      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89199",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AK, AL, AR, AZ, CA, FL, GU, HI, IA, IL, IN, KS, KY, MA, MI, MO, MS, NC, NE, NJ, NY, OH, OK,PA, PR, SD, TN, TX, WA, WV.",
      "recall_number": "Z-0499-2022",
      "product_description": "1. Sensis, material # 10764561; UDI: 04056869010137  2. Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199  3. Sensis Vibe Combo, material # 11007642; UDI: 04056869010205           The Sensis recording system is a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac, interventional Radiology, and surgical studies.",
      "product_quantity": "1047",
      "reason_for_recall": "Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.",
      "recall_initiation_date": "20211123",
      "center_classification_date": "20220114",
      "report_date": "20220126",
      "code_info": "1.  Sensis- 10764561; UDI: 04056869010137;  Serial Numbers: 103493, 103432, 103466  2.  Sensis Vibe Hemo, material # 11007641; UDI: 04056869010199; Serial Numbers: 103137, 103144, 103453, 103108, 103111, 103047, 103048, 103171, 103182, 103193, 103301, 103170, 103223,, 103225, 103247, 103313, 103315, 103196, 103198, 103166, 103173, 103177, 103184, 103186, 103199, 102326, 103139, 103161, 103167, 103168, 103172, 103178, 100263, 103424, 103319, 103320, 103321, 103323, 103325, 103328, 103330, 103333, 103334, 103335, 103344, 100216, 103337, 100416, 102008, 103413, 101083, 100208, 100209, 102138, 102145, 102147, 103160, 103314, 102199, 103079, 102221, 103123, 103124, 103131, 100163, 100306, 100309, 103300, 101099, 101100, 101101, 103265, 100422, 103019, 103331, 103362, 103393, 103402, 103419, 103252, 103261, 103267, 103269, 103270, 103276, 103277, 103280, 103281, 103022, 103381, 103194 103145, 103148, 103152, 100345, 102183, 103345, 103368, 103415, 103416, 103245, 103249, 103250, 103264, 103080, 103082, 103086, 103088, 103089, 103090, 100231, 103431, 103306, 103341, 103114, 100070, 100071, 100074, 102155, 102157, 103040, 103042, 103039, 102174, 102182, 103130, 103200, 103234, 103116, 103117, 103119, 103288, 103008, 103216, 103465, 103189, 103208, 102010, 102295, 102296, 102302, 100006, 100005, 103352, 103283, 103308, 103310, 103115, 103244, 103246, 103014, 103242, 102143, 103359, 103274, 103437   3.  Sensis Vibe Combo, material # 11007642; UDI: 04056869010205; Serial Numbers: 123021, 123004, 123047, 123053, 122011  ***Updated 1/18/22*** SENSIS Vibe Hemo, Serial #103185  ***Updated 2/24/22*** SENSIS Vibe Hemo Serial #103290 103510 103451 103461 103444 103450 103448 103445 103442 103449 103439 103457   SENSIS Vibe Combo Serial #123055 123056 123058   The following serial numbers have been removed from the scope of this recall:   SENSIS Vibe Hemo Serial #103301 103424  ***Updated 8/5/22***  The following serial numbers have been added to the recall: SENSIS Vibe Combo Serial numbers 103307, 103312"
    }
  ]
}