  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78677",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-0524-2018",
      "product_description": "Syngo.plaza systems with SW VB20A, Model Number - 10863171, 10863172, 10863173    Product Usage:  Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.",
      "product_quantity": "76",
      "reason_for_recall": "When a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other workflow steps. The other workflow steps will continue to show the initially loaded study.",
      "recall_initiation_date": "20170823",
      "center_classification_date": "20180205",
      "termination_date": "20181001",
      "report_date": "20180214",
      "code_info": "Model Number - 10863171, 10863172, 10863173"
    }
  ]
}