{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elizabethtown",
      "state": "KY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81475",
      "recalling_firm": "EUROTROL INC",
      "address_1": "850 N Black Branch Rd",
      "address_2": "N/A",
      "postal_code": "42701",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Austria, Australia, Bahrain, France, Hong Kong, Indonesia, Ireland, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Maldives, New Zealand, Mozambique, Pakistan, Poland, Slovenia, Africa, Spain, Thailand, Finland, Norway,  and UK.",
      "recall_number": "Z-0515-2019",
      "product_description": "HemoTrol Level 3- In Vitro Diagnostic Hematology quality control mixture.  Catalogue number: 171.003.002    Eurotrol HemoLin is an assayed hemoglobin control intended for professional use in the verification of the linearity of the HemoCue¿ Hb 201 systems.",
      "product_quantity": "2122",
      "reason_for_recall": "Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values",
      "recall_initiation_date": "20181018",
      "center_classification_date": "20181128",
      "termination_date": "20200107",
      "report_date": "20181205",
      "code_info": "Batch Number: 82467"
    }
  ]
}