  191650  084550  453810  191620  191630  276180  453790  367810  453800  688680  823800  911030  892820  093190  498410  498370  498400  498420  622110  317640  910770  236680  329880  686480  611700  686470  766080  037290  037280  686240  824660  824640  965550  946040  321670  965560  991780  334820  476440  490200  113220  511220  696970  671820  847280  024460  013420  175540  429810  454110  649410  649420  649390  852280  933990  132600  945360  143400  503920  745440  374190  506020  435660  682950  421150  919950  919960  254820  843980  379540  396960  368060  550900  082420  975920  927580  927590  214760  271920  347570  284940  214750  522170  563850  594840  403940  342260  681280  542790  453330  542780  815540  982470  927350  268150  927320  927570  498820  039140  297060  151700  294860  452960  845700  940610  583990  894920  050620  247840  981160  496810  825210  399810  416110  164490  741270  095060  769680  164500  652010  060010  548410  523030  115660  961870  212280  479190  705910  120680  881220  168980  079190  866360  531630  533580  845450  090570  595340  845480  090580  090590  090610  045440  045450  626360  626380  687850  866380  907610  251580  095980  191250  406800  042590  179690  314420  374080  374090  478790  478800  646330  304880  208030  188710  715810  700140  500710  431800  097420  279660  304530  304520  304560  097410  719660  097390  431790  097400  304550  324120  510880  304570  439250  469810  586310  469830  539930  469820  586340  539830  539960  586230  439240  467930  586320  746210  415990  438920  583490  608790  583410  876820  876810  876800  583430  876790  967110  967130  012890  013470  917640  044110  279910  512330  044090  013010  669230  143910  199730  384890  390050  391950  742820  391960  877280  545810  985610  018830  148750  251570  039120  552940  341550  431590  431600  677340  159200  837890  295670  295680  295650  295660  161990  289370  289350  642680  852620  580300  790080  710780  972010  140520  170480  187670  179710  334720  179660  992430  570190  373460  237080  632890  373480  639230  373470  568300  646360  802580  898010  898740  802570  277030  277040  478520  348600  431580  452890  302030  408360  302020  547630  406830  258100  183940  680530  482740  650320  715150  015690  996510  967270  968540  148030  182470  148020  968530  148010  597820  597690  399390  399400  297920  897640  897650  033660  058130  079670  079730  266440  355010  433340  434840  519460  501080  372120  569900  693620  195970  985090  985100  108480  500790  232350  400180  531280  948140  695720  574870  986890  191600  191610  084540  367790  367800  241000  496350  688640  457400  698580  688650  725220  944350  944340  037790  266860  355420  413900  413890  797840  236650  236660  329870  686460  766070  611690  318020  063750  686100  686230  437910  967220  768530  824630  965540  991760  809320  201940  824620  505710  505720  759480  006110  037620  006120  095360  160000  214620  558800  742590  861710  906240  024440  455010  406290  455110  718590  007560  945350  421140  785230  339690  506000  079870  919940  109370  471380  427230  561470  427260  561450  196120  139970  927560  769300  196020  214740  342250  196070  608160  214730  904540  423120  091560  927550  542760  542770  590970  498570  313900  555530  589820  370320  105040  875290  039100  297050  685020  673230  454490  642190  935670  763920  102140  949660  349900  368180  666420  903230  231800  164470  163410  343380  744060  080070  164480  651990  961850  115640  115650  661790  114790  436620  436630  479210  989100  841660  705900  881200  866310  168970  531620  584940  233720  558520  533560  584920  845430  249150  845440  560730  211390  045410  626330  420300  715800  754520  431750  431770  965500  307790  965490  052870  132040  306940  754580  539680  539780  539720  536370  886130  865870  819090  865850  886140  783740  013410  044060  044080  865880  512410  645530  553990  554000  603680  789330  971500  313570  805630  897890  969610  163440  302010  549540  984530  715140  650330  940300  968520  399370  125560  297840  673930  556720  880330  880530  121110  110400  332090  332070  434820  486740  372110  915690  749030  153450  531270  191730  367780  507100  669100  669110  911020  797830  910780  063760  686160  768540  965530  113210  490190  906220  458120  766980  421130  682940  682940R  109270  243870  550890  207590  973990  271910  542750  927290  927310  496940  343370  562390  401280  115630  212270  705890  866350  233700  276660  090750  090740  560720  045390  085450  719650  715780  027290  469790R  469780  364140  469770  469790  469800  585650  608230  586220  633280  632390  013600  044070  013490  044050  130140  780230  847450  561630  685700  120670  744620  533550  090730  019770  563730  589160  722600  084620  626500  420370  777480  420350  420380  211500  211510  746160  045470  084610  308510  626490  159240  211490  362640  405790  405800  544750  544770  681260  684450  626480  420340  845550  103840  211470  342030  362680  572910  196650  084590  845540  188020  229970  247610  305730  405780  540670  553000  808740  808750  918980  521370  018560  103810  103820  131610  209320  211460  323210  437960  556500  766160  959910  085900  086150  873360  736580  560790  029780  050050  076530  103800  119850  285340  315810  405770  430800  455630  683360  683380  719730  726350  043990  044780  049110  833870  833880  880280  880290  880310  880320  052940  159200  211450  317740  327530  450430  540540  944420  596950  845530  866480  084580  084560  308500  159120  308900  365800  405730  405740  405750  405760  408990  906680  917280  590200  159090  293200  456970  469870  489230  564010  571400  587700  616190  683800  779990  866000  885370  078470  895090  005880  086490  093240  093250  103770  103780  103790  117030  191480  232180  293180  293190  296860  354660  365420  366260  385910  444820  456950  489210  489220  571370  571380  571390  616160  616170  616180  683710  683750  683760  712950  729630  729640  779980  825530  855010  865930  969900  969910  087870  092050  308490  103730  103740  103750  103760  157950  208840  295700  295720  296130  297580  298700  301310  301790  303540  312420  318000  318090  318100  323200  335010  339120  354150  374140  383100  392330  392340  420280  420290  420300  420310  420320  420330  422320  426710  431130  431140  433080  445070  450980  457340  519680  556480  556490  560640  587690  648490  683020  683030  689740  726080  766140  766150  815620  825520  887120  898820  938830  945810  103720  159080  260750  295730  301090  329000  417070  418320  444810  583580  570080  514830  308480  560750  012020  044340  066810  103690  103700  103710  148070  177660  260740  293150  306340  307530  317970  323870  328990  329890  366250  397850  427610  430820  432520  444800  451570  489190  489200  571330  571340  571350  571360  587660  587670  587680  588540  616120  682960  766130  779960  779970  825510  870120  885360  898090  941040  761210  420310  308470  044330  103670  103680  107340  156880  312970  323860  328980  353340  384950  385030  385070  385140  385170  385180  385190  385210  385230  385240  385260  385270  385290  385300  385310  385320  385330  427600  439290  444790  456940  469850  469860  479160  563940  587650  588530  736720  766120  781840  825500  962870  103650  103660  147990  298690  323190  417300  431760  433070  683900  766110  825490  005870  159030  159040  159060  683890      UDI:    (01)00880304439962(17)180628(10)882090  (01)00880304439962(17)180728(10)942350  (01)00880304439962(17)180728(10)726180  (01)00880304439962(17)180828(10)111310  (01)00880304439962(17)180828(10)013150  (01)00880304439962(17)180828(10)868010  (01)00880304439962(17)190128(10)623340  (01)00880304439962(17)230128(10)028380  (01)00880304439962(17)230830(10)186120  (01)00880304439962(17)240204(10)409420  (01)00880304439962(17)240207(10)847030  (01)00880304439962(17)240221(10)679990  (01)00880304439962(17)240301(10)755600  (01)00880304439962(17)240505(10)409330  (01)00880304439962(17)240612(10)706270  (01)00880304439962(17)240708(10)169490  (01)00880304439962(17)251222(10)936000  (01)00880304439962(17)251229(10)400990  (01)00880304439962(17)260425(10)343410  (01)00880304439962(17)261121(10)235880  (01)00880304439917(17)180328(10)313330  (01)00880304439917(17)180328(10)324140  (01)00880304439917(17)180701(10)014640R  (01)00880304439917(17)180828(10)884080  (01)00880304439917(17)201228(10)838730  (01)00880304439917(17)230324(10)335900  (01)00880304439917(17)230325(10)684340  (01)00880304439917(17)230418(10)106100  (01)00880304439917(17)230709(10)105780  (01)00880304439917(17)240825(10)140000  (01)00880304439917(17)241126(10)447810  (01)00880304439917(17)250126(10)154700  (01)00880304439917(17)250219(10)911280  (01)00880304439917(17)260207(10)363670  (01)00880304439917(17)260317(10)102500  (01)00880304439917(17)260426(10)931980  (01)00880304439917(17)260518(10)519820  (01)00880304439917(17)260624(10)114830  (01)00880304439955(17)180628(10)634740  (01)00880304439955(17)180701(10)485340  (01)00880304439955(17)180728(10)969290  (01)00880304439955(17)180828(10)108310  (01)00880304439955(17)181028(10)808770  (01)00880304439955(17)200328(10)313520  (01)00880304439955(17)220228(10)698890  (01)00880304439955(17)220528(10)037810  (01)00880304439955(17)230116(10)028450  (01)00880304439955(17)230204(10)304010  (01)00880304439955(17)230213(10)491810  (01)00880304439955(17)230221(10)596490  (01)00880304439955(17)230319(10)684830  (01)00880304439955(17)251221(10)771630  (01)00880304439955(17)251222(10)731090  (01)00880304439955(17)251223(10)384570  (01)00880304439955(17)260208(10)343400  (01)00880304439955(17)260210(10)049790  (01)00880304439955(17)260318(10)057690  (01)00880304439955(17)260322(10)766800  (01)00880304439955(17)260427(10)513670  (01)00880304439955(17)270803(10)114820  (01)00880304439894(17)180601(10)420030R  (01)00880304439894(17)180628(10)808760  (01)00880304439894(17)180728(10)917350  (01)00880304439894(17)180728(10)629400  (01)00880304439894(17)180728(10)808720  (01)00880304439894(17)180728(10)014610  (01)00880304439894(17)230319(10)563690  (01)00880304439894(17)230327(10)671840  (01)00880304439894(17)230415(10)983580  (01)00880304439894(17)230520(10)250250  (01)00880304439894(17)250611(10)190920  (01)00880304439894(17)251016(10)231850  (01)00880304439894(17)251021(10)390870  (01)00880304439894(17)251222(10)757150  (01)00880304439894(17)251228(10)799950  (01)00880304439894(17)260509(10)517900  (01)00880304439894(17)260620(10)204080  (01)00880304439894(17)270305(10)764010  (01)00880304439894(17)270921(10)114810  (01)00880304439894(17)271016(10)249030  (01)00880304439924(17)180228(10)076550  (01)00880304439924(17)180601(10)540650R  (01)00880304439924(17)180828(10)584530  (01)00880304439924(17)191228(10)295500  (01)00880304439924(17)200128(10)296810  (01)00880304439924(17)220228(10)367880  (01)00880304439924(17)230326(10)028440  (01)00880304439924(17)241029(10)620410  (01)00880304439924(17)241117(10)620420  (01)00880304439924(17)241201(10)784160  (01)00880304439924(17)241204(10)046210  (01)00880304439924(17)241208(10)712950  (01)00880304439924(17)250112(10)443870  (01)00880304439924(17)250126(10)642180  (01)00880304439924(17)260523(10)503770  (01)00880304439924(17)280320(10)420330  (01)00880304439900(17)180201(10)917310  (01)00880304439900(17)180728(10)808660  (01)00880304439900(17)201028(10)712490  (01)00880304439900(17)210328(10)842280  (01)00880304439900(17)210428(10)912780  (01)00880304439900(17)220228(10)191740  (01)00880304439900(17)230121(10)824710  (01)00880304439900(17)230220(10)565110  (01)00880304439900(17)230416(10)759500  (01)00880304439900(17)230607(10)279840  (01)00880304439900(17)240715(10)751390  (01)00880304439900(17)241023(10)313950  (01)00880304439900(17)250130(10)297070  (01)00880304439900(17)250313(10)102460  (01)00880304439900(17)260114(10)399850  (01)00880304439900(17)251223(10)728220  (01)00880304439900(17)251229(10)741280  (01)00880304439900(17)260126(10)062670  (01)00880304439900(17)260216(10)599590  (01)00880304439900(17)260315(10)070530  (01)00880304439931(17)180328(10)193570  (01)00880304439931(17)180328(10)227680  (01)00880304439931(17)180701(10)674890  (01)00880304439931(17)180728(10)399620  (01)00880304439931(17)180728(10)485320  (01)00880304439931(17)180728(10)808640  (01)00880304439931(17)180728(10)055710  (01)00880304439931(17)180828(10)055720  (01)00880304439931(17)230215(10)991820  (01)00880304439931(17)230218(10)449680  (01)00880304439931(17)230225(10)192060  (01)00880304439931(17)260726(10)479170  (01)00880304439931(17)260729(10)738870  (01)00880304439931(17)280302(10)626460  (01)00880304439948(17)180628(10)553610  (01)00880304439948(17)240121(10)422710  (01)00880304439948(17)240421(10)751380  (01)00880304439948(17)240825(10)973120  (01)00880304439948(17)250210(10)763930  (01)00880304439948(17)250326(10)233200  (01)00880304439948(17)260524(10)503760  (01)00880304439948(17)260628(10)305760  (01)00880304439948(17)280301(10)211440  (01)00880304204829(17)180228(10)965510  (01)00880304204829(17)180528(10)586290  (01)00880304204829(17)191228(10)785680  (01)00880304204829(17)191228(10)852640  (01)00880304204829(17)200428(10)805470  (01)00880304204829(17)200528(10)134580  (01)00880304204829(17)200628(10)790650  (01)00880304204829(17)200628(10)302140  (01)00880304204829(17)210201(10)547610R  (01)00880304204829(17)210528(10)771130  (01)00880304204829(17)210528(10)372160  (01)00880304204829(17)220528(10)352910  (01)00880304204829(17)230218(10)475630  (01)00880304204829(17)230219(10)768550  (01)00880304204829(17)230219(10)768550R  (01)00880304204829(17)240209(10)465490  (01)00880304204829(17)240319(10)919980  (01)00880304204829(17)240518(10)485950  (01)00880304204829(17)240706(10)139990  (01)00880304204829(17)240721(10)284880  (01)00880304204829(17)240810(10)342280  (01)00880304204829(17)240916(10)266180  (01)00880304204829(17)241030(10)728880  (01)00880304204829(17)241201(10)875320  (01)00880304204829(17)250105(10)541510  (01)00880304204829(17)260216(10)652040  (01)00880304204829(17)260526(10)747580  (01)00880304204829(17)260630(10)331370  (01)00880304204829(17)270305(10)327140  (01)00880304204829(17)270419(10)169040  (01)00880304204829(17)280123(10)866470  (01)00880304204812(17)181201(10)373980  (01)00880304204812(17)190928(10)965160  (01)00880304204812(17)200501(10)906950  (01)00880304204812(17)210528(10)322670  (01)00880304204812(17)210528(10)471660  (01)00880304204812(17)210628(10)687610  (01)00880304204812(17)211128(10)531340  (01)00880304204812(17)230123(10)336150  (01)00880304204812(17)230128(10)946080  (01)00880304204812(17)230204(10)335880  (01)00880304204812(17)230204(10)201960  (01)00880304204812(17)230225(10)684430  (01)00880304204812(17)230228(10)505740  (01)00880304204812(17)230319(10)490180  (01)00880304204812(17)230327(10)759490  (01)00880304204812(17)230822(10)117200  (01)00880304204812(17)230826(10)024480  (01)00880304204812(17)230916(10)252230  (01)00880304204812(17)231011(10)406320  (01)00880304204812(17)240120(10)132610  (01)00880304204812(17)240120(10)143430  (01)00880304204812(17)240201(10)409410  (01)00880304204812(17)240221(10)925400  (01)00880304204812(17)240226(10)682980  (01)00880304204812(17)240227(10)435670  (01)00880304204812(17)240507(10)561500  (01)00880304204812(17)240706(10)342270  (01)00880304204812(17)240717(10)271940  (01)00880304204812(17)240806(10)827340  (01)00880304204812(17)240825(10)139770  (01)00880304204812(17)240828(10)161980  (01)00880304204812(17)240904(10)453360  (01)00880304204812(17)240905(10)542830  (01)00880304204812(17)241013(10)522760  (01)00880304204812(17)241203(10)875310  (01)00880304204812(17)241209(10)039180  (01)00880304204812(17)250415(10)473210  (01)00880304204812(17)250426(10)781530  (01)00880304204812(17)250430(10)610870  (01)00880304204812(17)250430(10)294890  (01)00880304204812(17)250519(10)947310  (01)00880304204812(17)250526(10)070400  (01)00880304204812(17)250527(10)069880  (01)00880304204812(17)250603(10)026340  (01)00880304204812(17)250618(10)372930  (01)00880304204812(17)250717(10)430830  (01)00880304204812(17)251006(10)819690  (01)00880304204812(17)251029(10)399840  (01)00880304204812(17)251106(10)496920  (01)00880304204812(17)251222(10)886220  (01)00880304204812(17)260108(10)980410  (01)00880304204812(17)260120(10)416120  (01)00880304204812(17)260125(10)363660  (01)00880304204812(17)260223(10)164590  (01)00880304204812(17)260324(10)821940  (01)00880304204812(17)260421(10)115710  (01)00880304204812(17)260425(10)503750  (01)00880304204812(17)260620(10)212310  (01)00880304204812(17)261107(10)462060  (01)00880304204812(17)261115(10)120690  (01)00880304204812(17)261215(10)881240  (01)00880304204812(17)270114(10)881180  (01)00880304204812(17)270305(10)079230  (01)00880304204812(17)270315(10)169030  (01)00880304204812(17)270319(10)079220  (01)00880304204812(17)270422(10)531660  (01)00880304204812(17)270825(10)327130  (01)00880304204812(17)271010(10)090650  (01)00880304204812(17)280108(10)211420  (01)00880304204812(17)280112(10)845510  (01)00880304204812(17)280117(10)866440  (01)00880304204812(17)280220(10)090660  (01)00880304204812(17)280305(10)045460  (01)00880304204812(17)280327(10)819530  (01)00880304204805(17)201228(10)680500  (01)00880304204805(17)220101(10)367830  (01)00880304204805(17)220828(10)329900  (01)00880304204805(17)221228(10)824680  (01)00880304204805(17)230101(10)991790  (01)00880304204805(17)230117(10)113230  (01)00880304204805(17)230128(10)201950  (01)00880304204805(17)230220(10)335870  (01)00880304204805(17)230320(10)671830  (01)00880304204805(17)230415(10)006130  (01)00880304204805(17)230508(10)085250  (01)00880304204805(17)230508(10)160010  (01)00880304204805(17)230606(10)359270  (01)00880304204805(17)230607(10)422910  (01)00880304204805(17)230607(10)423070  (01)00880304204805(17)230710(10)670240  (01)00880304204805(17)230824(10)970390  (01)00880304204805(17)230829(10)057550  (01)00880304204805(17)230829(10)906250  (01)00880304204805(17)230916(10)231030  (01)00880304204805(17)230930(10)455030  (01)00880304204805(17)230930(10)455360  (01)00880304204805(17)231011(10)266250  (01)00880304204805(17)231011(10)406310  (01)00880304204805(17)231018(10)162310  (01)00880304204805(17)231023(10)649440  (01)00880304204805(17)231109(10)649430  (01)00880304204805(17)231113(10)828560  (01)00880304204805(17)231124(10)766990  (01)00880304204805(17)231209(10)870250  (01)00880304204805(17)240109(10)945370  (01)00880304204805(17)240120(10)143410  (01)00880304204805(17)240121(10)143420  (01)00880304204805(17)240204(10)410710  (01)00880304204805(17)240206(10)421170  (01)00880304204805(17)240222(10)880810  (01)00880304204805(17)240226(10)506030  (01)00880304204805(17)240305(10)682970  (01)00880304204805(17)240316(10)919970  (01)00880304204805(17)240326(10)785240  (01)00880304204805(17)240413(10)551280  (01)00880304204805(17)240414(10)119550  (01)00880304204805(17)240422(10)751370  (01)00880304204805(17)240520(10)561490  (01)00880304204805(17)240522(10)485930  (01)00880304204805(17)240528(10)003950  (01)00880304204805(17)240604(10)409830  (01)00880304204805(17)240613(10)243880  (01)00880304204805(17)240618(10)649340  (01)00880304204805(17)240702(10)271930  (01)00880304204805(17)240706(10)139980  (01)00880304204805(17)240714(10)522000  (01)00880304204805(17)240806(10)453340  (01)00880304204805(17)240806(10)542810  (01)00880304204805(17)240813(10)542820  (01)00880304204805(17)240825(10)668240  (01)00880304204805(17)240825(10)453350  (01)00880304204805(17)240925(10)126380  (01)00880304204805(17)241008(10)589770  (01)00880304204805(17)241013(10)313940  (01)00880304204805(17)241023(10)313930  (01)00880304204805(17)241028(10)555550  (01)00880304204805(17)241029(10)712940  (01)00880304204805(17)241121(10)875300  (01)00880304204805(17)241211(10)039170  (01)00880304204805(17)241229(10)214050  (01)00880304204805(17)250124(10)154780  (01)00880304204805(17)250304(10)673520  (01)00880304204805(17)250306(10)685010  (01)00880304204805(17)250310(10)935690  (01)00880304204805(17)250331(10)294870  (01)00880304204805(17)250402(10)935680  (01)00880304204805(17)250403(10)411780  (01)00880304204805(17)250408(10)452970  (01)00880304204805(17)250421(10)580230  (01)00880304204805(17)250526(10)940620  (01)00880304204805(17)250625(10)299340  (01)00880304204805(17)250810(10)584000  (01)00880304204805(17)250904(10)819680  (01)00880304204805(17)250916(10)053300  (01)00880304204805(17)250916(10)819670  (01)00880304204805(17)251014(10)289370  (01)00880304204805(17)251103(10)496800  (01)00880304204805(17)251104(10)496770  (01)00880304204805(17)251218(10)496780  (01)00880304204805(17)251229(10)399830  (01)00880304204805(17)260115(10)095080  (01)00880304204805(17)260125(10)164510  (01)00880304204805(17)260126(10)095090  (01)00880304204805(17)260208(10)744250  (01)00880304204805(17)260208(10)652030  (01)00880304204805(17)260218(10)685720  (01)00880304204805(17)260322(10)095070  (01)00880304204805(17)260323(10)115700  (01)00880304204805(17)260328(10)115670  (01)00880304204805(17)260329(10)171340  (01)00880304204805(17)260419(10)115690  (01)00880304204805(17)260420(10)503740  (01)00880304204805(17)260421(10)115680  (01)00880304204805(17)260509(10)893580  (01)00880304204805(17)260518(10)747570  (01)00880304204805(17)260620(10)212300  (01)00880304204805(17)260715(10)436640  (01)00880304204805(17)260818(10)841670  (01)00880304204805(17)261125(10)705920  (01)00880304204805(17)",
      "center_classification_date": "20191122",
      "distribution_pattern": "US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.",
      "state": "IN",
      "product_description": "Copeland HA Resurfacing Head, MB/HA EAS Head; Item Nos. 11-114641  11-114642  11-114643  11-114644  11-114644S  11-114645  11-114646  11-114647  11-114648  11-114661  11-114662  11-114663  11-114664  11-114664S  11-114665  11-114666  11-114667  11-114668  11-114632  11-114633  Product Usage: Partial Shoulder Replacement.",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0527-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83783",
      "termination_date": "20210721",
      "more_code_info": "270314(10)866330  (01)00880304204805(17)270315(10)866320  (01)00880304204805(17)270316(10)169020  (01)00880304204805(17)270320(10)247060  (01)00880304204805(17)270407(10)547110  (01)00880304204805(17)270415(10)595380  (01)00880304204805(17)270417(10)595410  (01)00880304204805(17)270507(10)233730  (01)00880304204805(17)270524(10)531650  (01)00880304204805(17)270623(10)169000  (01)00880304204805(17)270817(10)595390  (01)00880304204805(17)270925(10)845490  (01)00880304204805(17)271022(10)090630  (01)00880304204805(17)271124(10)845500  (01)00880304204805(17)271218(10)090620  (01)00880304204805(17)280109(10)881170  (01)00880304204805(17)280226(10)866410  (01)00880304204805(17)280227(10)683040  (01)00880304204805(17)280228(10)090640  (01)00880304204805(17)280228(10)626430  (01)00880304204805(17)280305(10)211410  (01)00880304204805(17)280416(10)866400  (01)00880304204805(17)280416(10)626390  (01)00880304204805(17)280506(10)560760  (01)00880304204799(17)180401(10)540070  (01)00880304204799(17)180428(10)586240  (01)00880304204799(17)180428(10)586250  (01)00880304204799(17)180528(10)605810  (01)00880304204799(17)191128(10)656110  (01)00880304204799(17)191228(10)790070  (01)00880304204799(17)200328(10)374040  (01)00880304204799(17)200628(10)277070  (01)00880304204799(17)210428(10)673940  (01)00880304204799(17)210528(10)566540  (01)00880304204799(17)230219(10)335860  (01)00880304204799(17)230816(10)024470  (01)00880304204799(17)240225(10)682960  (01)00880304204799(17)240316(10)421160  (01)00880304204799(17)240421(10)485920  (01)00880304204799(17)240601(10)729180  (01)00880304204799(17)240606(10)925510  (01)00880304204799(17)240608(10)485940  (01)00880304204799(17)240706(10)424440  (01)00880304204799(17)240710(10)353450  (01)00880304204799(17)240714(10)000130  (01)00880304204799(17)240716(10)522010  (01)00880304204799(17)240818(10)542800  (01)00880304204799(17)240904(10)927390  (01)00880304204799(17)240917(10)268300  (01)00880304204799(17)241017(10)555540  (01)00880304204799(17)241202(10)039150  (01)00880304204799(17)250312(10)552140  (01)00880304204799(17)260210(10)164570  (01)00880304204799(17)260329(10)685710  (01)00880304204799(17)260819(10)989050  (01)00880304204799(17)270523(10)533590  (01)00880304204799(17)270718(10)531640  (01)00880304204799(17)280313(10)866390  (01)00880304204782(17)200628(10)302050  (01)00880304204782(17)210628(10)553980  (01)00880304204782(17)211101(10)574880  (01)00880304204782(17)220101(10)453810  (01)00880304204782(17)220528(10)498400  (01)00880304204782(17)230108(10)824640  (01)00880304204782(17)230123(10)321670  (01)00880304204782(17)230128(10)991780  (01)00880304204782(17)230207(10)334820  (01)00880304204782(17)230218(10)476440  (01)00880304204782(17)230218(10)490200  (01)00880304204782(17)230220(10)113220  (01)00880304204782(17)230220(10)511220  (01)00880304204782(17)230225(10)696970  (01)00880304204782(17)230320(10)671820  (01)00880304204782(17)230808(10)847280  (01)00880304204782(17)230820(10)024460  (01)00880304204782(17)230822(10)013420  (01)00880304204782(17)230823(10)175540  (01)00880304204782(17)230930(10)429810  (01)00880304204782(17)231011(10)454110  (01)00880304204782(17)231026(10)649410  (01)00880304204782(17)231104(10)649420  (01)00880304204782(17)231119(10)649390  (01)00880304204782(17)231121(10)852280  (01)00880304204782(17)231204(10)933990  (01)00880304204782(17)231204(10)132600  (01)00880304204782(17)240107(10)945360  (01)00880304204782(17)240113(10)143400  (01)00880304204782(17)240114(10)503920  (01)00880304204782(17)240211(10)745440  (01)00880304204782(17)240213(10)374190  (01)00880304204782(17)240213(10)506020  (01)00880304204782(17)240218(10)435660  (01)00880304204782(17)240228(10)682950  (01)00880304204782(17)240229(10)421150  (01)00880304204782(17)240304(10)919950  (01)00880304204782(17)240304(10)919960  (01)00880304204782(17)240310(10)254820  (01)00880304204782(17)240327(10)843980  (01)00880304204782(17)240331(10)379540  (01)00880304204782(17)240401(10)396960  (01)00880304204782(17)240413(10)368060  (01)00880304204782(17)240521(10)550900  (01)00880304204782(17)240604(10)082420  (01)00880304204782(17)240606(10)975920  (01)00880304204782(17)240610(10)927580  (01)00880304204782(17)240611(10)927590  (01)00880304204782(17)240616(10)214760  (01)00880304204782(17)240617(10)271920  (01)00880304204782(17)240619(10)347570  (01)00880304204782(17)240623(10)284940  (01)00880304204782(17)240706(10)214750  (01)00880304204782(17)240714(10)522170  (01)00880304204782(17)240714(10)563850  (01)00880304204782(17)240714(10)594840  (01)00880304204782(17)240714(10)403940  (01)00880304204782(17)240714(10)342260  (01)00880304204782(17)240806(10)681280  (01)00880304204782(17)240806(10)542790  (01)00880304204782(17)240807(10)453330  (01)00880304204782(17)240807(10)542780  (01)00880304204782(17)240811(10)815540  (01)00880304204782(17)240812(10)982470  (01)00880304204782(17)240902(10)927350  (01)00880304204782(17)240903(10)268150  (01)00880304204782(17)240904(10)927320  (01)00880304204782(17)240917(10)927570  (01)00880304204782(17)241007(10)498820  (01)00880304204782(17)241205(10)039140  (01)00880304204782(17)241229(10)297060  (01)00880304204782(17)250401(10)151700  (01)00880304204782(17)250407(10)294860  (01)00880304204782(17)250407(10)452960  (01)00880304204782(17)250512(10)845700  (01)00880304204782(17)250527(10)940610  (01)00880304204782(17)250807(10)583990  (01)00880304204782(17)250824(10)894920  (01)00880304204782(17)250910(10)050620  (01)00880304204782(17)251014(10)247840  (01)00880304204782(17)251024(10)981160  (01)00880304204782(17)251103(10)496810  (01)00880304204782(17)251216(10)825210  (01)00880304204782(17)251217(10)399810  (01)00880304204782(17)251222(10)416110  (01)00880304204782(17)260113(10)164490  (01)00880304204782(17)260126(10)741270  (01)00880304204782(17)260209(10)095060  (01)00880304204782(17)260218(10)769680  (01)00880304204782(17)260222(10)164500  (01)00880304204782(17)260226(10)652010  (01)00880304204782(17)260309(10)060010  (01)00880304204782(17)260426(10)548410  (01)00880304204782(17)260426(10)523030  (01)00880304204782(17)260426(10)115660  (01)00880304204782(17)260426(10)961870  (01)00880304204782(17)260620(10)212280  (01)00880304204782(17)260630(10)479190  (01)00880304204782(17)261118(10)705910  (01)00880304204782(17)261121(10)120680  (01)00880304204782(17)261214(10)881220  (01)00880304204782(17)270316(10)168980  (01)00880304204782(17)270319(10)079190  (01)00880304204782(17)270320(10)866360  (01)00880304204782(17)270617(10)531630  (01)00880304204782(17)270620(10)533580  (01)00880304204782(17)270923(10)845450  (01)00880304204782(17)270925(10)090570  (01)00880304204782(17)271001(10)595340  (01)00880304204782(17)271004(10)845480  (01)00880304204782(17)271023(10)090580  (01)00880304204782(17)271103(10)090590  (01)00880304204782(17)271122(10)090610  (01)00880304204782(17)280122(10)045440  (01)00880304204782(17)280126(10)045450  (01)00880304204782(17)280220(10)626360  (01)00880304204782(17)280301(10)626380  (01)00880304204782(17)280305(10)687850  (01)00880304204782(17)280422(10)866380  (01)00880304204775(17)180328(10)304550  (01)00880304204775(17)200628(10)302030  (01)00880304204775(17)210728(10)985090  (01)00880304204775(17)221228(10)965540  (01)00880304204775(17)230121(10)991760  (01)00880304204775(17)230123(10)809320  (01)00880304204775(17)230204(10)201940  (01)00880304204775(17)230204(10)824620  (01)00880304204775(17)230227(10)505710  (01)00880304204775(17)230227(10)505720  (01)00880304204775(17)230320(10)759480  (01)00880304204775(17)230410(10)006110  (01)00880304204775(17)230410(10)037620  (01)00880304204775(17)230412(10)006120  (01)00880304204775(17)230502(10)095360  (01)00880304204775(17)230503(10)160000  (01)00880304204775(17)230628(10)214620  (01)00880304204775(17)230701(10)558800  (01)00880304204775(17)230722(10)742590  (01)00880304204775(17)230731(10)861710  (01)00880304204775(17)230829(10)906240  (01)00880304204775(17)230830(10)024440  (01)00880304204775(17)230930(10)455010  (01)00880304204775(17)231011(10)406290  (01)00880304204775(17)231018(10)455110  (01)00880304204775(17)231113(10)718590  (01)00880304204775(17)231118(10)007560  (01)00880304204775(17)240113(10)945350  (01)00880304204775(17)240201(10)421140  (01)00880304204775(17)240221(10)785230  (01)00880304204775(17)240226(10)339690  (01)00880304204775(17)240226(10)506000  (01)00880304204775(17)240303(10)079870  (01)00880304204775(17)240310(10)919940  (01)00880304204775(17)240316(10)109370  (01)00880304204775(17)240406(10)471380  (01)00880304204775(17)240506(10)427230  (01)00880304204775(17)240506(10)561470  (01)00880304204775(17)240506(10)427260  (01)00880304204775(17)240601(10)561450  (01)00880304204775(17)240606(10)196120  (01)00880304204775(17)240611(10)139970  (01)00880304204775(17)240615(10)927560  (01)00880304204775(17)240616(10)769300  (01)00880304204775(17)240616(10)196020  (01)00880304204775(17)240702(10)214740  (01)00880304204775(17)240709(10)342250  (01)00880304204775(17)240714(10)196070  (01)00880304204775(17)240728(10)608160  (01)00880304204775(17)240806(10)214730  (01)00880304204775(17)240819(10)904540  (01)00880304204775(17)240819(10)423120  (01)00880304204775(17)240823(10)091560  (01)00880304204775(17)240829(10)927550  (01)00880304204775(17)240829(10)542760  (01)00880304204775(17)240905(10)542770  (01)00880304204775(17)241013(10)590970  (01)00880304204775(17)241013(10)498570  (01)00880304204775(17)241016(10)313900  (01)00880304204775(17)241017(10)555530  (01)00880304204775(17)241121(10)589820  (01)00880304204775(17)250203(10)370320  (01)00880304204775(17)241203(10)105040  (01)00880304204775(17)241208(10)875290  (01)00880304204775(17)241211(10)039100  (01)00880304204775(17)250112(10)297050  (01)00880304204775(17)250115(10)685020  (01)00880304204775(17)250129(10)673230  (01)00880304204775(17)250206(10)454490  (01)00880304204775(17)250219(10)642190  (01)00880304204775(17)250220(10)935670  (01)00880304204775(17)250224(10)763920  (01)00880304204775(17)250310(10)102140  (01)00880304204775(17)250310(10)949660  (01)00880304204775(17)250331(10)349900  (01)00880304204775(17)250407(10)368180  (01)00880304204775(17)250430(10)666420  (01)00880304204775(17)250430(10)903230  (01)00880304204775(17)251026(10)231800  (01)00880304204775(17)260111(10)164470  (01)00880304204775(17)260115(10)163410  (01)00880304204775(17)260118(10)343380  (01)00880304204775(17)260208(10)744060  (01)00880304204775(17)260209(10)080070  (01)00880304204775(17)260209(10)164480  (01)00880304204775(17)260209(10)651990  (01)00880304204775(17)260324(10)961850  (01)00880304204775(17)260406(10)115640  (01)00880304204775(17)260425(10)115650  (01)00880304204775(17)260428(10)661790  (01)00880304204775(17)260628(10)114790  (01)00880304204775(17)260628(10)436620  (01)00880304204775(17)260629(10)436630  (01)00880304204775(17)260815(10)479210  (01)00880304204775(17)260816(10)989100  (01)00880304204775(17)260907(10)841660  (01)00880304204775(17)261120(10)705900  (01)00880304204775(17)261215(10)881200  (01)00880304204775(17)270315(10)866310  (01)00880304204775(17)270324(10)168970  (01)00880304204775(17)270424(10)531620  (01)00880304204775(17)270520(10)584940  (01)00880304204775(17)270624(10)233720  (01)00880304204775(17)270803(10)558520  (01)00880304204775(17)270914(10)533560  (01)00880304204775(17)270920(10)584920  (01)00880304204775(17)271016(10)845430  (01)00880304204775(17)271030(10)249150  (01)00880304204775(17)271218(10)845440  (01)00880304204775(17)280213(10)560730  (01)00880304204775(17)280318(10)211390  (01)00880304204775(17)280418(10)045410  (01)00880304204775(17)280512(10)626330  (01)00880304204775(17)290104(10)420300  (01)00880304204768(17)180228(10)431770  (01)00880304204768(17)180328(10)754580  (01)00880304204768(17)200528(10)897890  (01)00880304204768(17)201128(10)650330  (01)00880304204768(17)210328(10)880330  (01)00880304204768(17)210528(10)372110  (01)00880304204768(17)210828(10)749030  (01)00880304204768(17)220228(10)669110  (01)00880304204768(17)221001(10)063760  (01)00880304204768(17)230115(10)965530  (01)00880304204768(17)230117(10)113210  (01)00880304204768(17)230219(10)490190  (01)00880304204768(17)230830(10)906220  (01)00880304204768(17)230930(10)458120  (01)00880304204768(17)231203(10)766980  (01)00880304204768(17)240130(10)421130  (01)00880304204768(17)240225(10)682940  (01)00880304204768(17)240225(10)682940R  (01)00880304204768(17)240316(10)109270  (01)00880304204768(17)240328(10)243870  (01)00880304204768(17)240518(10)550890  (01)00880304204768(17)240608(10)207590  (01)00880304204768(17)240616(10)973990  (01)00880304204768(17)240616(10)271910  (01)00880304204768(17)240813(10)542750  (01)00880304204768(17)240819(10)927290  (01)00880304204768(17)240918(10)927310  (01)00880304204768(17)251026(10)496940  (01)00880304204768(17)260209(10)343370  (01)00880304204768(17)260125(10)562390  (01)00880304204768(17)260126(10)401280  (01)00880304204768(17)260419(10)115630  (01)00880304204768(17)260705(10)212270  (01)00880304204768(17)261206(10)705890  (01)00880304204768(17)270319(10)866350  (01)00880304204768(17)270717(10)233700  (01)00880304204768(17)270921(10)276660  (01)00880304204768(17)271003(10)090750  (01)00880304204768(17)271005(10)090740  (01)00880304204768(17)280226(10)560720  (01)00880304204768(17)280319(10)045390  (01)00880304204768(17)280510(10)085450  (01)00880304204751(17)180128(10)719650  (01)00880304204751(17)180228(10)715780  (01)00880304204751(17)180328(10)027290  (01)00880304204751(17)180401(10)469790R  (01)00880304204751(17)180428(10)364140  (01)00880304204751(17)180428(10)469770  (01)00880304204751(17)180428(10)469790  (01)00880304204751(17)180428(10)469800  (01)00880304204751(17)180528(10)586220  (01)00880304204751(17)180628(10)632390  (01)00880304204751(17)180828(10)013490  (01)00880304204751(17)180828(10)044050  (01)00880304204751(17)250219(10)847450  (01)00880304204751(17)251218(10)561630  (01)00880304204751(17)260222(10)685700  (01)00880304204751(17)261206(10)120670  (01)00880304204751(17)270413(10)744620  (01)00880304204751(17)270414(10)533550  (01)00880304204751(17)271009(10)090730  (01)00887868346934(17)281113(10)019770",
      "recall_initiation_date": "20190821",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}