      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.",
      "address_2": "",
      "product_quantity": "219988 units",
      "code_info": "All lots manufactured between January 2008 - May 2019:     043240  073900  150190  236460  313350  391940  454660  460760  482880  498540  512280  544320  545330  633210  670360  762790  792260  792270  817590  921680  985030  985860  008330  008330R  107310  156490  181250  281130  310670  402480  501260  545340  545350  549190  678780  679390  720050  720060  755070  791580  816580  825010  954670  985290  989140  001130  039840  046290  114960  133540  157890  167740  177680  181260  199190  238670  242920  269770  281220  310920  428030  451480  465720  510770  510900  599300  696350  792280  821150  829820  830280  859420  921610  927330  947420  953480  010170  025880  046300  097120  133560  134450  157900  160130  177720  199200  206870  242580  281230  292790  310930  322360  349010  398080  428060  445800  445800R  477570  483190  495580  523750  588500  594850  599390  600450  615330  661100  696360  699090  715650  725930  730880  735850  794270  805190  821160  821960  836980  848500  864780  913390  925600  932080  932110  940500  947430  953490  953500  956980  958300  970180  993510  461290  461320  461330  807860  013580  017420  046280  046310  051290  056130  101630  134480  148320  157910  160140  178700  190510  199210  213830  214000  215840  217630  222400  242590  269780  279610  281250  292800  297290  305460  310950  348730  368120  384200  395710  395720  407270  409460  433130  454260  465450  508820  510910  523770  528600  531950  542190  559600  592990  593620  599410  603490  624040  647070  661110  667670  668880  671880  696370  706600  717020  742360  767240  771950  780790  803160  805200  821170  825060  864340  897610  902530  927380  940600  953510  957000  979870  980790  993520  997450  097130  156510  172900  354970  410340  427350  454640  457310  510920  549370  620980  696380  927860  971560  106160  156540  170710  209090  281290  295570  410350  410350R  414460  425010  460770  511140  523780  549380  596760  608580  618380  681060  696390  779820  897370  897650  952990  997920  007490  013540  026470  031560  098300  161050  222770  242640  242640R  242650  281300  311020  318200  319770  319770R  387180  395730  410360  415270  418850  425030  427360  439360  502600  510930  517230  523790  559610  592200  593120  601010  670550  687330  696400  706510  711830  718580  725940  727680  755770  763140  788380  814370  817690  817700  859660  881410  889500  903670  913680  953410  282760  343060  410370  523940  541700  670560  740380  788760  830340  897630  953420  989130  056140  066260  097140  160150  171140  178680  292850  311060  391950  599420  670570  775530  814930  821180  838430  841140  897670  050850  114730  115880  178690  242520  282790  311070  325360  382590  395740  523950  670580  711790  732150  821190  943520  951720  951730  148010  171130  248560  349020  389050  410380  425950  425970  428170  453000  502610  557910  559620  618390  740630  821200  888640  898020  953430  957010  004070  115070  131000  133600  133620  276850  281320  385700  389060  389070  410410  424720  428180  449590  454630  465780  600890  602020  614180  618400  704720  715750  729530  785000  845250  937260  942570  079640  107320  131030  158020  193720  195910  281350  294250  395750  431680  465790  465890  542410  664400  679400  711800  750440  771990  771990R  785010  953440  036750  085250  109640  115890  167470  211650  237270  248570  252320  256160  269810  409470  410470  473300  523980  561400  600460  645010  772110  849770  885870  975240  025430  082440  115900  162090  171150  175610  242550  269820  355580  358250  392020  428190  455570  501160  541710  561430  579120  585080  600480  633690  670110  715770  795180  805210  887070  914880  932120  937320  950900  953310  965790  012990  013910  043270  082350  089900  113890  158060  179680  209130  242490  242490R  248580  269840  290550  297920  311090  324980  353930  354110  361850  428210  460780  501170  518620  523990  541380  585090  585120  617590  632880  636640  636660  711810  763880  809880  915070  926740  927340  953300  969970  976190  040850  115910  158800  316060  318850  424260  501180  617600  648920  708510  837850  900320  927390  937360  994230  070980  156870  237960  281380  295580  303820  311120  319780  343990  384620  392250  428230  496650  501190  502660  599260  617610  655930  699320  745690  805220  837810  931040  996140  030740  056150  067410  156960  181220  281390  311130  391260  419750  428240  501200  555100  561440  571840  652070  670590  801860  854710  877850  897590  915080  915090  927490  957770  996150  156990  281440  349030  410990  478480  549440  641810  670600  830350  921360  932080  107460  214250  214900  349040  478490  519610  561490  585220  591300  668760  673590  719220  725950  772190  821520  848890  925580  964150  107480  269860  358400  415040  478500  502670  523740  561500  599270  652130  842950  866140  964130  115920  123670  234190  345140  450410  481030  503140  618300  671150  866220  933140  052040  087390  103350  151560  182770  240530  328360  328900  341970  383470  386230  413020  433020  482890  516120  597750  632270  692560  701330  701340  709510  781460  859600  868590  932370  981060  050270  078640  092350  103360  129860  234180  234290  242950  246730  320710  320720  327260  383490  391110  419660  419830  502790  503150  503170  563760  569920  632260  679640  715880  761510  781470  782290  785180  808060  808220  842170  864400  911380  919650  920060  952520  986220  008920  026070  345750  391100  039160  041320  052730  345740  455960  503160  543290  547170  563450  653630  675330  039170  049380  050300  052740  216270  228100  234320  234370  320730  345160  434850  469040  533050  619550  620600  688350  692570  808230  817600  842980  868600  881770  905590  945570  430570  868680  932430  037570  120370  123420  366130  546180  753300  766050  857750  072040  194010  216340  245470  248900  325120  375280  383510  420620  450870  534670  559630  602130  702920  716300  857760  868690  173910  582650  160990  311750  457820  628420  123400  194090  218330  248480  251850  255730  305450  314660  360620  384360  432740  479030  544530  546760  584900  632410  684270  843770  868710  886850  946780  965330  007060  017140  061340  061340R  135800  189730  246890  327090  433030  533910  553310  818830  838260  863030  049520  240980  246900  354630  424090  438530  448970  515700  569220  579670  598320  792160  838270  851360  118290  209950  360560  415070  432260  444840  482900  559560  808170  867400  884990  954690  118260  494790  728390  821540  934740  118270  171100  360540  728400  881480  934750  976530  115930  115950  118280  360520  494600  535830  932190  934770  465590  351230  410330  482870  830270  281100  829560  081090  698750  073810  570220  661080  698190  602000  791570  156420  026410  406960  912340  705860  793660  539100  067980  026400  066640  297960  699080  793560  032370  912410  066610  067960  428250  032360  032430  793550  981450  121500  912390  614400  775350  533410  888450  793690  288430  026420  127260  705850  775630  466430  793680  032450  288420  121490  026430  668660  632900  793670  067940  066490  903820  026370  067930  384460  066500  026360  066590  927940  927940R  793430  423630  423630R  684490  629760  055150  127130  472600  248520  793540  026390  127100  370020  026380  127090  911110  370030      UDI:    (01)00880304468351(17)200601(10)061340R  (01)00880304508965(17)201028(10)349020  (01)00880304509009(17)201128(10)617610  (01)00880304509078(17)210128(10)160130  (01)00880304509085(17)210128(10)821180  (01)00880304508903(17)210328(10)107460  (01)00880304509061(17)210328(10)696370  (01)00880304508996(17)210328(10)641810  (01)00880304509030(17)210528(10)109640  (01)00880304509115(17)210528(10)953440  (01)00880304509122(17)210528(10)652130  (01)00880304430259(17)210828(10)386230  (01)00880304508958(17)210901(10)410350R  (01)00880304509054(17)211128(10)599300  (01)00880304508989(17)211128(10)953420  (01)00880304508965(17)211128(10)410380  (01)00880304508934(17)211128(10)501260  (01)00880304509108(17)211128(10)501180  (01)00880304509009(17)211228(10)805220  (01)00880304508910(17)211228(10)501170  (01)00880304508965(17)220128(10)389050  (01)00880304509078(17)220828(10)281230  (01)00880304509115(17)220828(10)281350  (01)00880304508927(17)221028(10)385700  (01)00880304509016(17)221028(10)178690  (01)00880304508903(17)221028(10)668760  (01)00880304508941(17)221028(10)985030  (01)00880304509030(17)221128(10)248570  (01)00880304509078(17)221226(10)428060  (01)00880304509061(17)221226(10)510910  (01)00880304509092(17)221226(10)510920  (01)00880304509030(17)221226(10)523980  (01)00880304509009(17)221226(10)428230  (01)00880304508965(17)221227(10)559620  (01)00880304508996(17)221227(10)549440  (01)00880304508958(17)221228(10)618380  (01)00880304509047(17)221228(10)579120  (01)00880304508927(17)230107(10)704720  (01)00880304508972(17)230108(10)427360  (01)00880304508958(17)230111(10)549380  (01)00880304509085(17)230122(10)841140  (01)00880304508972(17)230131(10)559610  (01)00880304508965(17)230203(10)248560  (01)00880304508989(17)230205(10)523940  (01)00880304508972(17)230211(10)387180  (01)00880304508965(17)230212(10)618390  (01)00880304509047(17)230219(10)358250  (01)00880304508972(17)230227(10)439360  (01)00880304508958(17)230228(10)460770  (01)00880304461734(17)230301(10)375280  (01)00880304461734(17)230305(10)450870  (01)00880304509016(17)230306(10)523950  (01)00880304509139(17)230306(10)391260  (01)00880304509054(17)230306(10)451480  (01)00880304508941(17)230307(10)460760  (01)00880304508910(17)230308(10)460780  (01)00880304509061(17)230313(10)559600  (01)00880304509139(17)230326(10)652070  (01)00880304509139(17)230327(10)854710  (01)00880304508941(17)230409(10)512280  (01)00880304509078(17)230409(10)699090  (01)00880304508958(17)230409(10)511140  (01)00880304509061(17)230412(10)997450  (01)00880304509054(17)230412(10)428030  (01)00880304509054(17)230415(10)039840  (01)00880304509139(17)230415(10)030740  (01)00880304508972(17)230708(10)418850  (01)00880304508972(17)230731(10)711830  (01)00880304508910(17)230911(10)012990  (01)00880304509078(17)230911(10)025880  (01)00880304509061(17)230923(10)624040  (01)00880304509061(17)230924(10)217630  (01)00880304509061(17)231011(10)902530  (01)00880304508910(17)231107(10)082350  (01)00880304461727(17)231107(10)584900  (01)00880304508927(17)231109(10)424720  (01)00880304509030(17)231115(10)772110  (01)00880304508972(17)231115(10)601010  (01)00880304509061(17)231120(10)803160  (01)00880304509061(17)231121(10)771950  (01)00880304508927(17)231202(10)845250  (01)00880304509139(17)231203(10)927490  (01)00880304509115(17)231207(10)771990  (01)00880304509115(17)231207(10)771990R  (01)00880304509139(17)231209(10)067410  (01)00880304509108(17)231209(10)927390  (01)00880304509061(17)231216(10)395710  (01)00880304508910(17)231216(10)927340  (01)00880304508910(17)231220(10)926740  (01)00880304509047(17)231227(10)965790  (01)00880304509054(17)240116(10)167740  (01)00880304508972(17)240120(10)007490  (01)00880304508972(17)240127(10)395730  (01)00880304509061(17)240224(10)051290  (01)00880304509016(17)240224(10)395740  (01)00880304468375(17)240225(10)432260  (01)00880304509009(17)240227(10)502660  (01)00880304509061(17)240319(10)305460  (01)00880304509115(17)240319(10)395750  (01)00880304509122(17)240319(10)502670  (01)00880304509030(17)240320(10)256160  (01)00880304508958(17)240320(10)779820  (01)00880304508972(17)240320(10)502600  (01)00880304509061(17)240321(10)214000  (01)00880304509061(17)240330(10)395720  (01)00880304509061(17)240330(10)940600  (01)00880304509061(17)240403(10)215840  (01)00880304509078(17)240406(10)940500  (01)00880304509047(17)240507(10)541710  (01)00880304508965(17)240516(10)502610  (01)00880304509061(17)240529(10)864340  (01)00880304509061(17)240529(10)742360  (01)00880304508903(17)240609(10)772190  (01)00880304508910(17)240611(10)969970  (01)00880304509139(17)240611(10)181220  (01)00880304509047(17)240612(10)600480  (01)00880304509030(17)240613(10)085250  (01)00880304509078(17)240619(10)864780  (01)00880304508927(17)240619(10)131000  (01)00880304509078(17)240622(10)958300  (01)00880304509139(17)240622(10)801860  (01)00880304508989(17)240623(10)541700  (01)00880304509115(17)240624(10)131030  (01)00880304509115(17)240708(10)431680  (01)00880304508972(17)240714(10)222770  (01)00880304509108(17)240724(10)158800  (01)00880304508910(17)240805(10)518620  (01)00880304509061(17)240805(10)384200  (01)00880304509030(17)240811(10)409470  (01)00880304508910(17)240811(10)324980  (01)00880304509061(17)240813(10)409460  (01)00880304509078(17)240820(10)932080  (01)00880304468375(17)240828(10)808170  (01)00880304509054(17)240903(10)181260  (01)00880304509078(17)240903(10)932110  (01)00880304509047(17)240903(10)932120  (01)00880304508965(17)240924(10)425950  (01)00880304508965(17)240924(10)425970  (01)00880304508941(17)240929(10)985860  (01)00880304508934(17)240930(10)181250  (01)00880304461727(17)241023(10)218330  (01)00880304508910(17)241027(10)290550  (01)00880304509061(17)241029(10)717020  (01)00880304508927(17)241029(10)729530  (01)00880304509115(17)241029(10)542410  (01)00880304509078(17)241029(10)715650  (01)00880304508972(17)241029(10)593120  (01)00880304508927(17)241029(10)715750  (01)00880304509047(17)241029(10)715770  (01)00880304509139(17)241029(10)877850  (01)00880304509061(17)241203(10)433130  (01)00880304461734(17)241205(10)248900  (01)00880304508941(17)241209(10)817590  (01)00880304461734(17)241211(10)216340  (01)00880304509061(17)241222(10)199210  (01)00880304509078(17)250106(10)199200  (01)00880304508910(17)250227(10)158060  (01)00880304509061(17)250319(10)190510  (01)00880304509054(17)250319(10)199190  (01)00880304509078(17)250319(10)970180  (01)00880304508958(17)250319(10)997920  (01)00880304508972(17)250319(10)903670  (01)00880304508910(17)250319(10)013910  (01)00880304508996(17)250319(10)921360  (01)00880304509016(17)250324(10)943520  (01)00880304509115(17)250325(10)158020  (01)00880304509061(17)250405(10)508820  (01)00880304509061(17)250406(10)528600  (01)00880304508965(17)250416(10)740630  (01)00880304509078(17)250416(10)206870  (01)00880304508934(17)250420(10)816580  (01)00880304509009(17)250420(10)837810  (01)00880304509108(17)250420(10)837850  (01)00880304508910(17)250424(10)353930  (01)00880304509054(17)250424(10)114960  (01)00880304509139(17)250428(10)555100  (01)00880304509078(17)250501(10)594850  (01)00880304509061(17)250504(10)671880  (01)00880304509047(17)250504(10)455570  (01)00880304509054(17)250520(10)947420  (01)00880304508910(17)250520(10)915070  (01)00880304509139(17)250520(10)915080  (01)00880304509030(17)250601(10)849770  (01)00880304509009(17)250602(10)931040  (01)00880304509061(17)250602(10)454260  (01)00880304509139(17)250602(10)957770  (01)00880304509139(17)250602(10)915090  (01)00880304508972(17)250603(10)013540  (01)00880304509078(17)250604(10)947430  (01)00880304508910(17)250608(10)763880  (01)00880304509061(17)250609(10)013580  (01)00880304508972(17)250610(10)755770  (01)00880304509009(17)250611(10)070980  (01)00880304508972(17)250626(10)319770  (01)00880304508965(17)250626(10)148010  (01)00880304508972(17)250626(10)319770R  (01)00880304509054(17)250630(10)001130  (01)00880304509009(17)250707(10)303820  (01)00880304509009(17)250708(10)392250  (01)00880304508941(17)250708(10)391940  (01)00880304509085(17)250708(10)391950  (01)00880304509115(17)250708(10)195910  (01)00880304508972(17)250709(10)415270  (01)00880304508972(17)250709(10)318200  (01)00880304509009(17)250709(10)319780  (01)00880304509047(17)250713(10)392020  (01)00880304509009(17)250728(10)655930  (01)00880304509078(17)250730(10)848500  (01)00880304509122(17)250731(10)599270  (01)00880304509009(17)250810(10)599260  (01)00880304461734(17)250813(10)420620  (01)00880304509009(17)250820(10)699320  (01)00880304509061(17)250827(10)897610  (01)00880304508927(17)250902(10)942570  (01)00880304509092(17)250908(10)172900  (01)00880304508996(17)250908(10)932080  (01)00880304509078(17)250910(10)956980  (01)00880304508965(17)250911(10)957010  (01)00880304509061(17)250911(10)957000  (01)00880304509139(17)250916(10)897590  (01)00880304508941(17)250921(10)043240  (01)00880304509078(17)250921(10)133560  (01)00880304509061(17)250921(10)592990  (01)00880304508958(17)250921(10)209090  (01)00880304509054(17)250921(10)133540  (01)00880304508910(17)250921(10)043270  (01)00880304508910(17)250921(10)113890  (01)00880304508958(17)250923(10)295570  (01)00880304508910(17)250930(10)209130  (01)00880304509078(17)251001(10)322360  (01)00880304509061(17)251001(10)017420  (01)00880304508927(17)251013(10)276850  (01)00880304508927(17)251014(10)133620  (01)00880304509009(17)251014(10)295580  (01)00880304508989(17)251014(10)343060  (01)00880304461734(17)251026(10)072040  (01)00880304509092(17)251026(10)354970  (01)00880304508927(17)251118(10)600890  (01)00880304509054(17)260106(10)792280  (01)00880304509078(17)260114(10)821960  (01)00880304508941(17)260115(10)792260  (01)00880304508941(17)260115(10)792270  (01)00880304508972(17)260115(10)725940  (01)00880304508903(17)260119(10)821520  (01)00880304509078(17)260119(10)097120  (01)00880304509085(17)260119(10)097140  (01)00880304509092(17)260120(10)097130  (01)00880304509122(17)260126(10)269860  (01)00880304509061(17)260126(10)269780  (01)00880304508910(17)260126(10)269840  (01)00880304509054(17)260127(10)269770  (01)00880304509030(17)260127(10)269810  (01)00880304509047(17)260208(10)269820  (01)00880304508934(17)260216(10)678780  (01)00880304508903(17)260217(10)719220  (01)00880304461727(17)260217(10)384360  (01)00880304509078(17)260218(10)735850  (01)00880304509054(17)260219(10)510770  (01)00880304508934(17)260224(10)402480  (01)00880304508941(17)260329(10)498540  (01)00880304509078(17)260329(10)445800  (01)00880304509061(17)260329(10)046310  (01)00880304509078(17)260329(10)046300  (01)00880304509078(17)260329(10)445800R  (01)00880304509061(17)260330(10)046280  (01)00880304509061(17)260407(10)101630  (01)00880304509009(17)260415(10)745690  (01)00880304509061(17)260419(10)279610  (01)00880304509047(17)260429(10)025430  (01)00880304508941(17)260429(10)762790  (01)00880304509078(17)260503(10)588500  (01)00880304509078(17)260505(10)398080  (01)00880304509054(17)260505(10)046290  (01)00880304509030(17)260505(10)645010  (01)00880304509047(17)260505(10)175610  (01)00880304509061(17)260511(10)148320  (01)00880304508972(17)260601(10)763140  (01)00880304508972(17)260602(10)788380  (01)00880304509115(17)260602(10)750440  (01)00880304509061(17)260605(10)368120  (01)00880304508972(17)260606(10)718580  (01)00880304509061(17)260607(10)780790  (01)00880304509115(17)260607(10)664400  (01)00880304508910(17)260608(10)976190  (01)00880304508934(17)260705(10)989140  (01)00880304508934(17)260706(10)008330  (01)00880304509115(17)260706(10)294250  (01)00880304508910(17)260706(10)297920  (01)00880304508972(17)260706(10)031560  (01)00880304508934(17)260706(10)008330R  (01)00880304509061(17)260725(10)531950  (01)00880304508927(17)260725(10)614180  (01)00880304509047(17)260725(10)633690  (01)00880304509047(17)261228(10)082440  (01)00880304508927(17)261229(10)115070  (01)00880304509139(17)270107(10)571840  (01)00880304509139(17)270107(10)056150  (01)00880304509061(17)270120(10)056130  (01)00880304508972(17)270126(10)687330  (01)00880304509047(17)270707(10)242550  (01)00880304508972(17)270710(10)242640  (01)00880304508972(17)270710(10)242640R  (01)00880304508927(17)270722(10)602020  (01)00880304509115(17)270722(10)785010  (01)00880304509030(17)270723(10)885870  (01)00880304509061(17)270803(10)292800  (01)00880304508910(17)270809(10)242490  (01)00880304509061(17)270809(10)242590  (01)00880304508910(17)270809(10)242490R  (01)00880304509078(17)270901(10)292790  (01)00880304508910(17)270905(10)541380  (01)00880304509047(17)271014(10)585080  (01)00880304509009(17)271022(10)496650  (01)00880304509016(17)271022(10)282790  (01)00880304509047(17)271022(10)355580  (01)00880304508910(17)271022(10)585090  (01)00880304508958(17)271023(10)156540  (01)00880304508996(17)271024(10)156990  (01)00880304509139(17)271025(10)156960  (01)00880304508927(17)271101(10)785000  (01)00880304508989(17)271222(10)282760  (01)00880304509061(17)271117(10)980790  (01)00880304508934(17)271117(10)791580  (01)00880304508903(17)271119(10)585220  (01)00880304509078(17)271129(10)925600  (01)00880304508910(17)271201(10)711810  (01)00880304509054(17)271204(10)921610  (01)00880304509078(17)271212(10)953490  (01)00880304509016(17)271227(10)114730  (01)00880304509061(17)280108(10)979870  (01)00880304508934(17)280108(10)156490  (01)00880304509054(17)280109(10)953480  (01)00880304509061(17)280109(10)993520  (01)00880304509078(17)280111(10)177720  (01)00880304509016(17)280111(10)311070  (01)00880304509061(17)280111(10)178700  (01)00880304508910(17)280111(10)311090  (01)00880304509122(17)280112(10)842950  (01)00880304508910(17)280112(10)585120  (01)00880304430259(17)280112(10)981060  (01)00880304509085(17)280117(10)311060  (01)00880304509078(17)280117(10)953500  (01)00880304508965(17)280118(10)888640  (01)00880304508972(17)280118(10)311020  (01)00880304509054(17)280124(10)177680  (01)00880304509078(17)280124(10)993510  (01)00880304509061(17)280124(10)953510  (01)00880304508934(17)280126(10)310670  (01)00880304508972(17)280126(10)817690  (01)00880304509078(17)280129(10)310930  (01)00880304509078(17)280129(10)495580  (01)00880304508965(17)280129(10)453000  (01)00880304509047(17)280129(10)162090  (01)00880304509061(17)280210(10)310950  (01)00880304509054(17)280212(10)310920  (01)00880304508934(17)280216(10)720060  (01)00880304509016(17)280216(10)732150  (01)00880304509061(17)280216(10)825060  (01)00880304509078(17)280219(10)730880  (01)00880304508934(17)280219(10)825010  (01)00880304508972(17)280219(10)817700  (01)00880304508972(17)280219(10)814370  (01)00880304508910(17)280223(10)636660  (01)00880304508972(17)280224(10)913680  (01)00880304508972(17)280301(10)098300  (01)00880304508958(17)280301(10)952990  (01)00880304509016(17)280301(10)711790  (01)00880304508910(17)280302(10)636640  (01)00880304509139(17)280401(10)311130  (01)00880304509061(17)280405(10)407270  (01)00880304509009(17)280405(10)384620  (01)00880304509030(17)280406(10)975240  (01)00880304509085(17)280419(10)814930  (01)00880304509061(17)280423(10)706600  (01)00880304509061(17)280423(10)767240  (01)00880304509108(17)280423(10)900320  (01)00880304508972(17)280423(10)889500  (01)00880304508972(17)280424(10)706510  (01)00880304509122(17)280425(10)415040  (01)00880304509061(17)280428(10)297290  (01)00880304509085(17)280430(10)897670  (01)00880304260320(17)280604(10)842980  (01)00880304816374(10)461330  (01)00880304816374(10)461290  (01)00880304816374(10)461320  (01)00880304816374(10)807860  (01)00880304509023(17)210328(10)351230  (01)00880304509023(17)220528(10)830270  (01)00880304509023(17)250407(10)570220  (01)00880304509023(17)251118(10)661080  (01)00880304509023(17)260818(10)698190  (01)00880304509023(17)261228(10)602000  (01)00880304509023(17)271118(10)791570  (01)00880304509023(17)271212(10)156420  (01)00880304554900(17)230301(10)539100  (01)00880304554894(17)230128(10)032370  (01)00880304554887(17)230128(10)032360  (01)00880304554887(17)230128(10)032430  (01)00880304554887(17)230213(10)793550  (01)00880304554887(17)271101(10)981450  (01)00880304554931(17)230107(10)793690  (01)00880304554924(17)230121(10)793680  (01)00880304554924(17)230128(10)032450  (01)00880304554917(17)230115(10)793670  (01)00880304554832(17)230107(10)927940  (01)00887868346934(17)230107(10)927940R  (01)00880304554832(17)230306(10)423630  (01)00887868346934(17)230306(10)423630R  (01)00880304554832(17)230325(10)684490  (01)00880304554870(17)230121(10)793540",
      "center_classification_date": "20191122",
      "distribution_pattern": "US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.",
      "state": "IN",
      "product_description": "Compress, Mini Compress; Item Nos. 178350  178351  178352  178353  178354  178355  178356  178357  178358  178359  178360  178361  178362  178363  178364  178365  178366  178367  178368  178369  178370  178371  178372  178373  178464  178472  178480  178488  178496  178504  178575  178576  178577  178578  178579  178580  178730  178731  178732  178733  178734  178735  178354S  178754  178755  178756  178757  178758  178759    Product Usage: 1) Correction of revision of unsuccessful osteotomy, arthrodesis or previous joint replacement 2) Tumor resections 3) Revision of previously failed total joint arthroplasty 4) Trauma The ComPreSs Distal Femoral Replacement System is intended for uncemented use.",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-0519-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "83783",
      "termination_date": "20210721",
      "more_code_info": "",
      "recall_initiation_date": "20190821",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}