  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79088",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-0510-2018",
      "product_description": "BrightView X, Model #882478     Product Usage:  BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only).",
      "product_quantity": "1387 in total",
      "reason_for_recall": "During collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.",
      "recall_initiation_date": "20171215",
      "center_classification_date": "20180202",
      "termination_date": "20200410",
      "report_date": "20180214",
      "code_info": "BrightView X, Model # 882478  Serial Numbers:  11000044  11000006  11000033  11000025  11000105  11000016  11000072  11000116  11000003  11000018  11000066  11000047  11000101  11000065  11000114  11000078  11000051  11000046  11000026  11000007  11000010  11000027  11000032  11340001  11000055  11000070  11000084  11000088  11000068  11000052  11000057  11000050  11000067  11000102  11000061  11000095  11000023  11000094  11000060  11000054  11340004  11000083  11000091  11000098  4000570  11000063  11000062  11000013  11000015  11000035  11000081  11000009  11000017  11000110  11000090  11000107  11340003  11000056  11000029  11000100  11000019  11000034  11000020  11000112  11000086  11000045  11000092  11000037  11000106  11000038  11000059  11000064  11000039  11000087  11000004  11000058  11000030  11340005  11000113  11000117  11000011  11000118  11000074  11000075  11000036  11000104  11000103  11000071  11340002  11000089  11000012  11000041  11000042  11000043  11000049  11000014  11000099  11000022  11000031  11000005  11000028  11000073  11000040  11000077  11000008  11000048  11000093  11000096  11000097  11000069  11000079  11000080  11000024  11000076  11000085  11000108  11000109  11000115"
    }
  ]
}