{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Miami",
      "address_1": "11800 SW 147th Ave",
      "reason_for_recall": "When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review.  Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.",
      "address_2": "",
      "product_quantity": "1,056 (418 US, and 638 OUS)",
      "code_info": "Catalog Number 775222.  Serial numbers: AS39017, AS40018, AT35013, AT35014, AT35015, AT35017, AT35018, AT40020, AT40021, AT40022, AT40024, AT40025, AT40027, AT40030, AT45031, AT45032, AT45034, AT45035, AT49037, AT49041, AT49044, AT49045, AT49046, AT49047, AT49048, AT49049, AT49050, AT49051, AT49055, AT49057, AT49058, AT49059, AT49060, AT49061, AT49062, AT49063, AT49064, AT49065, AU01004, AU01005, AU01007, AU01009, AU01010, AU01011, AU01012, AU01014, AU01015, AU01016, AU01017, AU01020, AU01021, AU01022, AU01023, AU01024, AU01025, AU01028, AU01029, AU01030, AU08041, AU08043, AU08044, AU08047, AU08048, AU08049, AU08050, AU11051, AU11053, AU11057, AU11058, AU11059, AU11061, AU11062, AU11063, AU11064, AU11066, AU11068, AU11072, AU13073, AU13075, AU13077, AU13079, AU14083, AU14084, AU14085, AU14086, AU14087, AU14088, AU14090, AU16104, AU16111, AU16112, AU19113, AU19115, AU19118, AU19119, AU19120, AU20123, AU20124, AU20125, AU20128, AU20129, AU20130, AU20131, AU21133, AU21136, AU21137, AU21141, AU23146, AU23147, AU23149, AU23150, AU23151, AU23152, AU28153, AU28154, AU28160, AU28161, AU29163, AU29165, AU29167, AU29168, AU29169, AU29172, AU30173, AU30178, AU30180, AU30181, AU30182, AU35183, AU35185, AU35186, AU35187, AU35188, AU36189, AU36192, AU36193, AU36196, AU36198, AU36200, AU36201, AU36202, AU36203, AU36204, AU40205, AU40207, AU40209, AU40210, AU40211, AU40213, AU40214, AU40215, AU40216, AU40217, AU40218, AU40221, AU43223, AU43225, AU43226, AU43228, AU43229, AU43231, AU43232, AU43233, AU43234, AU43235, AU43237, AU43240, AU49254, AU49256, AU49259, AV01003, AV01004, AV01005, AV01007, AV01009, AV01010, AV01011, AV01012, AV01013, AV01014, AV01016, AV06018, AV06020, AV06023, AV06026, AV06028, AV06030, AV10033, AV10034, AV10035, AV10036, AV10038, AV10039, AV10042, AV10044, AV10045, AV12047, AV12049, AV12054, AV12055, AV12056, AV14057, AV14059, AV14061, AV14062, AV14063, AV16072, AV16073, AV16075, AV16077, AV16078, AV16080, AV16082, AV16083, AV16086, AV16088, AV16089, AV16090, AV16091, AV21092, AV21093, AV21095, AV21096, AV22097, AV22100, AV22102, AV22105, AV22106, AV23107, AV23109, AV23110, AV23111, AV23112, AV23113, AV23116, AV27117, AV27118, AV27120, AV27121, AV36127, AV36128, AV36129, AV36130, AV39131, AV39132, AV39133, AV39134, AV39135, AV40136, AV40137, AV40139, AV40140, AV40143, AV40145, AV42147, AV42149, AV42150, AV42151, AV42152, AV42153, AV42154, AV42155, AV44156, AV44157, AV44158, AV44160, AV44161, AV44162, AV44163, AV44164, AV47167, AV47168, AV47169, AV47170, AV48171, AV48173, AV48174, AV48176, AV48177, AV48178, AV48179, AV49181, AV49183, AV49184, AV49185, AV49187, AV49188, AV49189, AV51191, AV51192, AW01001, AW01002, AW01007, AW01008, AW01010, AW01011, AW01012, AW01013, AW01014, AW01015, AW06016, AW06018, AW06019, AW06020, AW06021, AW06023, AW06024, AW06025, AW06026, AW06027, AW06028, AW06029, AW09031, AW09032, AW09033, AW09034, AW09035, AW10036, AW10037, AW10038, AW10039, AW10040, AW10041, AW10044, AW10045, AW14046, AW14048, AW14049, AW14051, AW16052, AW16053, AW16054, AW16056, AW16057, AW16058, AW16059, AW16060, AW16061, AW20062, AW20063, AW20064, AW20065, AW20066, AW20067, AW20068, AW20069, AW20070, AW20071, AW23072, AW23073, AW23074, AW23075, AW23076, AW23077, AW23078, AW23079, AW28087, AW28088, AW28089, AW28090, AW28091, AW28094, AW28095, AW28096, AW28097, AW28098, AW31099, AW31100, AW31102, AW31103, AW31104, AW31105, AW31106, AW31107, AW31108, AW31109, AW31110, AW35111, AW35112, AW35113, AW35114, AW35115, AW35116, AW35117, AW35120, AW35121, AW35122, AW35123, AW35124, AW35125, AW35126, AW35127, AW35128, AW42129, AW42131, AW42132, AW42133, AW45134, AW45135, AW45136, AW45137, AW45138, AW45139, AW45140, AW45141, AW45143, AW45144, AW45145, AW45146, AW45147, AW47148, AW47149, AW47151, AW47152, AW47153, AW47154, AW47155, AW47156, AW47157, AW49159, AW49160, AW49161, AW49162, AW49163, AW49164, AW49165, AY01001, AY01002, AY01004, AY01005, AY01006, AY01007, AY01008, AY01009, AY01010, AY04011, AY04012, AY04013, AY04015, AY04016, AY04017, AY04018, AY04019, AY04020, AY08021, AY08022, AY08023, AY08024, AY08025, AY08026, AY08027, AY08028, AY08029, AY08032, AY08033, AY08034, AY10036, AY10037, AY10038, AY10039, AY10040, AY10041, AY10042, AY10044, AY10045, AY10046, AY10047, AY10048, AY10049, AY10051, AY10053, AY10054, AY10055, AY17056, AY17057, AY17058, AY17059, AY17060, AY17061, AY17062, AY17063, AY17064, AY17065, AY19066, AY19067, AY19069, AY19070, AY19071, AY19072, AY19073, AY19074, AY19075, AY23076, AY23077, AY23078, AY23079, AY23080, AY23081, AY30085, AY30086, AY30087, AY30088, AY30090, AY32092, AY32093, AY32094, AY32095, AY32096, AY32097, AY32098, AY32099, AY32100, AY32101, AY32102, AY32103, AY32104, AY32105, AY32106, AY35107, AY35108, AY35109, AY35110, AY35111, AY35112, AY35113, AY35114, AY35115, AY35116, AY35117, AY35118, AY35119, AY35120, AY35121, AY40123, AY40124, AY40125, AY40126, AY41127, AY41128, AY41129, AY41130, AY41132, AY41133, AY41134, AY41140, AY41141, AY51142, AY51143, AY51144, AY51145, AY51149, AY52150, AY52151, AY52152, AY52153, AY52154, AY52155, AY52156, AY52157, AY52158, AY52159, AZ02001, AZ02002, AZ02004, AZ02006, AZ02007, AZ02008, AZ02009, AZ02010, AZ06011, AZ06012, AZ06013, AZ06014, AZ06015, AZ06016, AZ06017, AZ06018, AZ06020, AZ06021, AZ06022, AZ08023, AZ08024, AZ08025, AZ08026, AZ08027, AZ08028, AZ08029, AZ08030, AZ08031, AZ12033, AZ12034, AZ12035, AZ12038, AZ12040, AZ12042, AZ12043, AZ12044, AZ17045, AZ17046, AZ17048, AZ17050, AZ17051, AZ17053, AZ18054, AZ18055, AZ18056, AZ18057, AZ19058, AZ19059, AZ19060, AZ19061, AZ20062, AZ20063, AZ20064, AZ20065, AZ20066, AZ20067, AZ20068, AZ20069, AZ20070, AZ22071, AZ22072, AZ22073, AZ22074, AZ22075, AZ23077, AZ23078, AZ23079, AZ23080, AZ24081, AZ24082, AZ24083, AZ24084, AZ24085, AZ25087, AZ25088, AZ25089, AZ25090, AZ28091, AZ28092, AZ28093, AZ28094, AZ28095, AZ28096, AZ28097, AZ28098, AZ28099, AZ28100, AZ34101, AZ34102, AZ34103, AZ34104, AZ34105, AZ34106, AZ34107, AZ34108, AZ34109, AZ34110, AZ34111, AZ34112, AZ34113, AZ34114, AZ34115, AZ39116, AZ39118, AZ39119, AZ39120, AZ39121, AZ39122, AZ39123, AZ39124, AZ39125, AZ41127, AZ41131, AZ41132, AZ41133, AZ41134, AZ41135, AZ43136, AZ43137, AZ43138, AZ43139, AZ43140, AZ43141, AZ43143, AZ46146, AZ46147, AZ46148, AZ46149, AZ46150, AZ46151, AZ46152, AZ46156, AZ46157, AZ46158, AZ46159, AZ48160, AZ48162, AZ48163, AZ48164, AZ48166, AZ48167, AZ49168, AZ49169, AZ49170, AZ49172, AZ49173, AZ49174, AZ49175, AZ49176, BA03001, BA03002, BA03003, BA03004, BA03005, BA03006, BA03007, BA03008, BA03009, BA03010, BA07011, BA07012, BA07013, BA07014, BA07015, BA07016, BA07017, BA07018, BA07019, BA07020, BA08021, BA08022, BA08023, BA08024, BA08025, BA10026, BA10027, BA10028, BA10029, BA10030, BA10031, BA13032, BA13034, BA13035, BA13036, BA13037, BA13038, BA13039, BA13040, BA13041, BA15042, BA15045, BA16047, BA16048, BA16049, BA16050, BA17052, BA17053, BA17054, BA18055, BA18056, BA18057, BA19060, BA19061, BA19062, BA21064, BA22067, BA22068, BA22069, BA22070, BA23071, BA23072, BA23073, BA24075, BA24076, BA24077, BA24078, BA25080, BA25081, BA25082, BA25083, BA25084, BA25085, BA27086, BA27087, BA27088, BA27089, BA27090, BA29091, BA29092, BA29093, BA29094, BA29095, BA31096, BA31097, BA31098, BA31099, BA31100, BA35101, BA35103, BA35104, BA35105, BA36106, BA36107, BA37109, BA37110, BA37111, BA37112, BA38113, BA38114, BA38115, BA40116, BA40117, BA40118, BA41119, BA41120, BA41121, BA41122, BA43124, BA43126, BA46127, BA46128, BA46129, BA46130, BA48131, BA48132, BA48133, BA49134, BA49135, BA49136, BA49137, BA50138, BA50139, BA50140, BA50141, BA50142, BB03001, BB03002, BB03003, BB03004, BB04005, BB04006, BB05007, BB05008, BB05009, BB06010, BB09011, BB09012, BB09013, BB09014, BB09015, BB16016, BB16017, BB16018, BB16019, BB16021, BB16022, BB16023, BB23024, BB23025, BB23026, BB23027, BB23028, BB27029, BB27030, BB27031, BB27032, BB27033, BB27034, BB47045, BB47046, BB47047, BB47048, BB47049, BB47050, BB47051, BB47052, BB47053, BB47054, BC01001, BC01002, BC01003, BC01004, BC04006, BC04007, BC04008, BC10016, BC10017, BC10018, BC10019, BC14021, BC14022, BC14023, BC14024, BC21026, BC21027, BC29030, BC29031, BC29032, BC29035, BC30036, BC30039.      Catalog Number B66705.     Serial Numbers:  AZ30002, AZ30003, AZ30004, AZ39005, BA14002.      Catalog Number C11477.  Serial Numbers:  BB11109, BB11110, BB12111, BB12117, BB12118, BB12119, BB12120, BB12121, BB12122, BB12123, BB18124, BB18125, BB18126, BB18127, BB18128, BB18129, BB18130, BB18133, BB23134, BB23135, BB23136, BB31145, BB31146, BB31147, BB31148, BB31149, BB31150, BB36157, BB36158, BB36159, BB36160, BB36161, BB36162, BB43163, BB43164, BB43165, BB43166, BB43167, BB43168, BB47169, BB47170, BB47171, BB47172, BB47173, BB47174, BB47176, BB47177, BC01101, BC01102, BC01103, BC01104, BC01105, BC04106, BC04108, BC04109, BC04110, BC04111, BC04112, BC04113, BC04114, BC04115, BC09117, BC09118, BC09119, BC09120, BC09121, BC09122, BC09123, BC09124, BC09125, BC09126, BC09127, BC09128, BC14131, BC14132, BC14133, BC14134, BC14135, BC14136, BC14137, BC14138, BC14140, BC14141, BC14142, BC14143, BC14144, BC14145, BC19146, BC19147, BC19148, BC19149, BC19150, BC19151, BC19152, BC19153, BC19154, BC19155, BC19156, BC19157, BC21158, BC21159, BC21160, BC21161, BC21162, BC21163, BC21164, BC21166, BC21169, BC21171, BC24172, BC24173, BC24174, BC24175, BC24176, BC24178, BC24180, BC24181, BC30182, BC30183, BC30184, BC30185, BC30186, BC30187, BC30188, BC30189, BC30190, BC30191, BC33192, BC33193, BC33194, BC33195, BC33196, BC33197, BC33198, BC33200, BC33201, BC33202, BC33203, BC33204, BC33206, BC36207, BC36208, BC36209, BC36210, BC36211, BC36212, BC36213, BC36214, BC36215.",
      "center_classification_date": "20191122",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Vietnam.",
      "state": "FL",
      "product_description": "UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.",
      "report_date": "20191204",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beckman Coulter, Inc.",
      "recall_number": "Z-0516-2020",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "84186",
      "termination_date": "20210427",
      "more_code_info": "",
      "recall_initiation_date": "20191007",
      "postal_code": "33196-2500",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}