  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elyria",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86604",
      "recalling_firm": "Invacare Corporation",
      "address_1": "1200 Taylor St",
      "address_2": "N/A",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US Nationwide, Canada, Australia, China, Germany, Guam, New Zealand, and Northern Mariana Islands",
      "recall_number": "Z-0514-2021",
      "product_description": "Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen   Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada),  All World:  IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5",
      "product_quantity": "384,767 units",
      "reason_for_recall": "Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.",
      "recall_initiation_date": "20201030",
      "center_classification_date": "20201201",
      "termination_date": "20241231",
      "report_date": "20201209",
      "code_info": "Model numbers:  IRC5PO2V from Serial # 14HF039890 to 18JF007415;  IRC5PO2VC from Serial # 14IF004028 to 18JF028633;    IRC5PO2VAW  from serial #17HF030338-18IF018522;   IRC5PO2VAWL2 from serial # 18BF007341-18IF024451;   IRC5PO2VAWL3 from serial # 17JF002563-19AF008559;   IRC5PO2VAWL5 from serial # 18EF003466-18EF003544.    DI Number: 00841447100249",
      "more_code_info": ""
    }
  ]
}