  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Roswell",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66859",
      "recalling_firm": "Kimberly-Clark Corporation",
      "address_1": "1400 Holcomb Bridge Rd",
      "address_2": "N/A",
      "postal_code": "30076-2190",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide Distribution",
      "recall_number": "Z-0518-2014",
      "product_description": "Kimberly-Clark Single Shot Epidural Pain Management Trays containing the BD LOR Syringe (BD Code 405198)    Product Usage:  The device is designed to provide regional anesthesia. The device may contain syringes, needles and drugs",
      "product_quantity": "389,300 units",
      "reason_for_recall": "Kimberly-Clark Single Shot Epidural Pain Management Trays contain the BD 7 mL Epilor plastic Luer-Lok LOR Syringes under recall by Becton Dickinson (BD) because the product may stall or stick when traveling within the barrel of the syringe.",
      "recall_initiation_date": "20131024",
      "center_classification_date": "20131216",
      "termination_date": "20170825",
      "report_date": "20131225",
      "code_info": "Lot numbers: 181024, 181129 181217, 181221, 181231, 182022, 182058, 182101,  182108, 182131, 182222, 182223, 182263, 182265, 182289, 182293, 182301, 182315, 181A074, 181A94, 181A154, 181A177, 181A286, 181A289, 181A369, 182A002, 182A005, 182A012, 182A055, 182A056, 182A057, 182A064, 182A074, 182A075/182A78, 182A085, 182090, 182A142, 182AA144, 182A151, 182A155, 182A178, 182A194, 182A196, 182A198, 182A202, 182A204, 182A212, 182A216 and 189A001 (syringe)."
    }
  ]
}