  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79088",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-0513-2018",
      "product_description": "Precedence 6 Model # 882351    Product Usage:  Precedence (cleared as Griffin) is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Precedence may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data for anatomical localization of the nuclear medicine data. Precedence may be used either as a separate single photon system, a separate CT system or as a combined CT and single photon system. The nuclear medicine and CT images may be transferred to other systems such as a radiation therapy planning system. The Precedence Imaging System should only be used by trained healthcare professionals.",
      "product_quantity": "1387 in total",
      "reason_for_recall": "During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.",
      "recall_initiation_date": "20171215",
      "center_classification_date": "20180202",
      "termination_date": "20200410",
      "report_date": "20180214",
      "code_info": "Precedence 6 Model # 882351  Serial Numbers :   KP06080001  KP06090002  3000027  3000028  747866A0  KP06020006  30007020021  KP06080010  3000030  KP06090012  KP06100013  KP06110015  30007030020  30007010017  30007030021  30007060024  30007070025  30007050023  30007050022  3000029  3000036  3000002  3000040  3000031  3000032  3000033  3000038  3000035  3000034  3000039  H3000037  3000042  3000041  3000043  3000044  3000046  3000045  30007030016  3000003  3000037  6111502  KP06070007  KP06080008"
    }
  ]
}