      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89260",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0516-2022",
      "product_description": "Artis icono biplane with software VE20C-diagnostic imaging angiography system  Model: 11327600",
      "product_quantity": "72 units",
      "reason_for_recall": "Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1.\tUpdated calibration data not saved with measurement after scene+/-; 2.\tNo x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap;  4.\tUnintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option",
      "recall_initiation_date": "20211012",
      "center_classification_date": "20220121",
      "termination_date": "20240411",
      "report_date": "20220202",
      "code_info": "Serial Numbers: 180062 180318 180054 180019 180065 180057 180041 180317 180034 180072 180349 180060 180337 180070 180048 180083 180369 180370 180345 180315 180343 180080 180301 180305 180067 180330 180040 180360 180333 180029 180352 180336 180316 180068 180363 180084 180334 180018 180350 180354 180046 180047 180335 180085 180035 180339 180357 180358 180361 180362 180081 180331 180021 180342 180321 180309 180356 180368 180341 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037  UDI: 04056869063317  Expanded Recall 12/20/21 S/N: 180056",
      "more_code_info": ""
    }
  ]
}