      "status": "Terminated",
      "city": "Southbridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66862",
      "recalling_firm": "Karl Storz Endoscopy America Inc",
      "address_1": "15 Wells St",
      "address_2": "N/A",
      "postal_code": "01550-4503",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution.",
      "recall_number": "Z-0526-2014",
      "product_description": "Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2.    \" Transoral intubation \" Transnasal intubation \" Evaluation of placement of endotracheal and double lumen endotracheal tubes \" Provide visual access to the larynx and tracheobronchial tree.",
      "product_quantity": "1516 units",
      "reason_for_recall": "KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.",
      "recall_initiation_date": "20131113",
      "center_classification_date": "20131217",
      "termination_date": "20140327",
      "report_date": "20131225",
      "code_info": "C1140, C1140E, C1160, C1160E, QKC1750, QKC1750E, QKC1694, QKC1694E"
    }
  ]
}