{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.",
      "address_2": "",
      "product_quantity": "219988 units (total)",
      "code_info": "All lots manufactured from January 2008 - May 2019    458330  805800  098150  098170  551700  582720  484570  885530  027260  549910  549930  549920  860410  057750  285620  535000  218160  132060  324940  047780  004120  004130  842140  193730  193600  842160  256170  256190  842150  382410  087590  584580  805780  904960  098160  098180  941910  241310  298790  300880  300890  408260  783420  848300  899690  528130  419980  448560  465460  465440  582640  669460  589030  817000  179900  788490  092880  453280  827050  827070  149670  090400  507780  522500  119230  197000  838570  103090  694060  047390  047400  047410  588340  683450  683480  037200  003870  869160  009630  209340  130810  399780  442740  425020  494290  494300  053330  785350  302010  545270  780360  780400  977760  970270  829750  222760  430770  671900  928740  860460  057730  203030  203040  427270  520000  829600  063830  064580  006930  991750  487720  227890  224470  626900  464300  964110  983020  184550  239200  783270  783290  403570  364850  054110  054160  960050  009820  438200  401880  408150  781210  763830  763790  544580  369920  537340  811220  811320  963080  958520  126020  197000  195230  489320  489590  489830  490700  490840  503260  503710  398840  677530  099100  501430  544030  725100  997570  997560  975970  418720  481160  280200  122050  276260  278030  595380  601710  601700  601730  595820  600810  636330  648400  600600  816490  482850  281380  577580  684070  051290  015910  233430  233270  359560  566560  445750  172350  030870  247430  597870  332340  331070  490540  920450  224420  251940  191060  547260  444650  080060  251170  390910  425180  341320  993180  390900  806420  110870  341340  121770  597460  977930  525350  868480  607660  464650  474060  443870  443160  826490  186970  589550  567500  568630  570100  487210  570690  506800  589820  721950  864570  726280  022340  600430  276040  890550  950370  186030  419000  187110  589260  236170  664770  159400  887090  167020  193000  071070  936700  605110  301000  468560  539170  013780  943830  584750  156710  697680  159500  986190  196320  332830  766240  207810  125050  235260  919740  757020  789410  528640  868980  645490  114740  130640  385520  216810  435600  486190  628720  370630  881420  256320  680170  057750  057730  680700  557830  423080  737810  056070  451490  532920  811110  581310  830450  584440  584430  162100  656120  332880  845140  421470  907980  737020  577210  865520  466780  275720  264380  541890  042360  150630  920760  545720  412190  694920  311430  402680  517200  550180  039790  063370  195230  997570  197000  823130  919710  105930  766960  622310  162390  598200  929000  612360  767390  806980  810230  554930  690100  287710  275620  420300  530260  660360  117440  292130  185920  292140  007740  712160  072110  784330  217050  809270  934930  837160  176270  529630  209030  020940  443320  468180  562130  463650  112000  544860  135220  612300  529580  401010  207310  382210  295690  615910  801240  384350  944240  328460  517460  939390  539600  733990  589930  932260  368050  711170  555310  339730  322770  403790  438780  376430  233310  889980  875070  414760  050470  568660  487550  001770  687820  165760  454530  110500  418750  334620  968360  131800  354670  278110  850570  618540  924720  695890  339610  938860  425020  137250  618690  925090  578700  321680  644290  408070  389150  891160  956430  603020  938090  978470  525860  909040  518060  276590  394320  291120  387410  966710  774530  074050  011200  556010  439900  093010  396670  131760  650240  747740  244810  666770  582360  033660  730550  137120  170300  822110  506410  177790  537650  042730  734550  991780  991760  576770  304580  645390  241310  457660  941750  532110  720320  109760  373390  879710  902470  919320  458790  930850  943880  255810  211580  537220  538610  281870  191850  798890  275460  274450  533350  272760  606240  401780  437560  152770  047250  088530  889870  822120  829750  430770  108430  122290  798870  798890  833440  558550  805800  382410  721040  420210  081280  937130  928170  439040  062360  205290  841780  955590  211790  717680  605400  123190  682900  897960  421560  650230  829600  130450  191230  181590  591130  700740  939060  822250  490720  659250  244660  150390  040310  134510  315890  374770  130760  287200  736180  825570  147110  266970  267000  512540  262940  363110  677760  996770  897650  116430  760430  762840  779480  196960  011170  378840  834140  246700  651520  559910  015290  920050  869970  479220  205310  442740  336130  507780  494300  494290  705110  173380  760180  077170  202310  750010  305500  944080  506680  235010  601270  653390  653380  070290  843970  873140  072740  242860  533360  676870  138470  622070  930110  961370  651380  484190  373430  443740  157100  703760  073790  605100  588610  836330  729530  667240  160530  719500  525900  653720  727120  316550  517650  421570  036600  286550  872930  286080  198320  738980  742200  811230  194530  586790  909350  884490  129270  722950  589610  670560  167600  950390  846930  273610  883440  946230  919080  017120  370170  049800  328010  216230  281420  571610  569280  526920  151740  370280  241640  786520  108390  638920  557020  952620  389180  902980  732250  635780  161640  375690  799700  525940  745480  386200  283280  285620  535000  582720  484570  027260  218160  458330  549930  549920  549910  132060  324940  930740  620300  668770  455300  332010  887060  734450  937270  114500  734480  170910  348950  834050  780870  675490  613890  699400  149940  156070  579190  122100  574740  006930  683480  626900  683450  464300  300890  300880  298790  037200  869290  878000  238180  246960  397420  788490  328050  003870  004130  584580  780400  780360  130810  399780  004120  877420  033770  007160  385020  408830  977760  159760  799840  687150  008820  260200  258920  159790  621700  269670  810210  634170  508820  940830  225720  397430  237190  179540  544250  753730  501080  527490  095490  333780  984840  404080  312780  179770  662570  522510  664400  142950  275880  260090  816030  374430  115360  109830  888690  178630  094650  129830  952590  771690  326450  098640  615650  414870  400370  986350  077180  728730  920800  921860  342330  429750  723620  506520  561110  929100  589950  741930  209350  209370  573460  099590  699240  618930  305470  639710  232930  831590  925250  946730  116440  597290  749240  569430  297250  748550  214490  673820  084110  699300  092170  921610  764020  595260  947750  990450  041270  417450  323120  795930  419480  921900  243500  316210  605120  973010  605180  270400  518370  733980  692970  125290  529300  123540  064550  376930  433010  932610  059590  130200  536290  590880  025100  571090  161760  435060  665920  926030  161870  473900  699860  070460  091600  845150  048170  806980  171930  084780  084800  968810  836780  058140  138580  708210  300230  023100  508060  518910  837960  095540  271590  005400  052840  925710  544110  932730  613030  505180  728920  475630  496780  781170  038760  997350  889540  725110  598210  527270  014860  126460  771190  206450  892830  499480  137800  883970  458730  528540  147100  402940  730730  505410  342380  069230  936450  907970  145390  848700  242850  078580  035560  035570  515780  515790  534580  515800  534570  730180  469150  969830  379800  454850  511120  523700  791960  115210  596200  995500  070640  345170  139040  321200  287360  345090  554820  300060  554550  656090  685380  035580  304560  461000  296390  509780  738050  714100  274320  205290  642420  065520  735890  071920  785130  501660  786380  379810  539650  873100  326980  417080  646120  620310  756540  232000  383150  364500  384310  506770  383160  567570  922270  138950  138960  190150  997290  120380  396530  291070  978020  176050  805760  383420  901190  042880  799710  808050  808050R  846060  187070  298310  805550  389120  260930R  216030  260930  498380  788910  277780  745290  782130  287350  760580  061970  274260  376160  797700  614300  910060  335240  194280  136900  173900  756220  254490  765720  131720  264670  656110  685410  087600  703570  290550  746550  461060  962150  740870  183180  668610  858920  573800  303910  110730  596500  011030  778370  194290  811300  955550  107080  816970  812050  827130  283180  781370  969510  264650  344010  514630  782110  782120  816010  909460  087590  339680  969350  340260  580250  899330  629280  827400  880470  368150  708980  803600  928780  805590  326740  863540  217940  983130R  422880  983130  437720  574730  809110  066880  768350  006110  787090  236430  417110  967140  969640  620350  364540  619080  179390  273120  131170  620320  684910  684920  706410  670240  312210  312250  291290  291340  781360  536360  264630  344000  450630  514620  450620  685390  685400  692890  692880  017780  098380  133900  098720  703550  907670  760570  150610  605520  296210  174440  689930  079790  208520  233310  835760  772950  618500  137940  341380  450880  714460  036720  020760  113540  205400  715520  260250  064750  637300  515060  406910  406920  614280  005400  418390  207430  733400  571350  571360  289970  858940  813760  156320  911240  101230  529780  809100  053120  918880  570950  406190  798630  136890  587290  811290  417100  050510  488670  967160  894410  811470  017390  620340  995510  800260  345210  724930  232270  389590  816960  922310  139050  139060  827100  704410  287370  855870  257830  468720  366760  221040  291250  440150  554940  085280  459190  211940  264620  343990  450610  529680  710030  710020  871090  871100  999540  339650  339640  441850  041420  336690  377590  703540  907660  990490  990500  320200  828380  759040  759010  660000  962130  509790  553080  618570  822890  740860  017890  085800  204390  355380  368140  632660  657280  714230  570820  928900  888200  641150  164590  435850  759420  818680  699240  642440  374780  655070  794210  168360  255490  064740  198650  370060  776090  614260  913400  413580  117570  437710  273810  066890  427250  475130  186750  595630  459600  674420  744730  037170  454620  175780  742470  570940  790240  494000  007650  741150  459660  136880  562600  053830  115220  488800  256560  863240  863240R  488660  685170  132680  488650  764820  764830  017380  006850  691220  506090  364530  567580  611310  345200  765710  816950  138930  138940  228850  527310  827120  724610  057740  483300  706400  243220  885080  885070  028590  291170  291240  554920  554930  542760  313320  678710  974090  974100  310470  384380  450600  971960  554560  637030  637040  656100  710010  994560  335180  139270  472700  703520  469140  732730  934310  990480  990470  344890  374000  721330  721320  399080  399100  242170  399090  525500  962120  640690  801520  790170  738300  085910  054680  179560  236540  506160  367650  945250  709030  709040  641090  925260  957460  910650  435830  901560  385170  642430  048390  065540  867030  065380  415780  867040  427260  427280  971470  406900  657610  991860  746450  761460  657900  657890  983120  417460  565990  504530  354220  784890  236420R  784910  110720  274030  006100  292140  784950  236420  450780  784880  491580  034030  725930  503470  201100  139670  389010  139550  798620  384390  943170  136870  873200  526710  136860  895710  408470  133760  132670  405560  646130  764810  026950  002860  017370  588600  620330  969730  296400  454540  364520  510220  383180  705750  551420  611300  940380  585990  032320  800220  099250  439010  922300  139000  119880  139020  139030  139010  131160  306930  228830  228840  546440  753900  963240  483290  981360  168700  150800  631840  291150  291110  554900  554890  929630  974070  974080  514610  842830  971940  971950  113240  287340  483190  304570  989380  098710  469180  607540  703510  021550  854330  990460  397660  721310  731740  027770  219420  525490  730630  829350  652210  232230  553070  629290  120020  570800  765640  367920  532900  709020  570560  972120  059880  142780  673680  617210  867010  805610  001030  705990  515050  339030  450080  789700  746440  099000  746430  610420  117540  566000  327230  417440  469440  784870  388190  273800  236410  627910  340640  596960  873190  079500  417090  619740  078350  847120  333600  758390  405550  846960  020390  784900  002850  279750  435310  811520  962770  444690  602190  620320  955540  955530  045920  766610  364510  628930  800250  345190  765700  055580  026050  232120  816940  922290  138910  138900  138920  753710  753730  225790  225780  670220  969230  258520  187010  113150  243200  028560  028550  440130  554870  667070  291100  554880  781290  504450  459210  315010  974050  974060  343960  343970  514600  745400  745230  761230  709980  709990  799700  999530  113230  551910  760560  955210  131960  532100  553230  769860  174430  659990  927190  860930  553060  484830  738290  835610  179340  981380  444410  368090  714310  641040  164580  448170  867000  065510  376140  910040  486060  614240  913390  733390  577800  878090  688070  469400  091650  469330  460470  225000  473070  345750  176160  429790  661430  778360  790120  570930  136850  873110  488790  654730  060660  404940  488640  178480  178510  811460  894400  596210  171360  506080  829310  383170  296390  454530  628920  345180  656590  724860  026040  922280  138990  138980  138970  753890  357370  225770  670210  116830  300060  052220  631830  269770  028540  440080  291090  554850  781210  781270  085270  459250  314980  209340  123110  976060  186690  863930  937040  505080  902760  158540  893790  891800  354210  580730  833190  245070  987770  101010  023240  752240  061630  464490  925810  931330  912610  143330  767860  040610  665130  457180  108410  010560  845100  385010  778320  778330  589440  242260  081730  077160  152510  333180  232750  018170  729400  922610  164870  012210  287220  113270  299760  494530  615350  663460  795130  866540  445600  987060  703660  570990  725610  865800  410790  323190  321240  062750  323220  392890  195790  678570  589430  549110  327840  527260  737150  811900  781450  185340  143070  102630  062800  927720  161540  469610  615310  106190  171940  877990  295510  261430  866620  020430  932330  345970  548320  041300  783830  830690  916100  014170  676460  526330  404790  560130  586830  252070  420330  072480  811880  359350  690080  061710  685590  745970  864430  266940  186880  458400  632470  150490  021720  953500  389510  709360  621270  589940  178950  436950  314680  801690  857460  782820  098160  098180  669460  077370  673650  131750  076460  198990  419500  431591  444655  459030  633280  921380  352090  553250  HA109-05  989490R  444510R  390410R  432706  484710  484720  136700  198960  198970  230290  230300  413443  413444  432705  984600R  120230  198910  198930  230280  413441  803810  803810R  131420  131430  912600  965370  856130R  120220  198850  341593  419499  422450  450710R  633250  120210  198820  436780  697210  722780  375410R  524230R  110510  139030  198790  230260  230270  349160  428342  428608  432704  553220  553230  553240  671740  671750  803800  068340  120400  131400  131410  230310  230320  2375300  341591  349159  418581  432702  439398  619620  799540  921370  139020  230200  341590  349158  435195  619610  637560  671570  768780  799530  912580  444070R  139010  198740  230190  410785  419769  440610  637540  671720  671460  684560R  092160R  341588  386580  388258  418573  435193  435194  634920  634930  634970  634990  635030  722770  817050  103150  139000  198690  230140  230150  321830  350841  352080  414693  440550  513080R  671700  737280  799520  866015  804340R  770890R  057520  403449  435191  458990  493800  493810  493820  493830  493850  493860  553200  553210  483640R  946040R  127040  127060  198640  198670  341585  388257  403451  553170  619550  619560  671640  671650  765760R  767020  855710  553160  418150  120200  671630  553150  074170  391466  131380  230110  364827  418854  435189  458970  553140  601620  684540R  825470  640240R  443940R  131360  131370  230090  230100  336811  403448  419768  698900  794200  912570  443930R  340650R  131340  131350  230070  230080  357718  372684  419672  436750  527900  527930  553110  553130  737270  779270  794190  921410R  601630R  186080R  129880  198560  375846  435186  443900R  527850  527860  553090  628090R  700090R  766950  HA063-04  946030R  230000  432695  435183  524190R  527830  698880  817010  861810R  118680R  889050R  352070  357715  388259  410780  414689  432693  436740  527780  553080  125040  068320  074200  357717  422227  435181  435182  527730  527740  671580  794180  443810R  692180R  002040R  002070R  129830  129840  229970  372010  403444  409810  409820  410778  410779  414688  428616  443830R  521810  521820  521830  521840  540760  553050  582610  601610R  671510  671530  671540  671560  782108  794170  834370  984450R  HA052-03  229940  388255  414687  432692  433033  436730  440520  484440R  521760  621000R  671490  921350  443710R  124990  242020R  272330  349154  359548  388254  422225  432691  436710  436720  521710  698870  740700R  799510  921340  068310  074160  124970  385586  409780  511600  730230  HA101-05  HA109-05  027512  116870  124960  127650  127660  129100  129110  129120  129130  229920  385592  403443  419767  432690  440500  511560  511580  511590  521660  541470  589620R  033569  048028  122670  129080  198280  341576  341578  349153  403441  403442  419669  425735  435175  435176  440470  458910  504000  511500  566230R  671470  737250  978468  257710  503960  651070  175930  204880  444661  826920  175900  352130  357720  459150  633370  288130  413445  413446  414698  424890  444665  826900  866016  142270  309770  336105  386590  419677  422510  440491  633330  633350  826890  136740  142220  175880  341596  390405  414696  422500  633320  767080  826880  865225  175810  175840  204870  336104  348055  357719  385566  436800  439399  444658  459120  553290  633300  803830  136720  142210  364828  377416  381912  422490  431656  459080  803820  826850  149270  149280  212710  385630  553280  826840  136710  175710  175720  348054  385628  422480  432707  553270  826820  142200  210556  257730  385625  444668  698850  826810  550410R  038800  175690  175700  375848  385561  431592  436790  441460  553260  741170  826800  856180R  352110  385624  431663  722790  795652  826520  142190  331526  385558  414695  419676  428618  435197  444657  459060  826430  142170  175630  431654  826420  001920  073990  730710  655330  660640  140230  121180  677860  905650  575390  244530  090680  147270  420270  488260  210230  576820  531770  280010  920470  848710  439910  907020  177330  146060  779010  510010  579400  256190  545270  064580  063830  256170  588340  239200  103090  053330  699870  785350  671900  438200  404800  016170  347890  698970  871360  287810  275770  710300  368990  920370  153480  455550  202440  403570  448560  904960  941910  842160  842150  805780  408260  842140  899690  276900  051110  978710  022960  127670  224710  700780  820600  099940  155730  026830  800590  352100  667280  228170  136500  233300  389460  791170  988720  053580  751900  382200  794330  884450  062690  715790  156950  597280  005850  167610  450140  548940  919950  766990  792680  465350  601970  001770  147130  207950  051130  258630  188450  862140  083520  550620  273990  666930  586800  101920  778340  768050  069360  925620  248270  333750  360480  874190  017680  409850  721050  114630  318280  699870  793740  825130  605350  279970  691640  542750  125300  946640  439000  556860  528650  354680  009120  667060  951270  516880  020520  740890  358060  829020  236430  196320  739560  613830  147120  608620  991420  532980  286390  675780  398110  277320  862060  752980  286450  610610  908670  746920  925800  130740  382190  040510  979370  921860  304830  863970  500860  498850  677870  019360  158490  205300  551840  447570  880920  315770  417220  408480  129730  208710  385380  797840  774020  926110  902930  295700  784490  663140  098820  681930  790470  720740  010420  516870  564390  665760  511240  418950  928780  998550  223770  330850  223840  977630  977640  977620  977650  680350  947760  395050  069910  331150  604630  227170  443270  602070  368810  568500  070320  236700  334000  641650  979480  019450  666430  153040  122120  708210  386280  198460  883800  215090  602080  676790  196980  211030  889870  745960  936850  769750  476890  860460  222760  783290  009630  983020  009820  960050  817000  184550  964110  848300  427270  991750  522500  419980  302010  047780  970270  465460  487720  465440  193600  783420  054110  453280  193730  092880  520000  054160  783270  696740    UDI:    (01)00880304035300(17)180326(10)087590  (01)00880304019133(17)181015(10)528130  (01)00880304019140(17)181210(10)589030  (01)00880304019140(17)190103(10)179900  (01)00880304019096(17)190305(10)827050  (01)00880304019119(17)190310(10)827070  (01)00880304019102(17)190429(10)149670  (01)00880304019102(17)190429(10)090400  (01)00880304019119(17)190619(10)119230  (01)00880304702981(17)200126(10)694060  (01)00880304813687(17)210212(10)928740  (01)00880304833890(17)210901(10)227890  (01)00887868166341(17)220802(10)364850  (01)00880304655188(17)230529(10)401880  (01)00880304655171(17)230606(10)408150  (01)00880304655454(17)230701(10)781210  (01)00880304655485(17)230701(10)763830  (01)00880304655478(17)230702(10)763790  (01)00887868259128(17)230731(10)544580  (01)00880304654532(17)230823(10)369920  (01)00880304655980(17)230916(10)537340  (01)00880304551022(17)231008(10)811220  (01)00880304551039(17)231008(10)811320  (01)00880304656222(17)231024(10)963080  (01)00880304656260(17)231029(10)958520  (01)00880304656284(17)231107(10)126020  (01)00880304748378(17)231220(10)489320  (01)00880304748170(17)240109(10)489590  (01)00880304748187(17)240109(10)489830  (01)00880304748194(17)240109(10)490700  (01)00880304748200(17)240109(10)490840  (01)00880304748224(17)240110(10)503260  (01)00880304748231(17)240110(10)503710  (01)00880304655171(17)240113(10)398840  (01)00880304748361(17)240113(10)677530  (01)00880304659551(17)240221(10)099100  (01)00880304719194(17)240325(10)501430  (01)00880304720701(17)240325(10)544030  (01)00880304717787(17)240427(10)725100  (01)00880304551022(17)240518(10)997560  (01)00880304720787(17)240521(10)975970  (01)00880304718128(17)240617(10)418720  (01)00880304718135(17)240617(10)481160  (01)00880304717435(17)240626(10)280200  (01)00880304724754(17)240814(10)122050  (01)00880304551022(17)240909(10)276260  (01)00880304717398(17)240909(10)278030  (01)00880304718753(17)240918(10)595380  (01)00880304718760(17)240918(10)601710  (01)00880304718777(17)240918(10)601700  (01)00880304718791(17)240918(10)601730  (01)00880304718807(17)240918(10)595820  (01)00880304718821(17)240918(10)600810  (01)00880304718838(17)240918(10)636330  (01)00880304718784(17)240918(10)648400  (01)00880304718814(17)240918(10)600600  (01)00880304718555(17)241003(10)816490  (01)00880304721029(17)241113(10)281380  (01)00880304690196(17)241211(10)577580  (01)00880304690202(17)241230(10)684070  (01)00880304700437(17)250213(10)051290  (01)00880304699953(17)250213(10)015910  (01)00880304704787(17)250303(10)233430  (01)00880304704770(17)250303(10)233270  (01)00880304705593(17)250304(10)359560  (01)00880304706729(17)250324(10)566560  (01)00880304709836(17)250511(10)445750  (01)00880304714281(17)250514(10)172350  (01)00880304714168(17)250515(10)030870  (01)00880304716759(17)250611(10)247430  (01)00880304730748(17)250717(10)597870  (01)00880304737419(17)250925(10)332340  (01)00880304737402(17)251005(10)331070  (01)00880304755635(17)251117(10)490540  (01)00880304801462(17)251120(10)920450  (01)00880304806597(17)251211(10)224420  (01)00880304807334(17)251211(10)251940  (01)00880304807327(17)251215(10)191060  (01)00880304808997(17)260115(10)547260  (01)00880304808393(17)260128(10)444650  (01)00880304814981(17)260303(10)080060  (01)00880304815940(17)260318(10)251170  (01)00880304816008(17)260323(10)390910  (01)00880304814936(17)260324(10)425180  (01)00880304816633(17)260401(10)341320  (01)00880304814950(17)260418(10)993180  (01)00880304815995(17)260428(10)390900  (01)00880304820302(17)260505(10)806420  (01)00880304824263(17)260517(10)110870  (01)00880304816756(17)260520(10)341340  (01)00880304825857(17)260526(10)121770  (01)00880304827882(17)260630(10)597460  (01)00880304824102(17)260911(10)977930  (01)00880304836426(17)260912(10)525350  (01)00880304849907(17)261027(10)868480  (01)00880304840294(17)261031(10)607660  (01)00880304856202(17)261230(10)464650  (01)00880304856240(17)261230(10)474060  (01)00880304858077(17)270313(10)443870  (01)00880304858077(17)270314(10)443160  (01)00880304907515(17)270331(10)826490  (01)00880304979895(17)270525(10)186970  (01)00880304992511(17)270601(10)589550  (01)00880304992030(17)270603(10)567500  (01)00880304992047(17)270603(10)568630  (01)00880304992054(17)270603(10)570100  (01)00880304992023(17)270606(10)487210  (01)00880304992061(17)270606(10)570690  (01)00880304992450(17)270607(10)506800  (01)00880304990166(17)270622(10)589820  (01)00887868241727(17)280218(10)721950  (01)00887868241062(17)280325(10)864570  (01)00887868241710(17)280430(10)726280  (01)00887868249020(17)280519(10)022340  (01)00880304753204(17)230916(10)276040  (01)00880304754126(17)240302(10)950370  (01)00880304751781(17)230205(10)419000  (01)00880304751750(17)230208(10)589260  (01)00880304705609(17)250219(10)236170  (01)00880304709041(17)250409(10)664770  (01)00887868226823(17)270920(10)159400  (01)00887868226366(17)270908(10)887090  (01)00880304723405(17)240825(10)193000  (01)00880304702721(17)250206(10)071070  (01)00887868248863(17)280413(10)936700  (01)00880304751897(17)230225(10)605110  (01)00887868245657(17)280314(10)301000  (01)00880304836334(17)260921(10)468560  (01)00880304754041(17)240123(10)539170  (01)00880304805316(17)251123(10)013780  (01)00887868248870(17)280417(10)943830  (01)00887868164248(17)270719(10)156710  (01)00887868245749(17)280623(10)159500  (01)00880304824096(17)260506(10)986190  (01)00880304826007(17)260524(10)196320  (01)00880304855618(17)261230(10)332830  (01)00880304849402(17)261019(10)766240  (01)00880304855175(17)261208(10)207810  (01)00880304821002(17)260531(10)125050  (01)00880304815841(17)260310(10)235260  (01)00880304830677(17)260728(10)919740  (01)00887868224843(17)270825(10)789410  (01)00880304810273(17)260112(10)528640  (01)00880304752917(17)230729(10)868980  (01)00880304751910(17)230214(10)645490  (01)00880304715363(17)250609(10)114740  (01)00887868243790(17)280307(10)130640  (01)00887868229633(17)271016(10)385520  (01)00880304716148(17)250605(10)216810  (01)00880304724006(17)241209(10)486190  (01)00880304865976(17)270306(10)370630  (01)00887868242779(17)280219(10)881420  (01)00887868240997(17)280124(10)680170  (01)00880304907775(17)270402(10)680700  (01)00887868240881(17)280117(10)557830  (01)00880304722309(17)240325(10)423080  (01)00880304811959(17)260126(10)737810  (01)00880304993006(17)270615(10)811110  (01)00880304808966(17)260112(10)581310  (01)00887868227653(17)270928(10)162100  (01)00880304817838(17)260418(10)656120  (01)00887868247439(17)280418(10)845140  (01)00880304992399(17)270602(10)421470  (01)00880304722620(17)240504(10)907980  (01)00880304709652(17)250416(10)737020  (01)00880304856790(17)270118(10)577210  (01)00880304752078(17)230313(10)865520  (01)00880304836327(17)260915(10)466780  (01)00880304826489(17)260603(10)275720  (01)00887868238222(17)271223(10)150630  (01)00887868242939(17)280205(10)920760  (01)00887868245329(17)280413(10)545720  (01)00880304755628(17)251001(10)412190  (01)00880304857025(17)270209(10)694920  (01)00880304722149(17)240319(10)311430  (01)00887868244889(17)280322(10)517200  (01)00880304993396(17)270617(10)550180  (01)00887868163555(17)270630(10)039790  (01)00887868238192(17)280110(10)063370  (01)00887868226441(17)270911(10)823130  (01)00880304713840(17)250429(10)919710  (01)00880304849389(17)261107(10)105930  (01)00880304849396(17)211109(10)766960  (01)00880304856851(17)270124(10)622310  (01)00880304854888(17)261209(10)162390  (01)00880304722408(17)240409(10)598200  (01)00887868248955(17)280416(10)929000  (01)00880304817807(17)260411(10)612360  (01)00880304754102(17)240211(10)806980  (01)00880304752856(17)230718(10)810230  (01)00880304724280(17)240403(10)554930  (01)00880304753464(17)231007(10)690100  (01)00880304689763(17)241117(10)275620  (01)00880304817449(17)260325(10)420300  (01)00880304808874(17)260112(10)530260  (01)00880304856769(17)270118(10)660360  (01)00880304815346(17)260301(10)117440  (01)00880304752344(17)230523(10)292130  (01)00880304988446(17)270519(10)185920  (01)00880304752351(17)230528(10)292140  (01)00887868238659(17)280112(10)007740  (01)00887868243943(17)280213(10)712160  (01)00880304733831(17)250909(10)072110  (01)00880304734753(17)200908(10)217050  (01)00887868163050(17)270616(10)809270  (01)00880304979765(17)270422(10)934930  (01)00887868242830(17)280212(10)837160  (01)00880304815834(17)260307(10)176270  (01)00880304810303(17)260115(10)529630  (01)00880304869479(17)270422(10)209030  (01)00880304980235(17)270425(10)020940  (01)00880304817500(17)260323(10)443320  (01)00880304856776(17)270118(10)562130  (01)00880304753402(17)231007(10)544860  (01)00880304832503(17)260812(10)135220  (01)00887868246036(17)280405(10)612300  (01)00880304854840(17)261228(10)207310  (01)00880304756199(17)251027(10)615910  (01)00880304731653(17)250804(10)801240  (01)00887868229879(17)271019(10)384350  (01)00887868163692(17)270717(10)944240  (01)00880304722248(17)240323(10)328460  (01)00887868163340(17)270626(10)939390  (01)00887868246692(17)280323(10)539600  (01)00880304840522(17)261014(10)589930  (01)00880304857155(17)270126(10)711170  (01)00880304722866(17)240617(10)339730  (01)00880304728097(17)250616(10)322770  (01)00880304808423(17)260108(10)403790  (01)00880304992016(17)270524(10)438780  (01)00880304729308(17)250622(10)376430  (01)00880304752290(17)230503(10)233310  (01)00880304811447(17)260208(10)875070  (01)00880304754201(17)240617(10)414760  (01)00880304833630(17)260818(10)050470  (01)00887868163951(17)270627(10)568660  (01)00880304858190(17)270213(10)001770  (01)00880304840614(17)261004(10)687820  (01)00880304866119(17)270328(10)454530  (01)00880304820890(17)260516(10)110500  (01)00887868227097(17)271010(10)418750  (01)00880304754263(17)240908(10)334620  (01)00880304989740(17)270517(10)278110  (01)00880304756267(17)251028(10)618540  (01)00880304993372(17)270616(10)695890  (01)00880304820616(17)260518(10)938860  (01)00887868238413(17)271217(10)137250  (01)00887868229947(17)271026(10)618690  (01)00887868239038(17)271204(10)925090  (01)00887868244636(17)280326(10)578700  (01)00880304816572(17)260325(10)321680  (01)00887868230127(17)271025(10)644290  (01)00880304817081(17)260401(10)408070  (01)00880304855731(17)270104(10)389150  (01)00887868231919(17)271116(10)891160  (01)00880304979505(17)270427(10)956430  (01)00880304810297(17)260111(10)603020  (01)00887868242793(17)280217(10)938090  (01)00880304907560(17)270329(10)525860  (01)00880304979512(17)270415(10)909040  (01)00887868245107(17)280315(10)518060  (01)00880304864696(17)270302(10)276590  (01)00880304722811(17)240612(10)394320  (01)00887868244094(17)280328(10)291120  (01)00880304979932(17)270506(10)387410  (01)00880304804180(17)251120(10)966710  (01)00887868226786(17)270916(10)074050  (01)00880304988323(17)270513(10)011200  (01)00880304856745(17)270110(10)556010  (01)00880304852747(17)261123(10)093010  (01)00880304907492(17)270316(10)396670  (01)00880304833685(17)260825(10)244810  (01)00887868246876(17)280409(10)730550  (01)00887868238444(17)271208(10)137120  (01)00880304989573(17)270510(10)170300  (01)00880304752887(17)230724(10)822110  (01)00887868240973(17)280110(10)506410  (01)00887868237799(17)271130(10)042730  (01)00880304907867(17)270407(10)734550  (01)00887868237997(17)271129(10)991780  (01)00880304824300(17)260518(10)991760  (01)00880304752337(17)230430(10)304580  (01)00880304722989(17)240716(10)645390  (01)00887868229329(17)271013(10)457660  (01)00887868163401(17)220630(10)941750  (01)00880304812963(17)260926(10)532110  (01)00887868240911(17)280215(10)720320  (01)00880304807754(17)260104(10)109760  (01)00880304816862(17)260323(1",
      "center_classification_date": "20191122",
      "distribution_pattern": "US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.",
      "state": "IN",
      "product_description": "Various Custom Products, Item Nos. CP102940  CP102941  CP102942  CP102943  CP102944  CP102945  CP102946  CP102947  CP102948  CP112662  CP112663  CP112664  CP112674  CP112675  CP113771  CP113777  CP113778  CP113782  CP113783  CP114083  CP114090  CP114112  CP114118  CP114148  CP114213  CP114388  CP114436  CP114463  CP114472  CP114478  CP114495  CP114532  CP114561  CP114563  CP114605  CP114695  CP114756  CP114782  CP114783  CP114933  CP114987  CP115008  CP115048  CP115049  CP115050  CP115166  CP115209  CP115212  CP115312  CP115313  CP115314  CP115318  CP115319  CP115320  CP115322  CP115357  CP115358  CP115359  CP115360  CP115362  CP115363  CP115371  CP161304  CP161305  CP161306  CP161307  CP161522  CP161560  CP161735  CP161741  CP161751  CP161752  CP161753  CP161768  CP161793  CP161795  CP161831  CP161867  CP162006  CP162007  CP162008  CP162013  CP162017  CP162042  CP162043  CP162055  CP162084  CP162085  CP162086  CP162121  CP162149  CP162150  PM0000711  PM103016  PM103022  PM103060  PM155537  PM155577  PM155578  PM155580  PM155590  PM155591  PM155592  PM155594  PM155595  PM155602  PM155603  PM155604  PM155610  PM155612  PM155615  PM155616  PM155618  PM155623  PM155630  PM155635  PM155638  PM155640  PM155641  PM155643  PM155644  PM155646  PM155650  PM155653  PM155654  PM155656  PM155660  PM155662  PM155663  PM155677  PM155678  PM155685  PM155686  PM155687  PM155694  PM155698  PM155702  PM155710  PM155711  PM155712  PM155715  PM155717  PM155724  PM155725  PM155727  PM155728  PM155730  PM155734  PM155737  PM155738  PM155742  PM155746  PM155750  PM155751  PM155753  PM155756  PM155757  PM155761  PM155762  PM155767  PM155771  PM155774  PM155780  PM155781  PM155785  PM155789  PM155791  PM155792  PM155793  PM155795  PM155804  PM155806  PM155808  PM155809  PM155812  PM155813  PM155815  PM155818  PM155820  PM155823  PM155825  PM155828  PM155829  PM155840  PM155841  PM155842  PM155843  PM155847  PM155849  PM155850  PM155854  PM155857  PM155860  PM155863  PM155867  PM155868  PM155871  PM155872  PM155879  PM155882  PM155887  PM155888  PM155890  PM155892  PM155893  PM155894  PM155895  PM155897  PM155903  PM155905  PM155909  PM155913  PM155922  PM155926  PM155927  PM155928  PM155929  PM155933  PM155938  PM155939  PM155940  PM155942  PM155945  PM155947  PM155948  PM155950  PM155952  PM155963  PM155965  PM155972  PM155974  PM155979  PM155982  PM155983  PM155986  PM155988  PM155989  PM155991  PM155993  PM155997  PM155998  PM156000  PM156002  PM156007  PM156008  PM156012  PM156035  PM156038  PM156039  PM156041  PM156043  PM156044  PM156047  PM156048  PM156060  PM156061  PM156072  PM156079  PM156085  PM156092  PM156095  PM156101  PM156104  PM156110  PM156116  PM156132  PM156133  PM156143  PM156147  PM156154  PM156165  PM156166  PM156173  PM156184  PM156188  PM156192  PM156193  PM156197  PM156205  PM156207  PM156219  PM156235  PM156236  PM156244  PM156248  PM156254  PM156266  PM156275  PM156280  PM156281  PM156290  PM156292  PM156293  PM156320  PM156329  PM156340  PM156343  PM156345  PM156351  PM156352  PM156356  PM156357  PM156358  PM156362  PM156377  PM156379  PM156382  PM156384  PM156387  PM156388  PM156390  PM156391  PM156392  PM156393  PM156397  PM156402  PM156403  PM156405  PM156406  PM156409  PM156410  PM156431  PM156434  PM156435  PM156438  PM156444  PM156448  PM156453  PM156457  PM156458  PM156460  PM156477  PM156485  PM156489  PM156492  PM156497  PM156501  PM156510  PM156511  PM156520  PM156529  PM156532  PM156533  PM156539  PM156544  PM156546  PM156551  PM156560  PM156561  PM156563  PM156565  PM156567  PM156570  PM156572  PM156573  PM156578  PM156584  PM156589  PM156595  PM156596  PM156604  PM156605  PM156609  PM156610  PM156620  PM156632  PM156641  PM156643  PM156656  PM156663  PM156666  PM156668  PM156674  PM156675  PM156676  PM156680  PM156681  PM156682  PM156684  PM156688  PM156691  PM156692  PM156693  PM156698  PM156700  PM156704  PM156705  PM156713  PM156718  PM156803  PM156814  PM156821  PM156829  PM156830  PM156831  PM156834  PM15",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0526-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83783",
      "termination_date": "20210721",
      "more_code_info": "0)373390  (01)00880304814523(17)260210(10)902470  (01)00880304722934(17)240630(10)458790  (01)00880304733336(17)250901(10)930850  (01)00887868243707(17)280215(10)943880  (01)00887868245404(17)280315(10)255810  (01)00880304753013(17)230827(10)211580  (01)00887868227141(17)221012(10)537220  (01)00887868247217(17)280320(10)538610  (01)00887868239175(17)271228(10)281870  (01)00880304820296(17)260419(10)798890  (01)00887868244032(17)280306(10)275460  (01)00880304865853(17)270313(10)274450  (01)00880304756359(17)251020(10)606240  (01)00887868245084(17)280327(10)401780  (01)00887868242410(17)280115(10)437560  (01)00880304815827(17)260303(10)152770  (01)00880304694514(17)250120(10)889870  (01)00880304752894(17)230730(10)822120  (01)00880304752993(17)230808(10)108430  (01)00887868243639(17)280223(10)122290  (01)00880304820210(17)260427(10)833440  (01)00887868240867(17)280111(10)558550  (01)00880304816855(17)260323(10)420210  (01)00880304980020(17)270427(10)937130  (01)00880304724327(17)240429(10)928170  (01)00880304827547(17)260616(10)439040  (01)00880304722040(17)240210(10)062360  (01)00880304722170(17)240320(10)205290  (01)00880304713048(17)250423(10)841780  (01)00887868231025(17)271117(10)955590  (01)00887868243868(17)280228(10)123190  (01)00880304756755(17)251106(10)682900  (01)00887868238673(17)271207(10)897960  (01)00880304753976(17)240109(10)421560  (01)00880304723856(17)241117(10)130450  (01)00880304860001(17)270228(10)191230  (01)00887868224317(17)270816(10)591130  (01)00880304814691(17)210225(10)939060  (01)00880304992184(17)270526(10)490720  (01)00887868248221(17)280517(10)659250  (01)00887868238918(17)271217(10)244660  (01)00880304715653(17)250601(10)150390  (01)00887868240836(17)271228(10)315890  (01)00880304706132(17)250317(10)374770  (01)00880304722064(17)240331(10)130760  (01)00880304866058(17)270327(10)287200  (01)00880304756762 (17)251105(10)736180  (01)00880304757721(17)251110(10)825570  (01)00880304753730(17)231122(10)147110  (01)00887868241024(17)280117(10)512540  (01)00887868163937(17)270719(10)262940  (01)00880304722224(17)240317(10)363110  (01)00880304731233(17)250708(10)677760  (01)00880304723269(17)240810(10)897650  (01)00887868248290(17)280513(10)760430  (01)00887868240249(17)280130(10)762840  (01)00880304819818(17)260608(10)779480  (01)00887868166358(17)270726(10)196960  (01)00880304729469(17)250709(10)378840  (01)00880304754157(17)240507(10)834140  (01)00880304716902(17)250616(10)246700  (01)00880304907874(17)270413(10)869970  (01)00880304722194(17)240324(10)205310  (01)00880304855694(17)261230(10)336130  (01)00880304907782(17)270402(10)705110  (01)00887868163272(17)271012(10)173380  (01)00880304914742(17)270412(10)760180  (01)00880304702837(17)250220(10)202310  (01)00880304753495(17)231024(10)750010  (01)00880304752375(17)230523(10)305500  (01)00887868249396(17)280416(10)944080  (01)00880304992177(17)270526(10)506680  (01)00880304816169(17)260309(10)235010  (01)00880304993358(17)270626(10)601270  (01)00887868243578(17)280221(10)070290  (01)00880304732315(17)250803(10)843970  (01)00880304734234(17)250828(10)072740  (01)00880304705586(17)250304(10)242860  (01)00887868249440(17)280523(10)676870  (01)00880304858503(17)270221(10)138470  (01)00887868237676(17)271118(10)930110  (01)00880304814554(17)260216(10)961370  (01)00880304810662(17)260125(10)651380  (01)00880304816725(17)260323(10)373430  (01)00880304817586(17)260412(10)443740  (01)00880304752252(17)230417(10)157100  (01)00887868246883(17)280404(10)703760  (01)00887868248207(17)280622(10)073790  (01)00880304751798(17)230207(10)605100  (01)00880304817791(17)260413(10)588610  (01)00880304979901(17)270413(10)836330  (01)00880304865396(17)270415(10)729530  (01)00880304840492(17)261005(10)667240  (01)00887868163999(17)270720(10)160530  (01)00887868241550(17)280228(10)727120  (01)00887868245091(17)280320(10)517650  (01)00880304753969(17)240109(10)421570  (01)00880304989672(17)270510(10)286550  (01)00880304814899(17)260303(10)872930  (01)00880304816176(17)260314(10)286080  (01)00880304734357(17)250916(10)198320  (01)00880304755178(17)181014(10)738980  (01)00880304755185(17)231014(10)742200  (01)00887868247422(17)280406(10)811230  (01)00880304859982(17)270301(10)194530  (01)00880304723528(17)240925(10)586790  (01)00887868163395(17)270622(10)909350  (01)00880304812390(17)260129(10)722950  (01)00880304817814(17)260411(10)589610  (01)00880304752795(17)230718(10)670560  (01)00880304753044(17)230816(10)167600  (01)00880304728806(17)240225(10)950390  (01)00880304753532(17)231016(10)846930  (01)00880304702844(17)250304(10)273610  (01)00887868226229(17)270908(10)946230  (01)00880304713697(17)250513(10)919080  (01)00880304858367(17)270210(10)017120  (01)00880304724013(17)241210(10)370170  (01)00880304806542(17)251214(10)049800  (01)00880304752955(17)230814(10)216230  (01)00880304816473(17)260315(10)281420  (01)00880304828162(17)210715(10)569280  (01)00887868246500(17)280406(10)526920  (01)00880304715660(17)250611(10)151740  (01)00880304865754(17)270324(10)370280  (01)00887868164156(17)270718(10)241640  (01)00880304756892(17)251106(10)786520  (01)00880304806634 (17)251209(10)108390  (01)00887868241154(17)280113(10)557020  (01)00880304737761(17)250925(10)389180  (01)00880304804449(17)251124(10)902980  (01)00880304756809(17)251103(10)732250  (01)00880304828445(17)260713(10)635780  (01)00880304816183(17)260316(10)375690  (01)00880304721975(17)230724(10)799700  (01)00887868239182(17)271221(10)283280  (01)00887868237683(17)271117(10)930740  (01)00887868249273(17)280516(10)620300  (01)00880304729827(17)250710(10)455300  (01)00880304991606(17)270523(10)332010  (01)00887868226359(17)270909(10)887060  (01)00880304907799(17)270407(10)734450  (01)00880304824089(17)260506(10)937270  (01)00880304751682(17)230102(10)114500  (01)00880304980044(17)270407(10)734480  (01)00880304989979(17)270518(10)170910  (01)00880304704886(17)250317(10)348950  (01)00880304753396(17)231016(10)675490  (01)00880304756205(17)251020(10)613890  (01)00880304727458(17)241001(10)699400  (01)00887868245725(17)280622(10)156070  (01)00887868243745(17)280301(10)122100  (01)00887868248139(17)280513(10)574740  (01)00880304801431(17)251202(10)869290  (01)00880304752115(17)230312(10)878000  (01)00880304989313(17)270513(10)238180  (01)00887868238925(17)271228(10)246960  (01)00880304689497(17)241209(10)397420  (01)00880304857018(17)270131(10)877420  (01)00887868238000(17)271201(10)007160  (01)00880304827325(17)260613(10)408830  (01)00887868245763(17)280627(10)159760  (01)00880304829770(17)261005(10)799840  (01)00880304840607(17)260929(10)687150  (01)00880304729445(17)250612(10)260200  (01)00880304826427(17)260607(10)258920  (01)00887868247156(17)280627(10)159790  (01)00880304752849(17)230717(10)810210  (01)00880304840409(17)260925(10)634170  (01)00880304730564(17)250709(10)508820  (01)00887868163425(17)270628(10)940830  (01)00880304689442(17)241209(10)397430  (01)00880304715875(17)250617(10)237190  (01)00887868224898(17)220919(10)544250  (01)00887868230257(17)271107(10)753730  (01)00880304723191(17)240810(10)095490  (01)00880304717220(17)250616(10)312780  (01)00880304805958(17)251209(10)179770  (01)00880304756779(17)251021(10)664400  (01)00880304752207(17)230416(10)142950  (01)00880304833876(17)260824(10)275880  (01)00880304716933(17)250617(10)260090  (01)00887868239311(17)280110(10)374430  (01)00880304723801(17)241107(10)115360  (01)00880304806511(17)251223(10)178630  (01)00880304693319(17)250206(10)771690  (01)00880304732872(17)250819(10)098640  (01)00887868227776(17)271020(10)615650  (01)00887868227202(17)271012(10)400370  (01)00887868164019(17)270716(10)986350  (01)00880304830646(17)260726(10)728730  (01)00887868237850(17)271219(10)920800  (01)00887868237874(17)271204(10)921860  (01)00887868237867(17)271218(10)342330  (01)00880304729506(17)250626(10)429750  (01)00887868243523(17)280207(10)723620  (01)00880304856738(17)270113(10)561110  (01)00880304979840(17)270425(10)929100  (01)00880304817869(17)210426(10)589950  (01)00880304722392(17)240421(10)573460  (01)00880304820937(17)260531(10)099590  (01)00880304727465(17)240930(10)699240  (01)00880304756335(17)251027(10)618930  (01)00880304752368(17)230516(10)305470  (01)00887868214493(17)270816(10)639710  (01)00880304816435(17)260311(10)232930  (01)00887868247354(17)280420(10)831590  (01)00880304696006(17)250206(10)925250  (01)00887868226250(17)270908(10)946730  (01)00880304722491(17)240422(10)597290  (01)00880304993013(17)270612(10)749240  (01)00880304808782(17)260112(10)569430  (01)00880304855465(17)261223(10)297250  (01)00880304993037(17)270615(10)748550  (01)00887868224454(17)270823(10)673820  (01)00880304820876(17)260518(10)084110  (01)00880304723603(17)240930(10)699300  (01)00880304752139(17)230319(10)921610  (01)00880304849334(17)261014(10)764020  (01)00880304806580(17)251203(10)990450  (01)00880304855649(17)261228(10)417450  (01)00887868244537(17)280312(10)323120  (01)00887868231100(17)271111(10)795930  (01)00887868227875(17)271011(10)419480  (01)00887868238949(17)271217(10)243500  (01)00880304752269(17)230516(10)316210  (01)00880304751606(17)230220(10)605120  (01)00887868248900(17)280417(10)973010  (01)00880304756342(17)251020(10)605180  (01)00880304852877(17)261206(10)270400  (01)00887868245916(17)280328(10)518370  (01)00880304756502(17)251021(10)692970  (01)00880304722651(17)240521(10)123540  (01)00880304865822(17)270311(10)376930  (01)00887868214264(17)270816(10)433010  (01)00887868243196(17)280221(10)932610  (01)00880304807907(17)260107(10)130200  (01)00880304992573(17)270607(10)590880  (01)00880304817630(17)260404(10)571090  (01)00887868223969(17)270817(10)435060  (01)00887868229404(17)271026(10)665920  (01)00887868242922(17)280212(10)926030  (01)00880304816732(17)260325(10)473900  (01)00880304989689(17)270524(10)070460  (01)00880304832350(17)260812(10)091600  (01)00887868247644(17)280516(10)845150  (01)00887868238154(17)271217(10)048170  (01)00887868226526(17)270913(10)806980  (01)00880304722767(17)240611(10)171930  (01)00880304907959(17)270422(10)968810  (01)00880304764125(17)251117(10)836780  (01)00880304722576(17)240504(10)058140  (01)00880304815681(17)260223(10)138580  (01)00880304724419(17)241010(10)708210  (01)00887868245640(17)280319(10)300230  (01)00887868163616(17)270708(10)023100  (01)00880304752610(17)230618(10)508060  (01)00887868240966(17)280320(10)518910  (01)00880304753501(17)231016(10)837960  (01)00880304850477(17)261117(10)005400  (01)00880304805347(17)251202(10)052840  (01)00887868243561(17)280213(10)925710  (01)00887868166464(17)270816(10)544110  (01)00887868243202(17)280302(10)932730  (01)00880304756151(17)251020(10)613030  (01)00880304724297(17)240410(10)728920  (01)00880304723559(17)240919(10)475630  (01)00880304817593(17)260329(10)496780  (01)00880304857476(17)270203(10)781170  (01)00880304752177(17)230405(10)997350  (01)00880304979628(17)270415(10)889540  (01)00880304691100(17)250115(10)725110  (01)00880304722330(17)240409(10)598210  (01)00880304706774(17)250323(10)527270  (01)00880304832176(17)260803(10)014860  (01)00880304700833(17)250220(10)126460  (01)00880304693890(17)250119(10)892830  (01)00880304815032(17)260301(10)137800  (01)00880304732339(17)250811(10)883970  (01)00880304722910(17)240630(10)458730  (01)00880304753549(17)231127(10)147100  (01)00880304753723(17)231206(10)402940  (01)00880304992283(17)270607(10)505410  (01)00887868240409(17)271228(10)342380  (01)00880304700420(17)250217(10)069230  (01)00880304980709(17)270428(10)936450  (01)00880304754164(17)240507(10)907970  (01)00880304752023(17)230319(10)848700  (01)00880304703261(17)250309(10)242850  (01)00880304806566(17)251210(10)078580  (01)00880304183025(17)180828(10)515790  (01)00880304183025(17)181228(10)469150  (01)00880304183025(17)230214(10)379800  (01)00880304183025(17)231209(10)791960  (01)00880304183025(17)240212(10)115210  (01)00880304183025(17)240817(10)596200  (01)00880304183025(17)241216(10)995500  (01)00880304183025(17)250324(10)070640  (01)00880304183025(17)250407(10)345170  (01)00880304183025(17)251007(10)139040  (01)00880304183025(17)260202(10)321200  (01)00880304183025(17)260415(10)287360  (01)00880304183025(17)270727(10)345090  (01)00880304183025(17)280126(10)554820  (01)00880304183032(17)180828(10)304560  (01)00880304183032(17)231022(10)539650  (01)00880304183032(17)231220(10)873100  (01)00880304183032(17)240123(10)326980  (01)00880304183032(17)240222(10)417080  (01)00880304183032(17)240529(10)646120  (01)00880304183032(17)241013(10)620310  (01)00880304183032(17)241112(10)756540  (01)00880304183032(17)241201(10)232000  (01)00880304183032(17)241217(10)383150  (01)00880304183032(17)241219(10)364500  (01)00880304183032(17)241220(10)384310  (01)00880304183032(17)250109(10)506770  (01)00880304183032(17)250114(10)383160  (01)00880304183032(17)250209(10)567570  (01)00880304183032(17)250828(10)922270  (01)00880304183032(17)250917(10)138950  (01)00880304183032(17)250917(10)138960  (01)00880304183032(17)260324(10)190150  (01)00880304183032(17)260330(10)997290  (01)00880304183032(17)261230(10)120380  (01)00880304183032(17)270410(10)396530  (01)00880304183032(17)271220(10)291070  (01)00880304183032(17)280130(10)978020  (01)00880304183155(17)250129(10)383420  (01)00880304183155(17)250303(10)901190  (01)00880304183155(17)250409(10)042880  (01)00880304183148(17)241120(10)808050  (01)00880304183148(17)241120(10)808050R  (01)00880304183131(17)240707(10)389120  (01)00880304183124(17)180701(10)260930R  (01)00880304183124(17)280430(10)498380  (01)00880304183117(17)231212(10)194280  (01)00880304183117(17)240323(10)136900  (01)00880304183117(17)240403(10)173900  (01)00880304183117(17)240512(10)756220  (01)00880304183117(17)250318(10)254490  (01)00880304183117(17)250527(10)765720  (01)00880304183100(17)231023(10)011030  (01)00880304183100(17)231209(10)778370  (01)00880304183100(17)231212(10)194290  (01)00880304183100(17)240302(10)811300  (01)00880304183100(17)241118(10)955550  (01)00880304183100(17)250305(10)107080  (01)00880304183100(17)250819(10)816970  (01)00880304183100(17)251214(10)812050  (01)00880304183100(17)260215(10)827130  (01)00880304183100(17)270908(10)283180  (01)00880304183100(17)280312(10)781370  (01)00880304183094(17)221101(10)983130R  (01)00880304183094(17)240225(10)417110  (01)00880304183094(17)240615(10)967140  (01)00880304183094(17)241112(10)969640  (01)00880304183094(17)241120(10)620350  (01)00880304183094(17)250108(10)364540  (01)00880304183094(17)250108(10)619080  (01)00880304183094(17)250616(10)179390  (01)00880304183094(17)250924(10)273120  (01)00880304183094(17)250929(10)131170  (01)00880304183094(17)251211(10)620320  (01)00880304183094(17)260204(10)684910  (01)00880304183094(17)260210(10)684920  (01)00880304183094(17)260509(10)706410  (01)00880304183094(17)260513(10)670240  (01)00880304183094(17)270816(10)312210  (01)00880304183094(17)270830(10)312250  (01)00880304183094(17)271127(10)291290  (01)00880304183094(17)280202(10)291340  (01)00880304183094(17)280224(10)781360  (01)00880304183087(17)230128(10)053120  (01)00880304183087(17)231109(10)918880  (01)00880304183087(17)231118(10)570950  (01)00880304183087(17)231118(10)406190  (01)00880304183087(17)240109(10)798630  (01)00880304183087(17)240201(10)136890  (01)00880304183087(17)240219(10)587290  (01)00880304183087(17)240222(10)811290  (01)00880304183087(17)240304(10)417100  (01)00880304183087(17)240330(10)050510  (01)00880304183087(17)240601(10)488670  (01)00880304183087(17)240701(10)967160  (01)00880304183087(17)240817(10)894410  (01)00880304183087(17)240819(10)811470  (01)00880304183087(17)240902(10)017390  (01)00880304183087(17)241014(10)620340  (01)00880304183087(17)241113(10)995510  (01)00880304183087(17)250216(10)800260  (01)00880304183087(17)250324(10)345210  (01)00880304183087(17)250325(10)724930  (01)00880304183087(17)250616(10)232270  (01)00880304183087(17)250722(10)389590  (01)00880304183087(17)250819(10)816960  (01)00880304183087(17)250915(10)922310  (01)00880304183087(17)250929(10)139050  (01)00880304183087(17)250930(10)139060  (01)00880304183087(17)260210(10)827100  (01)00880304183087(17)260211(10)704410  (01)00880304183087(17)260416(10)287370  (01)00880304183087(17)260516(10)855870  (01)00880304183087(17)270206(10)257830  (01)00880304183087(17)270407(10)468720  (01)00880304183087(17)270720(10)366760  (01)00880304183087(17)271121(10)221040  (01)00880304183087(17)271216(10)291250  (01)00880304183087(17)280115(10)440150  (01)00880304183087(17)280126(10)554940  (01)00880304183087(17)280223(10)085280  (01)00880304183087(17)280417(10)459190  (01)00880304183070(17)230604(10)186750  (01)00880304183070(17)230701(10)595630  (01)00880304183070(17)230702(10)459600  (01)00880304183070(17)230718(10)674420  (01)00880304183070(17)230722(10)744730  (01)00880304183070(17)230828(10)037170  (01)00880304183070(17)231003(10)454620  (01)00880304183070(17)231021(10)175780  (01)00880304183070(17)231024(10)742470  (01)00880304183070(17)231116(10)570940  (01)00880304183070(17)231118(10)790240  (01)00880304183070(17)231204(10)494000  (01)00880304183070(17)231212(10)007650  (01)00880304183070(17)240103(10)741150  (01)00880304183070(17)240207(10)459660  (01)00880304183070(17)240213(10)136880  (01)00880304183070(17)240229(10)562600  (01)00880304183070(17)240302(10)053830  (01)00880304183070(17)240302(10)115220  (01)00880304183070(17)240303(10)488800  (01)00880304183070(17)240327(10)256560  (01)00880304183070(17)240330(10)863240  (01)00880304183070(17)240331(10)863240R  (01)00880304183070(17)240424(10)488660  (01)00880304183070(17)240428(10)685170  (01)00880304183070(17)240505(10)132680  (01)00880304183070(17)240508(10)488650  (01)00880304183070(17)240617(10)764820  (01)00880304183070(17)240625(10)764830  (01)00880304183070(17)240902(10)017380  (01)00880304183070(17)240930(10)006850  (01)00880304183070(17)240930(10)691220  (01)00880304183070(17)240930(10)506090  (01)00880304183070(17)241215(10)364530  (01)00880304183070(17)250115(10)567580  (01)00880304183070(17)250127(10)611310  (01)00880304183070(17)250325(10)345200  (01)00880304183070(17)250505(10)765710  (01)00880304183070(17)250824(10)816950  (01)00880304183070(17)250918(10)138930  (01)00880304183070(17)251002(10)138940  (01)00880304183070(17)260125(10)228850  (01)00880304183070(17)260125(10)527310  (01)00880304183070(17)260210(10)827120  (01)00880304183070(17)260215(10)724610  (01)00880304183070(17)260309(10)057740  (01)00880304183070(17)260418(10)483300  (01)00880304183070(17)260509(10)706400  (01)00880304183070(17)270203(10)243220  (01)00880304183070(17)270804(10)885080  (01)00880304183070(17)270811(10)885070  (01)00880304183070(17)270813(10)028590  (01)00880304183070(17)271219(10)291170  (01)00880304183070(17)271229(10)291240  (01)00880304183070(17)280124(10)554920  (01)00880304183070(17)280129(10)554930  (01)00880304183070(17)280324(10)542760  (01)00880304183063(17)230101(10)236420R  (01)00880304183063(17)230517(10)201100  (01)00880304183063(17)230522(10)139670  (01)00880304183063(17)230530(10)389010  (01)00880304183063(17)230531(10)139550  (01)00880304183063(17)231213(10)798620  (01)00880304183063(17)231219(10)384390  (01)00880304183063(17)231220(10)943170  (01)00880304183063(17)231220(10)136870  (01)00880304183063(17)240108(10)873200  (01)00880304183063(17)240214(10)526710  (01)00880304183063(17)240221(10)136860  (01)00880304183063(17)240225(10)895710  (01)00880304183063(17)240304(10)408470  (01)00880304183063(17)240323(10)133760  (01)00880304183063(17)240330(10)132670  (01)00880304183063(17)240507(10)405560  (01)00880304183063(17)240520(10)646130  (01)00880304183063(17)240605(10)764810  (01)00880304183063(17)240609(10)026950  (01)00880304183063(17)240716(10)002860  (01)00880304183063(17)240916(10)017370  (01)00880304183063(17)241013(10)588600  (01)00880304183063(17)241028(10)620330  (01)00880304183063(17)241112(10)969730  (01)00880304183063(17)241229(10)296400  (01)00880304183063(17)241230(10)454540  (01)00880304183063(17)250105(10)364520  (01)00880304183063(17)250106(10)510220  (01)00880304183063(17)250112(10)383180  (01)00880304183063(17)250120(10)705750  (01)00880304183063(17)250128(10)551420  (01)00880304183063(17)250209(10)611300  (01)00880304183063(17)250216(10)940380  (01)00880304183063(17)250217(10)585990  (01)00880304183063(17)250305(10)032320  (01)00880304183063(17)250306(10)800220  (01)00880304183063(17)250326(10)099250  (01)00880304183063(17)250423(10)439010  (01)00880304183063(17)250901(10)922300  (01)00880304183063(17)250916(10)139000  (01)00880304183063(17)250917(10)119880  (01)00880304183063(17)250917(10)139020  (01)00880304183063(17)250918(10)139030  (01)00880304183063(17)250918(10)139010  (01)00880304183063(17)250929(10)131160  (01)00880304183063(17)251005(10)306930  (01)00880304183063(17)260125(10)228830  (01)00880304183063(17)260126(10)228840  (01)00880304183063(17)260203(10)546440  (01)00880304183063(17)260210(10)753900  (01)00880304183063(17)260221(10)963240  (01)00880304183063(17)260421(10)483290  (01)00880304183063(17)260601(10)981360  (01)00880304183063(17)260614(10)168700  (01)00880304183063(17)270113(10)150800  (01)00880304183063(17)270302(10)631840  (01)00880304183063(17)271204(10)291150  (01)00880304183063(17)280127(10)291110  (01)00880304183063(17)280127(10)554900  (01)00880304183063(17)280130(10)554890  (01)00880304183063(17)280223(10)929630  (01)00880304183056(17)230228(10)273800  (01)00880304183056(17)230531(10)340640  (01)00880304183056(17)231029(10)596960  (01)00880304183056(17)231206(10)873190  (01)00880304183056(17)231212(10)079500  (01)00880304183056(17)240214(10)417090  (01)00880304183056(17)240214(10)619740  (01)00880304183056(17)240217(10)078350  (01)00880304183056(17)240316(10)847120  (01)00880304183056(17)240429(10)333600  (01)00880304183056(17)240512(10)758390  (01)00880304183056(17)240512(10)405550  (01)00880304183056(17)240525(10)846960  (01)00880304183056(17)240528(10)020390  (01)00880304183056(17)240602(10)784900  (01)00880304183056(17)240625(10)002850  (01)00880304183056(17)240701(10)279750  (01)00880304183056(17)240721(10)435310  (01)00880304183056(17)240814(10)811520  (01)00880304183056(17)240904(10)962770  (01)00880304183056(17)240929(10)444690  (01)00880304183056(17)241013(10)602190  (01)00880304183056(17)241022(10)620320  (01)00880304183056(17)241029(10)955540  (01)00880304183056(17)241030(10)955530  (01)00880304183056(17)241105(10)045920  (01)00880304183056(17)241113(10)766610  (01)00880304183056(17)241215(10)364510  (01)00880304183056(17)250209(10)628930  (01)00880304183056(17)250217(10)800250  (01)00880304183056(17)250325(10)345190  (01)00880304183056(17)250501(10)765700  (01)00880304183056(17)250602(10)055580  (01)00880304183056(17)250610(10)026050  (01)00880304183056(17)250618(10)232120  (01)00880304183056(17)250819(10)816940  (01)00880304183056(17)250911(10)922290  (01)00880304183056(17)250917(10)138910  (01)00880304183056(17)250918(10)138900  (01)00880304183056(17)250918(10)138920  (01)00880304183056(17)260210(10)753710  (01)00880304183056(17)260210(10)753730  (01)00880304183056(17)260315(10)225790  (01)00880304183056(17)260329(10)225780  (01)00880304183056(17)260509(10)670220  (01)00880304183056(17)260520(10)969230  (01)00880304183056(17)260704(10)258520  (01)00880304183056(17)261230(10)187010  (01)00880304183056(17)270327(10)113150  (01)00880304183056(17)270407(10)243200  (01)00880304183056(17)270802(10)028560  (01)00880304183056(17)270821(10)028550  (01)00880304183056(17)270830(10)440130  (01)00880304183056(17)280127(10)554870  (01)00880304183056(17)280127(10)667070  (01)00880304183056(17)280129(10)291100  (01)00880304183056(17)280201(10)554880  (01)00880304183056(17)280224(10)781290  (01)00880304183056(17)280421(10)504450  (01)00880304183056(17)280423(10)459210  (01)00880304183056(17)280430(10)315010  (01)00880304183049(17)190728(10)659990  (01)00880304183049(17)230606(10)460470  (01)00880304183049(17)230606(10)225000  (01)00880304183049(17)230606(10)473070  (01)00880304183049(17)230606(10)345750  (01)00880304183049(17)230930(10)176160  (01)00880304183049(17)231014(10)429790  (01)00880304183049(17)231024(10)661430  (01)00880304183049(17)231112(10)778360  (01)00880304183049(17)231116(10)790120  (01)00880304183049(17)231118(10)570930  (01)00880304183049(17)240113(10)136850  (01)00880304183049(17)240214(10)873110  (01)00880304183049(17)240222(10)488790  (01)00880304183049(17)240225(10)654730  (01)00880304183049(17)240401(10)060660  (01)00880304183049(17)240429(10)404940  (01)00880304183049(17)240605(10)488640  (01)00880304183049(17)240615(10)178480  (01)00880304183049(17)240630(10)178510  (01)00880304183049(17)240813(10)811460  (01)00880304183049(17)240813(10)894400  (01)00880304183049(17)240817(10)596210  (01)00880304183049(17)241006(10)171360  (01)00880304183049(17)241006(10)506080  (01)00880304183049(17)241201(10)829310  (01)00880304183049(17)241219(10)383170  (01)00880304183049(17)250114(10)296390  (01)00880304183049(17)250122(10)454530  (01)00880304183049(17)250127(10)628920  (01)00880304183049(17)250410(10)345180  (01)00880304183049(17)250505(10)656590  (01)00880304183049(17)250512(10)724860  (01)00880304183049(17)250610(10)026040  (01)00880304183049(17)250828(10)922280  (01)00880304183049(17)250917(10)138990  (01)00880304183049(17)251002(10)138980  (01)00880304183049(17)251005(10)138970  (01)00880304183049(17)260212(10)753890  (01)00880304183049(17)260404(10)357370  (01)00880304183049(17)260405(10)225770  (01)00880304183049(17)260524(10)670210  (01)00880304183049(17)260601(10)116830  (01)00880304183049(17)260613(10)300060  (01)00880304183049(17)270718(10)052220  (01)00880304183049(17)270729(10)631830  (01)00880304183049(17)270731(10)269770  (01)00880304183049(17)270821(10)028540  (01)00880304183049(17)270821(10)440080  (01)00880304183049(17)271209(10)291090  (01)00880304183049(17)280205(10)554850  (01)00880304183049(17)280219(10)781210  (01)00880304183049(17)280223(10)781270  (01)00880304183049(17)280302(10)085270  (01)00880304183049(17)280416(10)459250  (01)00880304183049(17)280419(10)314980  (01)00887868243875(17)280221(10)123110  (01)00880304696563(17)250213(10)976060  (01)00880304722897(17)240623(10)505080  (01)00880304804258(17)251123(10)902760  (01)00880304825970(17)260520(10)158540  (01)00880304814677(17)260218(10)893790  (01)00887868237690(17)271117(10)891800  (01)00880304729339(17)250618(10)354210  (01)00887868250774(17)280513(10)580730  (01)00880304857759(17)270202(10)833190  (01)00887868238963(17)271217(10)245070  (01)00880304713987(17)250514(10)987770  (01)00880304815001(17)260223(10)101010  (01)00887868242984(17)280228(10)023240  (01)00880304856783(17)270124(10)752240  (01)00887868237713(17)271202(10)061630  (01)00880304816992(17)260329(10)464490  (01)00887868241147(17)280213(10)931330  (01)00887868231049(17)271108(10)767860  (01)00887868228247(17)271013(10)457180  (01)00880304806641(17)251209(10)108410  (01)00880304815162(17)260219(10)010560  (01)00880304762336(17)251123(10)845100  (01)00887868214868(17)270816(10)589440  (01)00880304751712(17)230124(10)242260  (01)00880304858268(17)270216(10)081730  (01)00887868163852(17)270712(10)152510  (01)00880304990234(17)270517(10)333180  (01)00880304855120(17)261206(10)232750  (01)00887868238895(17)271219(10)018170  (01)00887868247330(17)280412(10)729400  (01)00880304715028(17)250513(10)922610  (01)00880304989580(17)270503(10)012210  (01)00880304723238(17)240810(10)113270  (01)00880304753860(17)231212(10)299760  (01)00887868240805(17)280110(10)494530  (01)00880304731097(17)250803(10)615350  (01)00880304979802(17)270404(10)663460  (01)00880304849457(17)261102(10)866540  (01)00880304856073(17)261221(10)445600  (01)00887868238994(17)271202(10)987060  (01)00880304751958(17)230228(10)703660  (01)00880304752719(17)230620(10)570990  (01)00880304722514(17)240422(10)725610  (01)00880304993044(17)270621(10)865800  (01)00880304752436(17)230604(10)410790  (01)00880304721876(17)240626(10)323190  (01)00880304721869(17)240626(10)321240  (01)00887868226496(17)270909(10)062750  (01)00880304721883(17)240626(10)323220  (01)00880304728615(17)240626(10)392890  (01)00887868246852(17)280412(10)678570  (01)00880304992689(17)270607(10)589430  (01)00880304723993(17)241204(10)327840  (01)00880304706767(17)250326(10)527260  (01)00880304762077(17)251106(10)811900  (01)00880304848900(17)261014(10)781450  (01)00880304722675(17)240611(10)185340  (01)00880304852921(17)261206(10)143070  (01)00880304852907(17)261207(10)102630  (01)00880304852976(17)261207(10)161540  (01)00887868223938(17)270823(10)469610  (01)00887868227646(17)271020(10)615310  (01)00880304722774(17)240611(10)171940  (01)00880304752085(17)230325(10)877990  (01)00880304826533(17)260607(10)295510  (01)00880304716919(17)250615(10)261430  (01)00880304753471(17)231021(10)866620  (01)00880304733572(17)250827(10)020430  (01)00887868242908(17)280306(10)932330  (01)00880304808294(17)251215(10)345970  (01)00880304992238(17)270617(10)548320  (01)00887868243592(17)280208(10)783830  (01)00880304728752(17)240210(10)830690  (01)00880304723696(17)241021(10)916100  (01)00880304838635(17)260915(10)526330  (01)00880304723511(17)240916(10)404790  (01)00880304856806(17)270123(10)560130  (01)00880304723566(17)240925(10)586830  (01)00880304992467(17)270609(10)420330  (01)00880304858114(17)270213(10)072480  (01)00880304753426(17)231018(10)690080  (01)00887868241093(17)280207(10)685590  (01)00880304722569(17)240501(10)745970  (01)00880304849303(17)261018(10)864430  (01)00880304737105(17)250915(10)266940  (01)00880304689633(17)241209(10)458400  (01)00887868238468(17)271217(10)150490  (01)00880304723771(17)241027(10)021720  (01)00880304804425(17)251124(10)953500  (01)00880304857100(17)270124(10)709360  (01)00880304752764(17)230701(10)621270  (01)00880304817821(17)260411(10)589940  (01)00887868238987(17)271228(10)178950  (01)00880304753273(17)230923(10)436950  (01)00880304752306(17)230508(10)314680  (01)00887868230974(17)271109(10)801690  (01)00880304852419(17)261118(10)077370  (01)00880304709522(17)250506(10)673650  (01)00880304833616(17)260811(10)076460  (01)00887868104152(17)190801(10)989490R  (01)00887868104152(17)200401(10)444510R  (01)00887868104138(17)190515(10)856130R  (01)00887868104091(17)180801(10)375410R  (01)00887868104091(17)241119(10)524230R  (01)00887868104077(17)200401(10)444070R  (01)00887868104022(17)190201(10)684560R  (01)00887868104039(17)190515(10)804340R  (01)00887868104039(17)250212(10)770890R  (01)00887868103940(17)180915(10)483640R  (01)00887868103940(17)190615(10)946040R  (01)00887868103926(17)190115(10)640240R  (01)00887868103926(17)200301(10)443940R  (01)00887868103919(17)200401(10)443930R  (01)00887868103919(17)220228(10)340650R  (01)00887868103902(17)190115(10)601630R  (01)00887868103902(17)260527(10)186080R  (01)00887868103834(17)190615(10)946030R  (01)00887868103889(17)180615(10)118680R  (01)00887868103889(17)190515(10)889050R  (01)00887868103810(17)200301(10)443810R  (01)00887868103810(17)230318(10)692180R  (01)00887868103810(17)280326(10)002040R  (01)00887868103810(17)280326(10)002070R  (01)00887868103780(17)200301(10)443710R  (01)00887868103230(17)240618(10)589620R  (01)00887868104244(17)211201(10)550410R  (01)00887868104237(17)190501(10)856180R  (01)00880304734241(17)250827(10)073990  (01)00880304811393(17)260126(10)655330  (01)00880304753389(17)231001(10)660640  (01)00880304752979(17)230808(10)140230  (01)00880304752962(17)230808(10)121180  (01)00880304732254(17)250806(10)677860  (01)00880304752931(17)230730(10)905650  (01)00880304751842(17)231119(10)244530  (01)00880304806610(17)251211(10)090680  (01)00880304852884(17)261221(10)147270  (01)00880304817753(17)260325(10)420270  (01)00887868247347(17)280912(10)488260  (01)00880304702783(17)250309(10)210230  (01)00880304817845(17)260414(10)576820  (01)00887868244070(17)280315(10)280010  (01)00887868242915(17)280212(10)920470  (01)00880304752030(17)230325(10)848710  (01)00887868231018(17)271117(10)907020  (01)00887868239144(17)271219(10)177330  (01)00880304852969(17)261222(10)146060  (01)00880304820227(17)260425(10)779010  (01)00880304856370(17)261229(10)510010  (01)00887868214554(17)270824(10)579400  (01)00880304751255(17)230724(10)699870  (01)00880304754270(17)240917(10)404800  (01)00880304988293(17)270502(10)016170  (01)00880304914766(17)270422(10)871360  (01)00880304737259(17)251004(10)287810  (01)00880304723986(17)241202(10)275770  (01)00887868229916(17)271106(10)710300  (01)00880304705999(17)250319(10)368990  (01)00887868244179(17)280219(10)920370  (01)00880304753778(17)231127(10)153480  (01)00887868227820(17)271117(10)455550  (01)00880304724464(17)241113(10)202440  (01)00880304833852(17)260822(10)276900  (01)00880304734548(17)250827(10)051110  (01)00887868243363(17)280220(10)022 960  (01)00880304825802(17)260521(10)127670  (01)00880304723412(17)240905(10)224710  (01)00880304848139(17)261004(10)700780  (01)00887868244155(17)281106(10)099940  (01)00887868164040(17)270719(10)155730  (01)00880304733664(17)250826(10)026830  (01)00887868230882(17)271108(10)800590  (01)00880304833883(17)260829(10)352100  (01)00880304709140(17)200416(10)667280  (01)00880304989290(17)270503(10)228170  (01)00880304988897(17)270505(10)136500  (01)00880304752283(17)230503(10)233300  (01)00880304811423(17)260203(10)791170  (01)00880304714076(17)250511(10)988720  (01)00880304858275(17)270330(10)053580  (01)00880304992740(17)270608(10)751900  (01)00887868241475(17)280125(10)715790  (01)00880304852945(17)261209(10)156950  (01)00880304722477(17)240410(10)597280  (01)00880304753556(17)231113(10)005850  (01)00880304753105(17)230821(10)167610  (01)00880304753334(17)230919(10)450140  (01)00880304731226(17)250710(10)548940  (01)00887868241581(17)280205(10)919950  (01)00880304728677(17)240129(10)766990  (01)00887868242427(17)280116(10)465350  (01)00887868245664(17)280405(10)601970  (01)00880304806306(17)251202(10)001770  (01)00880304753754(17)231121(10)147130  (01)00880304734098(17)250916(10)051130  (01)00887868163289(17)270624(10)862140  (01)00887868227103(17)271001(10)083520  (01)00880304992412(17)270608(10)550620  (01)00880304826038(17)260617(10)273990  (01)00880304755024(17)230208(10)666930  (01)00880304723542(17)240922(10)586800  (01)00880304852716(17)261120(10)101920  (01)00887868231032(17)271103(10)768050  (01)00880304987067(17)270503(10)069360  (01)00880304832312(17)260809(10)925620  (01)00880304753846(17)231127(10)248270  (01)00880304855526(17)261221(10)333750  (01)00887868166259(17)270725(10)360480  (01)00887868242991(17)280213(10)874190  (01)00880304753150(17)231105(10)017680  (01)00887868214875(17)270804(10)409850  (01)00880304816466(17)260314(10)318280  (01)00887868224485(17)270823(10)793740  (01)00880304732100(17)250806(10)825130  (01)00887868244100(17)280301(10)279970  (01)00880304817685(17)260408(10)542750  (01)00887868226267(17)270912(10)946640  (01)00880304753945(17)231219(10)439000  (01)00880304828346(17)260712(10)556860  (01)00880304808942(17)260108(10)528650  (01)00880304856950(17)270118(10)667060  (01)00880304752634(17)230606(10)516880  (01)00887868240188(17)271213(10)020520  (01)00880304989719(17)270518(10)358060  (01)00887868163067(17)270617(10)829020  (01)00880304734838(17)250917(10)196320  (01)00880304756793(17)251104(10)739560  (01)00880304722507(17)240421(10)613830  (01)00880304753747(17)231121(10)147120  (01)00880304817876(17)260415(10)608620  (01)00880304752900(17)230724(10)991420  (01)00880304839359(17)260921(10)532980  (01)00880304990111(17)270517(10)286390  (01)00880304731493(17)250714(10)675780  (01)00880304834842(17)260906(10)277320  (01)00887868163241(17)270620(10)862060  (01)00887868231056(17)271103(10)752980  (01)00880304989641(17)270526(10)286450  (01)00880304756168(17)251020(10)610610  (01)00887868163418(17)270626(10)908670  (01)00880304812420(17)260202(10)746920  (01)00880304857773(17)270215(10)040510  (01)00880304723726(17)241029(10)921860  (01)00880304723450(17)240911(10)304830  (01)00880304730540(17)250706(10)500860  (01)00880304730304(17)250706(10)498850  (01)00880304731240(17)250806(10)677870  (01)00880304832336(17)210816(10)019360  (01)00887868226762(17)270920(10)158490  (01)00880304722187(17)240323(10)205300  (01)00880304732711(17)250722(10)551840  (01)00880304724501(17)241201(10)447570  (01)00880304723184(17)240727(10)880920  (01)00880304989504(17)270509(10)315770  (01)00887868166440(17)270726(10)417220  (01)00887868166457(17)270808(10)408480  (01)00880304992085(17)270524(10)385380  (01)00880304752054(17)230405(10)926110  (01)00887868226335(17)270909(10)902930  (01)00887868243615(17)280207(10)784490  (01)00880304752818(17)230710(10)663140  (01)00887868163944(17)270715(10)098820  (01)00880304993716(17)270626(10)681930  (01)00887868226212(17)270901(10)790470  (01)00887868224515(17)270816(10)720740  (01)00887868230905(17)271130(10)010420  (01)00880304752535(17)230619(10)516870  (01)00887868245718(17)280413(10)564390  (01)00880304709485(17)250427(10)665760  (01)00887868238451(17)271206(10)928780  (01)00880304723795(17)241029(10)998550  (01)00880304735835(17)250914(10)223770  (01)00880304816527(17)260317(10)330850  (01)00880304699977(17)250212(10)977630  (01)00880304699984(17)250213(10)977640  (01)00880304699960(17)250213(10)977620  (01)00880304699991(17)250212(10)977650  (01)00880304731486(17)250730(10)680350  (01)00880304855762(17)270105(10)395050  (01)00887868163593(17)270705(10)069910  (01)00880304753891(17)231212(10)331150  (01)00887868245688(17)280412(10)604630  (01)00880304735866(17)250911(10)227170  (01)00880304992436(17)270606(10)443270  (01)00880304707108(17)250327(10)602070  (01)00887868238178(17)271227(10)368810  (01)00887868224621(17)270928(10)568500  (01)00887868163562(17)270705(10)070320  (01)00880304729872(17)250617(10)236700  (01)00880304707115(17)250406(10)641650  (01)00880304814561(17)260228(10)979480  (01)00880304814578(17)210301(10)019450  (01)00880304907751(17)270402(10)666430  (01)00880304715677(17)250610(10)153040  (01)00887868243769(17)280228(10)122120  (01)00887868245787(17)280412(10)708210  (01)00880304992092(17)270602(10)386280  (01)00880304753051(17)230821(10)198460  (01)00880304700635(17)250126(10)883800  (01)00880304735767(17)250915(10)215090  (01)00880304706996(17)250406(10)602080  (01)00887868249471(17)280518(10)676790  (01)00880304808034(17)251221(10)196980  (01)00880304702820(17)250220(10)211030  (01)00880304831032(17)260802(10)889870  (01)00880304722385(17)240507(10)745960  (01)00880304752146(17)230405(10)936850  (01)00880304813670(17)260202(10)769750  (01)00880304728080(17)241219(10)476890  (01)00887868241000(17)280202(10)696740",
      "recall_initiation_date": "20190821",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}