  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.",
      "address_2": "",
      "product_quantity": "219988 units",
      "code_info": "All lots manufactured between January 2008 - May 2019:    064270  080270  128980  137290  143730  153340  159530  167310  167350  178160  202110  204480  215370  217060  222160  222300  252860  267090  276540  276620  307580  308360  314480  343540  350100  356980  367640  384320  422970  427440  454950  497760  512310  515990  534870  549000  551140  558390  563890  574720  607950  616850  627550  627670  630310  636470  643170  653820  673480  685260  694200  699690  715590  716060  752360  762320  775690  775740  799160  799210  812890  822070  832770  832990  833090  833800  835910  836580  851290  880580  892250  914760  917410  924500  925720  929750  929920  951620  953380  957790  960130  960370  968400  969900  004360  023060  043830  045520  051210  055590  064990  095300  124710  133900  149180  160150  161960  167340  181730  187990  190220  211220  252870  268490  274460  281780  330640  349640  351240  351930  352140  356860  358350  363840  470270  474740  474930  491250  491500  520870  522390  523010  529290  535000  539710  571640  610480  619900  623270  688980  698110  699300  740680  752370  752440  777920  786080  789370  792540  799710  823190  870400  892330  895480  917430  924470  927880  929680  952270  956910  960110  011300  012720  019650  051200  052410  058980  066080  123760  134630  143340  167520  173700  185670  191120  207860  260270  268110  324460  347380  357400  376030  406600  423030  454990  477630  485320  496850  497960  529740  548340  587230  593200  636020  636500  643150  655850  686280  703180  723850  743140  775680  792530  825560  827640  830750  895670  907690  953070  071930  076970  081800  102960  167760  216010  268400  322260  351880  356850  357070  416780  474720  480850  519470  522360  558810  568890  618000  629320  650880  674130  688180  694130  697660  723950  769670  787810  845930  850940  851720  852020  908150  929720  956850  968430  010060  092230  114350  129700  200490  206190  218850  263300  290500  369280  422940  480830  488250  558990  562180  567600  580040  643110  692400  701250  762020  812790  830960  839300  849950  853000  856740  943120  967840  039610  098550  100120  148100  176460  229820  302910  347590  367510  529250  562210  569080  584790  617620  629560  658740  670090  674470  723890  751930  753690  799660  800670  840540  848450  849960  895470  929670  993440  004500  005830  006420  017720  029970  071450  072240  119200  125230  197520  206830  279080  378700  408300  416760  441990  470220  546100  546130  549240  568320  576640  658720  746070  781590  804070  818500  872060  914220  993830  997450  026630  035380  048830  087040  109920  134610  176820  178470  186220  186280  218860  339680  431080  434970  555280  561530  563900  592740  596280  611560  688840  688860  698540  703820  716070  740690  751270  753440  801740  849910  865390  891360  895400  895800  914680  952280  998880  026830  175640  179010  333650  409120  427110  431110  463440  487960  517180  520860  520880  611620  615840  762160  777740  866140  876870  890010  896900  917240  924420  026770  234980  366070  366220  367610  367820  410080  457030  520910  522070  549760  564330  629370  673500  683820  692420  738210  809840  812370  849970  901110  919090  923920  014050  026760  082270  109930  114700  135390  173830  228160  234970  295470  311300  333640  366990  409080  510020  546140  549320  593740  617640  662930  684880  692440  721040  746090  845940  887860  895520  943820  962830  020690  098320  235000  409060  458630  496890  540020  543800  554210  554230  634010  654010  688830  688910  738280  801610  818510  895780  975850  013350  024580  026670  048890  055240  074450  094220  112030  117160  123950  128950  149270  168760  174320  178960  181550  209250  216860  231370  262290  266540  272890  296650  301140  314730  347060  347480  349290  351000  383330  383340  383360  383370  383880  388040  398290  406510  408000  423040  426080  426990  428200  452640  455670  459460  459620  460190  463510  478430  478450  493610  499820  501120  502520  508750  525100  543530  565750  584800  596890  611630  630290  631130  645630  669860  674340  681930  684290  691380  700700  705510  706540  731080  734480  742580  752450  755340  758280  788330  833930  907990  909490  909890  924200  924750  930340  930500  949700  952220  962840  963750  994010  995550  095380  823830  584570  011170  022010  035430  040980  071060  072150  079190  082420  082580  082740  087290  100740  131400  136990  140070  167330  174750  181720  186110  204560  204650  219300  239570  243110  255410  268540  273830  295070  314860  317680  360230  382790  388790  401460  404210  405530  423050  429800  437840  460470  465210  472530  487710  499800  499810  512680  512930  514470  521780  534970  559820  570600  590190  631170  698650  701910  701930  748400  769380  801090  803150  846170  857920  874900  905640  906760  915810  924970  953870  954700  964490  974950  983960  99392020940  027080  035330  038150  082400  148800  150470  153820  223910  244850  257330  268550  318180  337550  427000  432670  445070  464260  477630  478440  487780  499840  634900  634940  499910  500790  518770  569160  574890  578360  582510  604920  607860  610090  706720  710850  755630  763590  770980  812790  818140  822440  827080  854770  865790  887710  918370  960300  057740  173070  670740  090090  102330  157180  167160  174270  209140  229410  240020  257340  263850  330530  330720  384010  388740  420610  433210  478420  514240  517760  517790  524620  531420  538740  582150  591390  596880  680560  731270  732010  744920  762230  795950  810580  851380  856550  876730  887700  910550  918830  941420  968000  350780  4100  026580  056240  102350  131770  153520  205880  209830  249870  364690  380610  383530  411980  466440  524320  524580  524600  530450  543580  576850  610170  610200  715380  731450  799290  830810  831150  915100  952190  101890  176800  182230  197230  207400  239520  244030  270190  311470  321590  404220  420730  475210  491150  538140  576670  628470  674380  685990  700890  737990  753480  833840  864610  927240  940230  967360  974930  450760  025540  044070  061800  191040  249930  257360  274520  356970  388030  426960  427010  433060  437970  463490  475230  493860  524340  559090  559730  577170  584580  617780  630010  656430  698390  705550  711870  732090  741580  770770  810330  811010  954650  963670  994170  042060  071000  158810  209850  254620  263800  364710  370400  432690  432700  433160  454100  481430  524170  524180  524370  559080  560200  610150  637230  762190  777340  861940  871980  960080  961450  843000  670720  874810  055720  061870  070970  086430  164300  172600  172610  182200  208380  292660  295490  314440  408020  432710  433220  440110  474820  514450  518600  530310  550040  577260  588140  588200  594820  610160  634430  641240  684520  731550  779270  803010  824650  863570  924100  924450  014310  037970  061660  061700  081040  112040  130280  153780  237420  250450  256440  256950  256970  383830  454060  530440  550000  550020  559100  628510  663820  680510  739490  797600  838980  918820  924350  924410  173110  035570  038040  038180  042090  050500  070990  109780  277060  283310  340220  381500  415750  420560  436710  454050  465180  469460  487100  524390  524570  530460  559170  560150  610100  610140  628610  680730  681240  791210  856590  882070  882140  926740  967280  981250  061680  061770  082300  118730  130260  182220  191100  212300  277070  283240  283300  295450  295480  364610  432680  508630  524130  524330  524530  550030  586050  596790  596820  610070  610080  628420  628520  699640  699650  712670  721170  798590  846180  863580  925920  927090  941380  941480  961430  966830  995690  096720  109880  152490  162200  234910  234930  268580  310540  374490  408550  441920  456760  474800  499070  561400  591450  650610  688880  692390  763700  771950  801540  901050  933470  944050  960180  038250  042070  061620  061710  061860  071610  128170  130310  215150  244020  250830  256940  283260  283280  295500  307360  307410  310220  349160  351660  351670  359660  364680  426940  433100  433120  436750  469080  481610  504500  524140  524430  524590  524610  524660  532550  549960  559150  583270  610260  610270  628460  628530  641230  720980  797620  799020  804180  811180  822430  869180  881340  901140  922740  924330  924390  924430  926910  945040  960090  961420  973650  981210  026800  056170  074580  135370  139150  169250  306700  351980  367000  409130  410070  431140  457490  458620  497010  499250  558950  591490  593720  676560  688850  692380  692450  696830  704110  838490  849890  866620  866730  895450  901060  901070  914250  924120  943080  026570  028880  038030  038100  038170  044490  070960  071580  071650  082290  092770  172620  182190  208330  208360  209810  243990  250430  250470  256960  263760  277050  283180  283220  293100  295470  307390  310200  310250  310270  319980  321550  321560  321600  334330  334350  369540  381490  405160  426910  433150  004270  026780  026820  109890  109900  129110  152680  158010  206750  234920  275830  281770  281790  281810  336700  408010  408750  465450  496950  496980  549300  561360  561380  561390  561500  561520  591460  639240  639310  688890  728820  745940  787630  843590  850040  857590  857610  914200  919070  943830  959410  964900  026810  109910  135350  135360  135380  135400  135410  135420  135430  137390  146000  146020  148140  204640  234940  234950  234960  234990  259760  281800  295460  349580  408590  454200  456900  458640  458650  458660  460070  487050  487070  496940  520850  520890  520900  520920  520930  561410  561510  561540  566470  591470  591480  592890  593760  606320  688810  688870  692360  692370  692410  692430  746130  762310  807600  838460  849900  849930  849940  850000  850030  850050  850060  887620  902300  914210  914230  914240  914260  914270  914280  933440  933480  933490  933500  943100  943110  943140  943840  943850  943860  107230  107270  215430  295450  349560  349610  352130  498980  517170  696930  738200  746050  804400  867230  913230  925270  997880  020740  075040  107290  107310  109940  127930  151280  151320  151380  152400  226850  446650  546090  567680  597960  636040  643250  746040  788180  818440  867220  895420  926900  943960  004290  004510  014060  045540  066690  107260  156810  227010  302900  324360  351960  371600  377760  431100  456710  522090  560850  610690  656120  674040  696920  702780  723880  746120  746150  746160  783300  848550  004470  075050  111120  125290  129160  141120  145970  148120  148130  151540  151670  152330  152630  153040  259790  357270  374140  408370  494690  532210  561880  568310  611510  673870  696980  702730  788270  801530  845960  845970  866680  892200  895680  909610  983920  004760  014040  111140  112280  118120  141110  146030  152260  152540  152620  177590  206600  206800  209980  215440  229940  281190  306770  314070  314080  349460  352150  374120  387150  433420  456880  494730  494740  549260  560860  561890  633990  659260  675320  675350  675360  675450  701390  701840  731020  746110  771980  777100  784940  801660  848610  865190  866760  909590  995210  004300  015240  041400  082030  091530  092290  107280  107490  118070  118090  125260  151240  151350  170830  175710  206760  219830  234640  315570  315780  349510  351900  371570  377790  424980  433380  499320  568370  592920  593750  610680  612650  632280  639200  746060  761990  787900  838520  845980  876340  914620  938650  984280  004340  004400  047030  075060  080260  107510  129070  151310  157980  162150  171570  177790  187540  209650  214980  215000  264390  319480  352120  357330  408380  408440  408500  408690  431020  456820  457040  457290  500280  500390  514990  517220  532190  571390  571410  593730  610970  658660  676550  701340  707420  740080  745910  745990  761970  801480  895810  895820  914740  938620  953690  977310  994370  995510  996430  004600  047010  075090  092240  092330  111130  129050  142260  145960  151860  152550  156780  170790  177490  177640  177730  206610  206620  206640  206680  215480  229330  229400  229870  281200  287300  302940  302970  306650  327640  349500  349550  349570  444750  446620  457010  457020  457220  457470  487080  489140  494720  499290  517450  569060  610950  638130  639370  675400  740850  745970  746000  746020  746030  762030  763610  772000  787860  787920  790550  828520  848530  876310  876360  901040  901090  907920  914630  948580  951230  978190  980510  984160  984240  984260  004330  014070  019090  020640  020750  092300  092310  092350  129120  141040  142240  145950  148170  150800  151270  151510  152420  160950  161060  162180  206650  206690  226910  230120  230190  234670  239540  264430  271360  281220  305630  314090  314240  314260  336620  336630  336640  336650  336660  349400  349420  349520  374170  392800  392840  392970  393010  393020  408660  422300  444710  444740  444990  445010  456730  456780  456790  456830  456930  457200  457210  457500  457530  512210  522110  527180  532120  532230  532240  546110  560840  563290  567670  576840  610930  616830  628870  633970  633980  634000  639220  639410  675430  702770  738230  745950  745960  761980  762060  763690  787600  788100  801600  801760  801770  838450  843620  845910  848540  865120  876300  876350  876380  907890  909540  914730  948640  959400  984290  004310  004390  004410  004590  004610  004650  047020  047110  075080  075100  092320  107250  111160  125180  125310  125370  125420  129180  141010  141060  145990  151340  151370  151710  151870  152560  153600  153610  153620  160920  162170  171550  177660  180720  180750  209620  210000  229380  229420  229930  229950  230040  234650  259800  264440  264460  271630  271850  305560  306640  306710  306760  310560  314110  336600  349410  349430  349540  349620  352030  352060  352200  357460  392850  392860  393040  408420  408430  408480  408710  422260  433400  444730  444970  445000  445020  456720  456810  457480  457520  490850  499050  499080  517420  532100  543840  563250  563320  567610  567630  569070  576670  592900  592930  606290  610720  612610  617310  639140  643080  658620  673850  676520  701800  738170  740810  745770  745930  763600  763670  787980  788030  788320  801490  801500  801570  845950  848520  848580  848600  848630  866710  895430  895750  901120  902320  905550  905570  907880  907930  913200  948590  948620  948630  953610  953670  959390  984120  984140  996380  004420  004480  004490  004630  004660  020610  020660  022940  047050  047090  075070  075110  087450  092840  105230  107300  107320  107330  107500  107520  107540  118080  125130  125320  125390  129140  141020  142140  142200  146040  148110  148150  148190  150680  150780  150790  151050  151250  151290  151300  151490  151810  152530  153060  162120  170810  170840  171510  177460  177540  177740  177760  177820  180730  180740  180760  206570  206580  206790  209970  229410  229440  229830  229850  230010  230060  230100  230110  234600  264420  264470  264480  271300  271380  271650  306690  306730  306740  310570  314060  314100  314220  314230  314250  314270  314280  315770  327650  327680  327690  336720  336730  342820  349380  351910  352010  352040  352170  357280  357310  392790  392810  408390  408610  408730  422230  435000  444700  444850  445030  456740  456910  456960  456970  456980  457050  457300  486230  489170  489180  489200  490860  499010  499060  499310  500330  500470  517210  518090  522100  527400  532170  532200  543830  549280  563330  567620  568340  569090  571370  576600  576660  576730  577590  588260  596400  601650  601880  606170  606260  606310  606330  606350  610960  611570  611590  612640  612660  617540  638060  638070  638080  638100  639430  639450  639460  643220  643230  673830  677860  683200  698510  715490  715510  715550  724670  727280  740820  740830  740840  740860  740870  740880  745920  762000  762040  763590  763660  763680  763710  763730  788350  801550  801590  801620  801650  801720  801750  811010  812440  818480  818550  818570  838480  838510  845920  848510  866670  866690  866720  866750  870140  870220  876320  876400  895700  901100  905560  914770  944090  948610  951220  953630  953640  953680  984170  984270  994130  995200  996390  996420  004260  004350  004380  004620  014030  019180  019190  020620  020730  047040  047100  054050  060760  060770  060780  092220  092270  092280  092340  092900  105270  107210  107240  107550  129040  129130  129150  129170  141050  141130  142170  144110  145980  146010  146050  146080  148160  148180  151020  151060  151220  151530  151820  151910  153090  161000  161040  161050  162140  162210  171520  171560  171580  177470  177500  177510  177550  177610  180770  180780  180790  180800  206630  206670  206700  206810  209940  209950  209960  209990  210010  210020  215460  215470  215490  215500  229350  229390  229430  229810  229840  229980  271280  271310  271330  271340  271370  271410  271440  271450  271470  271490  271510  271570  271620  271860  271880  302960  306660  306780  310500  314120  327630  335880  336680  336690  336710  349370  349450  349530  349630  351890  352000  352160  357320  357340  377130  392950  392980  408450  408580  422240  422280  431010  456800  456920  456940  456950  457250  457270  457310  489130  489160  493790  500290  500360  500400  500410  500480  500490  500510  500520  500530  514550  517190  517200  532110  532140  532150  532160  546080  563230  563260  563270  563300  563310  563340  569040  576580  576610  576620  576630  576680  576800  576810  576830  576860  592910  592950  593810  600350  606160  606210  606240  606250  606270  606280  606340  610670  610710  610730  611530  611540  611580  611600  611610  611630  611640  612070  612410  616820  617290  618330  628880  636000  638040  638050  638110  638170  638180  639290  639360  639420  642970  643240  659250  673860  673890  675380  676500  676540  676880  701830  701850  701860  701870  701880  702830  703160  715500  715540  715620  727240  727270  731010  731030  738160  738190  738240  740100  740110  740620  740650  740660  749740  762070  777050  787950  790080  790130  798150  801510  801520  801560  801640  801730  801780  801820  801830  801840  810980  812320  813460  818530  818560  820650  820810  838470  845880  866700  876330  876390  895380  901130  901140  905540  907910  909560  909600  914640  914650  914660  914720  914750  938670  953600  953650  953660  953700  959370  959380  959420  959430  975910  977440  984100  984110  994380  995190  996400  000430  027120  040250  040300  041160  118170  151720  170300  187480  201560  207440  258310  291010  303000  319540  357430  357970  371620  371660  424910  424950  435010  435260  486140  494710  517070  517410  519860  596870  612060  628930  629760  696940  697020  707400  707450  707470  721000  721020  739580  807630  820620  843610  874540  892120  902360  902990  939200  945250  017520  017560  017580  040220  066670  087380  118140  148750  151700  165530  166690  170320  170360  190200  190270  201500  209590  215010  226820  234610  239400  239480  239520  284070  302860  322610  342490  374160  388570  424840  424890  431040  431060  431170  446640  452300  487060  490790  490870  527090  563920  571380  577580  597640  628940  629780  632320  650580  650630  676850  683140  683190  696870  697110  729110  739590  778600  778610  782280  798290  828500  838360  838370  863290  870210  874250  877600  887580  913240  924700  924760  938570  939170  945210  951180  960040  967280  980430  983880  983950  017550  017620  025540  040190  040240  040270  041360  041420  041620  066740  082000  082010  082020  082040  082050  082060  082070  082080  094120  095720  095730  112270  118150  125410  148740  148770  148780  148790  148800  148810  151660  151730  153080  158000  158050  158070  165540  166650  166660  166670  170370  175730  175790  175800  184810  187520  190250  190260  201530  207450  214990  219800  219820  219840  219850  219860  219880  226900  227020  234660  234680  234690  239510  239550  258340  258360  258370  287310  290990  291030  302780  302850  302990  315550  315560  319460  322600  342420  342440  357470  357480  357920  357960  371580  371630  374150  377060  377740  388500  388510  388520  415120  415170  424830  424850  424860  424880  431130  433410  433430  434950  434960  435170  435200  435220  435230  435250  446670  483840  487040  517430  517440  519820  519870  527080  527110  527380  527440  543770  543780  543810  549290  549750  549790  549820  554200  554220  557600  563660  564120  564130  569420  577570  588280  595690  612080  612090  617550  628330  628370  628960  629450  629460  629690  629730  629750  629790  629810  656100  676910  683120  683150  683650  683700  683720  696820  696840  696860  696970  697050  697120  697160  698530  707410  707460  720790  720980  720990  721010  722650  724630  724640  724680  728830  728840  731040  777040  778590  778670  778690  783310  790110  790170  801380  801420  807640  820580  820670  820680  828480  865130  870120  870170  870200  877520  887630  892180  909200  913220  913310  938680  939120  939130  945240  945260  945290  951190  951200  951210  951240  951250  951260  959900  960010  960020  972830  976950  983890  983900  983910  983930  993620  993650  000360  000370  000380  000390  000400  000420  000440  017510  017530  017540  017570  017590  017600  017610  017630  038830  038840  038850  038860  038870  038880  038890  038900  040200  040230  040310  041170  041180  041190  041350  041380  041410  041430  041440  041450  041610  041630  041640  041650  053980  053990  054000  054010  054020  054030  054040  054060  056150  066680  066700  066710  066720  066730  066750  081020  087390  094110  094130  095770  096730  096740  096750  096760  096770  096780  100730  112250  112260  118100  118110  118130  118160  144090  144100  144120  148760  151650  151690  152250  152270  152280  152290  152300  152310  152320  152340  153030  153050  153070  153100  153110  156790  156800  157990  158020  158030  158040  158060  158080  158090  164260  164270  166680  166700  170310  170330  170340  170350  170380  175720  175740  175750  175760  175780  175810  184750  184760  184770  184780  184790  184800  184820  187460  187470  187490  187500  187510  187530  190170  190190  190280  201510  201520  201540  201550  209560  209600  209610  209630  209640  209670  209690  209700  209710  209720  209730  219780  219790  219810  219870  226810  226830  226840  226860  226870  226880  226890  226920  227000  234620  234630  234750  239470  239490  239500  239530  239560  239570  246490  246500  258320  258330  258350  258380  258390  258400  259770  259780  284030  284040  284050  284060  284080  284090  284100  287320  287330  287340  287350  287360  287370  287380  290970  290980  291000  291020  291040  291050  291070  302810  302820  302840  302880  302930  302980  305540  305550  305580  305590  305600  305610  305640  310470  310480  310490  310510  310520  310530  310550  310580  310590  315580  315760  315790  315800  319470  319490  322620  322630  322640  322660  322670  342430  342450  342460  342470  342480  342760  342770  342780  342790  342800  342810  342830  342840  342850  357410  357420  357440  357450  357930  357940  371590  371610  371640  371670  374110  374130  377050  377070  377080  377090  377100  377110  377120  377730  377750  377770  377780  377800  377810  377820  388530  388540  388550  388560  388580  388590  415070  415080  415090  415140  415160  422800  422810  422820  424870  424900  424920  424930  424940  424960  424970  424990  425000  425010  425030  425040  425050  425060  425070  427080  427090  427100  431030  431050  431070  431090  431120  431150  431160  434980  435150  435160  435180  435190  435210  435270  435290  446590  446600  446610  446630  446660  447580  453650  453660  453670  453680  453690  453700  453710  453720  453730  483830  483850  483870  483890  483910  483920  483930  486150  486160  486170  486180  486190  486200  486210  486220  490800  490810  490820  490830  490840  490880  493800  493810  493820  494670  494680  494700  498970  498990  499000  499020  519790  519800  519810  527100  527120  527130  527150  527160  527170  527190  527350  527370  527410  527420  527450  543790  543820  549250  549270  549310  549720  549730  549740  549770  549800  549810  554240  554250  554260  554270  554280  557610  557630  563630  563680  563700  563800  563840  563870  569410  577550  577600  577610  577620  577630  588270  588290  617490  617500  617510  617520  617530  617560  617570  617580  617590  618290  618310  628280  628290  628310  628340  628350  628360  628380  628860  628890  628900  628910  628950  629720  632290  632310  635960  635970  635990  636010  650570  650590  650600  650620  650640  650650  656050  656060  656070  656090  656130  656140  656150  656160  659230  659240  659270  676820  676830  676840  676860  676870  676890  676900  676920  676930  683130  683160  683170  683180  683210  683660  683670  683680  683690  683710  696800  696810  696850  696890  696910  696950  696960  697000  697010  697030  697070  697080  697100  697130  697150  698520  701300  701310  701320  701330  701350  701360  701370  701380  707430  707440  707480  720770  720780  720800  721030  721050  721060  721070  721080  721090  722520  722530  722540  722550  722560  722570  722580  722590  722600  722610  722620  722630  722640  722660  722670  722680  724600  724610  724620  724650  728810  731050  739600  739610  749700  749710  749720  749730  749750  771960  771970  771990  777030  777060  777070  777080  777090  777110  778620  778630  778650  778660  778680  782270  782290  783220  783230  783240  783250  783260  783270  783280  783290  783320  790090  790100  790120  790140  790150  790160  790180  790190  790530  790540  790560  793480  793530  796410  798070  801350  801360  801370  801390  801400  801410  801440  807580  807590  807610  807650  820600  820610  820630  820640  820690  820710  820720  820750  820780  828450  828460  828470  828490  828510  838280  838300  838320  838330  838340  838350  838380  838390  843640  863220  863230  863240  863250  863260  863270  863300  865140  865150  865160  865170  865180  865200  865210  865220  870010  870150  870160  870180  870420  874050  874090  874320  877140  877410  887560  887570  887590  887600  887610  887640  887650  892130  892140  892150  892160  892190  892210  902310  902330  902340  902350  902970  902980  909210  913210  913260  913270  913280  913290  924670  924690  924710  924730  924740  924780  924790  924810  926340  926840  927410  938590  938600  938630  938640  939090  939100  939250  939270  944060  944070  944080  945160  945180  945190  945200  945220  945230  945270  945280  959890  959940  960000  960030  967260  967270  972820  972840  972850  975190  975820  975830  975840  975880  975890  975940  976710  976790  976840  977180  977390  978180  978200  978210  978220  980450  980470  980480  980500  980540  980550  980580  980620  980630  983940  983960  993570  993590  993610  993640  993660  993670  993690  993700  678310  151030  004660R  148160R  171520R  239540R  349510R  388590R  577630R  639430R  801530R  801660R  838460R  616860  454030  454090  454110  454160  469100  469140  481570  498430  508570  517770  517800  524110  524120  524240  524380  524440  524460  524520  524540  524550  524630  524640  537120  548140  549940  559110  559120  559140  560100  560170  575960  576450  591420  596800  599260  628580  628620  628630  630470  634310  660180  699580  699670  704860  715450  791240  797580  797590  797640  799060  799140  807840  822560  830750  830780  830820  831260  831270  832600  846210  856600  866600  869200  879320  880990  881430  881780  882080  882160  896940  918770  924440  926990  941360  941540  960100  961490  973680  973850  981230  670730  005780  026590  026600  026610  026620  026640  026650  028850  028860  028870  028890  028910  028920  037980  038020  038060  038140  038160  038200  038240  042100  042110  065840  065850  069020  069030  069040  069050  069060  069070  069080  069090  069100  069110  069120  070930  070940  070980  071510  071520  071530  071550  071560  071570  071590  071600  071620  071630  071640  082260  082280  082310  082320  092760  092780  092790  094500  094510  094520  094530  094540  094550  094560  094570  103570  103580  109750  109760  109770  109790  109800  130230  130240  130250  130270  131610  131620  131640  131650  131670  131680  139710  139720  172630  172640  172650  177740  177750  177760  182180  182210  182240  182250  182260  182270  182280  208320  208340  208350  208370  208390  208400  209790  209800  209820  209840  215030  215040  215050  215060  215070  215080  215090  215100  215110  215120  215130  215160  215170  236560  239620  239970  239990  243950  243960  243970  243980  244000  244040  244050  244060  244070  250440  250840  256920  256930  257320  257350  263770  263780  263790  263810  263820  263830  263840  272120  282440  283100  283110  283170  283200  283230  283270  292630  292670  293080  293090  295460  307370  307380  307400  310210  310230  310240  310260  313860  319880  319890  319900  319910  319920  319930  319950  319960  319970  319990  321510  321520  321530  321540  321570  321580  321610  321620  321630  334320  334340  340650  340670  340680  351680  351690  359680  359690  359700  359710  359870  359880  359890  359900  359910  362440  362450  362460  364730  370130  370140  372760  381270  381280  381290  381300  381480  381520  400840  404400  404410  405100  405110  405120  405130  405140  405150  420550  420570  420580  420590  420600  420700  422710  426880  426920  426950  426970  426980  433070  433110  433130  433140  433170  433180  433190  433200  433230  433240  433250  445780  445790  445800  445810  445820  445830  454040  454070  454080  454120  454130  454140  454170  459330  459340  469070  469090  469110  469120  469130  469150  469160  469170  469190  481580  481590  481600  498390  498400  498410  498420  498440  504430  504440  504450  504460  504470  504490  504530  504540  504790  504800  508580  508600  508620  517780  517810  517820  517840  522070  524150  524510  524560  524650  530430  532480  532490  532560  535980  535990  536000  536020  536040  548050  548190  548220  549950  549970  549990  550010  550050  553580  559070  559130  559160  560110  560120  560130  560140  560160  560180  560190  560210  560220  575950  575970  575980  575990  576440  577240  577250  583220  583230  583250  583260  583280  583290  588170  588270  596810  596830  596840  596850  599250  605380  610120  610180  610190  610240  615920  628480  628500  628590  628600  628650  628670  629640  629650  629670  629680  629690  630020  636510  641260  641270  641280  641290  653470  660190  660200  660210  660220  660230  660240  662030  662040  662050  667310  667320  667330  680570  680580  680590  685970  685980  686000  686010  689320  689340  699590  699620  699660  699680  702110  704840  704850  704870  704880  708090  708120  720890  720900  720910  721090  721100  721120  721130  721140  721150  721160  731440  751750  751760  751770  751780  751790  751800  751810  751820  777330  780260  791160  791170  791180  791200  791230  796240  797630  798580  798600  798990  799000  799010  799030  799040  799050  799070  799080  799090  799100  799110  799120  799130  799150  799160  803020  803030  803190  803280  806810  806820  807830  807850  807860  807870  811030  811190  811200  811210  811240  811260  830730  830830  830840  832610  835630  848330  848340  848350  848360  848370  849360  849370  849380  849390  856560  856570  856580  856620  863350  863360  863370  863560  863590  863600  866580  866610  866620  866630  869160  869170  869190  880740  881130  881220  881290  881350  881910  882130  895250  895260  896960  912160  912170  912180  912190  912200  912210",
      "center_classification_date": "20191122",
      "distribution_pattern": "US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.",
      "state": "IN",
      "product_description": "COMP RVRS Shoulder Baseplates, including Custom Products, Item numbers 115331S  115331  115330S  115330  110040610  110040620  010000589  CP561861  PM0001456  PM551047  PM551063  PM555137  PM555160  PM555209  PM555322  PM555406  PM555415    UDI:    (01)00880304475342(17)211128(10)704850  (01)00880304475342(17)220128(10)094500  (01)00880304475342(17)220328(10)856620  (01)00880304475342(17)220528(10)487710  (01)00880304475342(17)220628(10)028860  (01)00880304475342(17)220628(10)028870  (01)00880304475342(17)220628(10)799030  (01)00880304475342(17)220628(10)799070  (01)00880304475342(17)220628(10)929900  (01)00880304532465(17)220628(10)850060  (01)00880304475342(17)220628(10)803010  (01)00880304532465(17)220728(10)026780  (01)00880304475342(17)220828(10)208370  (01)00880304475342(17)220901(10)811210R  (01)00880304475342(17)221028(10)960160  (01)00880304532465(17)221028(10)943820  (01)00880304475342(17)240616(10)237420  (01)00880304475342(17)230909(10)500790  (01)00880304475342(17)221128(10)469390  (01)00880304475342(17)221128(10)864610  (01)00880304532465(17)221128(10)234970  (01)00880304475342(17)221204(10)038140  (01)00880304475342(17)221212(10)630530  (01)00880304475342(17)221212(10)508750  (01)00880304475342(17)221212(10)769380  (01)00880304475342(17)221213(10)801090  (01)00880304475342(17)221213(10)831150  (01)00880304475342(17)221217(10)812790  (01)00880304475342(17)221218(10)543530  (01)00880304475342(17)221219(10)630010  (01)00880304532465(17)221228(10)458650  (01)00880304532465(17)221228(10)752440  (01)00880304532465(17)221228(10)786080  (01)00880304475342(17)230103(10)913540  (01)00880304475342(17)230103(10)628460  (01)00880304475342(17)230103(10)250450  (01)00880304475342(17)230104(10)913530  (01)00880304475342(17)230104(10)994010  (01)00880304475342(17)230107(10)591420  (01)00880304475342(17)230107(10)913550  (01)00880304475342(17)230108(10)732010  (01)00880304475342(17)230111(10)913570  (01)00880304475342(17)230111(10)035570  (01)00880304475342(17)230114(10)918370  (01)00880304475342(17)230117(10)042060  (01)00880304475342(17)230118(10)042100  (01)00880304475342(17)230118(10)014090  (01)00880304475342(17)230121(10)014110  (01)00880304475342(17)230124(10)081050  (01)00880304475342(17)230128(10)967360  (01)00880304532465(17)230128(10)014050  (01)00880304475342(17)230129(10)967280  (01)00880304475342(17)230204(10)913560  (01)00880304475342(17)230205(10)460190  (01)00880304475342(17)230205(10)207400  (01)00880304475342(17)230206(10)081040  (01)00880304532465(17)230207(10)014060  (01)00880304475342(17)230207(10)081030  (01)00880304532465(17)230207(10)843610  (01)00880304532465(17)230207(10)319470  (01)00880304475342(17)230207(10)081060  (01)00880304475342(17)230207(10)274520  (01)00880304475342(17)230208(10)165560  (01)00880304532465(17)230208(10)165530  (01)00880304475342(17)230211(10)165570  (01)00880304532465(17)230211(10)207450  (01)00880304532465(17)230211(10)520910  (01)00880304532465(17)230214(10)112280  (01)00880304475342(17)230214(10)460470  (01)00880304532465(17)230214(10)555280  (01)00880304532465(17)230214(10)173830  (01)00880304532465(17)230215(10)014040  (01)00880304532465(17)230215(10)387150  (01)00880304475342(17)230215(10)165550  (01)00880304475342(17)230215(10)207430  (01)00880304475342(17)230215(10)319510  (01)00880304532465(17)230215(10)512310  (01)00880304532465(17)230215(10)124710  (01)00880304532465(17)230215(10)274460  (01)00880304532465(17)230215(10)322260  (01)00880304532465(17)230215(10)562180  (01)00880304532465(17)230215(10)006420  (01)00880304532465(17)230218(10)319480  (01)00880304475342(17)230218(10)433460  (01)00880304532465(17)230219(10)433400  (01)00880304475342(17)230219(10)319500  (01)00880304532465(17)230219(10)487040  (01)00880304532465(17)230219(10)491250  (01)00880304475342(17)230220(10)446680  (01)00880304532465(17)230220(10)433430  (01)00880304532465(17)230221(10)446660  (01)00880304532465(17)230222(10)433380  (01)00880304475342(17)230222(10)594640  (01)00880304532465(17)230225(1",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0520-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83783",
      "termination_date": "20210721",
      "more_code_info": "922750  924400  926050  926200  926340  926700  929900  941230  941350  941400  941440  941490  941500  941510  941520  941530  941590  959630  960040  960060  960120  960130  960140  960160  961410  961440  961460  961470  961480  966910  973640  973660  973670  973690  973700  973710  973720  973820  973830  973840  973860  973870  973880  973890  981180  981190  981200  981220  981240  981260  981270  983710  983720  983730  983740  071660  148330  314590  459590  482030  550550  550560  576500  632620  636520  670700  014090  014100  014110  250460  296020  364470  364670  364700  364830  469280  469390  469490  469590  508590  551120  591370  591400  591410  591430  618360  628490  628640  628660  630280  630340  630390  630510  630530  780160  786790  810590  866600  901150  374180  254510  427090  487110  652680  713220  810280  846240  992120  056230  164280  207420  229690  254540  432580  436670  517390  810440  881400  082390  157930  427120  564100  779360  803170  804150  804170  004110  004200  093120  295920  295930  295940  295950  295990  296000  296010  296040  436690  436700  436720  436730  436740  553340  583860  634280  657770  713230  788630  966890  056190  153700  677460  702890  712710  772020  786780  846220  004080  153710  165570  306790  306820  410430  677450  789450  790160  790200  803500  892870  947880  550540  004240  144130  153690  153790  254530  348100  447550  490400  586100  697740  779300  780100  803220  803230  804120  811670  846150  966880  004170  044040  125470  153760  398520  427140  634380  708130  712570  712660  779330  788720  789440  803270  846130  846190  861920  087480  093210  164330  254500  254590  415630  415640  422790  427200  433460  677400  712580  715410  761270  780120  780190  786800  803260  804190  846140  929360  938000  004030  004050  004090  004150  041130  041140  044080  081030  156840  229460  230180  254470  306830  306860  410410  415590  419540  460140  490430  507410  536930  677420  697810  697820  697850  702900  708460  713260  713300  715460  779260  779280  786740  789490  789520  789550  803180  803290  810300  846200  861910  966840  978250  004060  004160  254480  254550  415560  415570  415580  419480  427160  427170  427180  499030  536910  536950  536980  586060  586110  634290  634300  634330  697770  708070  708160  712600  712720  712750  712770  713270  713290  715390  715440  780030  780070  786770  788620  803470  803480  804130  811660  811680  827450  835640  881330  909220  916650  918780  951790  966850  972900  004040  004190  004210  055510  093180  093190  246480  254490  254520  410400  454150  455880  493870  569470  586080  616190  634340  634350  634390  634400  634450  652690  677430  697750  697780  697830  698380  698410  699500  712640  712700  712730  713280  713310  715470  772010  786810  803110  803120  803310  803400  803410  811710  861950  881380  892890  916570  916630  918840  054100  125430  155960  156820  165560  229490  230200  232260  348160  348170  407860  416550  446680  455910  487140  490420  549330  550170  577210  594620  612110  790180  881360  881390  886520  886540  903010  951780  966820  966870  972920  055540  041120  041660  054070  055500  056200  095340  125450  165550  207410  229440  246470  246990  319500  348180  371150  455920  455930  487090  487130  493880  507390  550140  571430  577150  577200  579290  618340  790190  790230  810390  823760  823800  827490  835650  881370  886490  886500  886510  901110  909260  951760  951770  951800  951820  951830  972890  371670  041670  044020  055480  055490  055530  087460  094170  095350  095380  153720  153800  157840  157870  164290  164310  207430  215020  229470  247050  247060  259900  344880  344890  344910  348080  348120  407870  415410  415430  415460  416570  416580  416590  427130  433440  447590  447650  455860  487120  490410  490440  507210  507400  517380  519910  549350  550180  550200  550210  569450  571420  571440  577180  588300  588320  612100  612120  612130  652700  652710  652730  657780  657800  661490  661500  708430  810250  810410  811750  823790  835610  886530  892840  892910  892920  913570  947920  966860  003650  041150  041680  041690  044030  044050  044060  054080  054090  055470  055520  056220  081050  081060  087470  094140  094150  094160  094180  094190  095310  095320  095330  095360  095390  125440  125460  125480  144140  146100  153730  153740  153770  153810  153830  156830  157040  157080  157900  157970  164320  166710  166720  166730  166740  215200  229450  229480  229500  229510  229520  230140  230150  230160  230170  230190  230210  230220  232240  232250  232270  232280  246450  246460  247000  247010  247020  247030  247040  247070  259810  259830  259840  259860  259870  259890  319510  322710  322720  322730  322750  344870  344900  348090  348110  348130  348140  348150  371120  371130  371140  407820  407830  407840  407850  407880  415400  415420  415440  415450  415470  415490  415500  416560  416600  422780  427150  447520  447530  447560  447570  447580  447600  447610  447620  447630  455870  455890  470290  470310  470320  470330  470350  470360  470370  470390  470410  470420  493830  493840  493850  499040  507200  507220  507230  507240  507250  507260  507420  517360  517370  517400  519880  519890  519900  519920  519930  549340  550150  550160  550190  569430  569440  569460  577140  577160  577220  577230  577270  577280  577290  579300  579310  579320  588310  588330  588340  594540  594640  594680  612140  618350  652660  652670  652720  657810  661470  661480  698370  698400  708400  708410  708420  708440  708450  790210  810350  810460  811690  811700  811720  811730  811740  823770  823780  827480  835620  881760  892810  892830  892940  901100  901120  901130  903000  903020  909230  909240  909250  913530  913540  913550  913560  947860  947900  947940  947950  951810  951840  966900  972860  972870  972880  972910  972930  978230  978240  995970  657790  661510  259890R  577280R  616190R  677420R  697820R  702890R  780190R  786810R  811210R  291010  291060  291090  503830  503950  872840  081640  998060  621790  393520  982920  272200  391320",
      "recall_initiation_date": "20190821",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}