  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79088",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-0514-2018",
      "product_description": "Precedence 16 Model # 882350    Product Usage:  Precedence (cleared as Griffin) is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Precedence may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data for anatomical localization of the nuclear medicine data. Precedence may be used either as a separate single photon system, a separate CT system or as a combined CT and single photon system. The nuclear medicine and CT images may be transferred to other systems such as a radiation therapy planning system. The Precedence Imaging System should only be used by trained healthcare professionals.",
      "product_quantity": "1387 in total",
      "reason_for_recall": "During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.",
      "recall_initiation_date": "20171215",
      "center_classification_date": "20180202",
      "termination_date": "20200410",
      "report_date": "20180214",
      "code_info": "Precedence 6 Model # 882351  Serial Numbers :   KP05100002  KP06040012  KP05040006  KP05080007  KP05110003  KP0510008  30008  KP06110019  K06110017  6080522  30007060027  3000034  3000032  30007060029  3000006  KP06020011  KP06050013  KP06080008  KP05110004  KP06110018  30007040026  KP06070015  30007060028  3000033  30007030025  3000031  3000035  3000040  3000041  30058  3000049  3000054  3000039  3000038  3000042  3000093  3000044  3000036  3000043  3000047  KP06070048  3000074  3000059  3000055  340090  3000057  3000058  3000053  3000050  3000056  3000052  3000061  3000051  3000045  3000048  3000069  3000066  3000078  3000060  3000064  3000068  3000065  30007070030  3000071  3000063  3000067  3000062  300072  3000095  3000073  3000085  30007080011  30007040024  3000102  3000075  3000076  7021  3000092  3000098  3000087  3000077  3000082  3000080  3000081  3000088  3000079  3000083  3000084  3000089  3000091  3000096  3000097  3000100  3000086  3000090  3000099  3000094  3000104  3000101  3000103  3000105  3000106  3000037  3000708005  30006120020  30007020023"
    }
  ]
}