  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89243",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of NY, PA, and WI.",
      "recall_number": "Z-0501-2022",
      "product_description": "Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300",
      "product_quantity": "5",
      "reason_for_recall": "26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.",
      "recall_initiation_date": "20211216",
      "center_classification_date": "20220114",
      "termination_date": "20230417",
      "report_date": "20220126",
      "code_info": "UDI 04048675041498 Serial Numbers  ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002"
    }
  ]
}