  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Golden",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86047",
      "recalling_firm": "CME America, LLC",
      "address_1": "14998 W 6th Ave Ste 830",
      "address_2": "N/A",
      "postal_code": "80401-5025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.",
      "recall_number": "Z-0525-2021",
      "product_description": "CME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).",
      "product_quantity": "Total Pumps = 4,182 (all models)",
      "reason_for_recall": "Multiple reasons:  1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated.  2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.",
      "recall_initiation_date": "20200730",
      "center_classification_date": "20201203",
      "termination_date": "20220308",
      "report_date": "20201209",
      "code_info": "Model TR34L - Part Number(s)  100-100PSLA; 100-101PSLA  Serial Number:  58766  58765  26479  69547  69548  26428  26299  26301  26295  26294  26304  26296  26298  26302  26297  26300  26130  26133  26132  58906  58907  58909  58910  58911  58912  58913  58914  58915  58916  58917  58918  58919  58920  58921  58922  58923  58924  58925  58926  58998  59003  59007  58999  59006  59000  59004  59002  59005  59001  69618  69619  69628  69629  69630  69637  69659  69666  69638  69579  70119  70134  70136  70142  70144  70145  70146  70147  70148  70162  70167  70169  70172  70173  70174  70175  70143  S25610  S25720  S25725  S25729  S69776  S69785  S70157  S70160  S70165  S70176  S70192  S70203  S70206  S70207  S70208  S70209  S70210  S70211  S70129  S70177  S70181  S70205  S70114  S70115  S70117  S70118  S70120  S70121  S70124  S70125  S70126  S70127  S70128  S70130  S70131  S70132  S70138  S70139  S70140  S70163  S70164  S70166  S70182  S70183  S70185  S70187  S70188  S70190  S70191  S70194  S70195  S70196  A12200  A12201  S69760  A100795  A100796  A100797  A100798  A100799  A100800  A100801  A100802  A100803      UDI Code:  B101100100PSLA0  100-101 PSLA was retired prior to UDI",
      "more_code_info": ""
    }
  ]
}