  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rhondda Cynon Taff",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86484",
      "recalling_firm": "Flexicare Medical Ltd.",
      "address_1": "Cynon Valley Business Park, Mountain Ash",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "US nationwide distribution.",
      "recall_number": "Z-0497-2021",
      "product_description": "ProVu Single Use Video Stylet with ET Tube, Part/Size: 038-990-080U/ 8.0, 038-990-075U/ 7.5, 038-990-070U/ 7.0, 038-990-065U/ 6.5      *Note: product list updated per 12/23/2020 letter to customers",
      "product_quantity": "2087",
      "reason_for_recall": "When the video stylet cuff is over-inflated there is potential for occlusion of the corrugated section of the ETT, and increased resistance to gas flow and difficulty in ventilating the patient.",
      "recall_initiation_date": "20200922",
      "center_classification_date": "20201126",
      "termination_date": "20240729",
      "report_date": "20201202",
      "code_info": "Part/ Lot: 038-990-080U/ 191102901, 200300486, 200300539, 200500068, 200500069, 200803210, 191201567, 200904316, 200904319;     038-990-075U/ 191102900, 200100157, 200300535, 200300536, 200300537, 200402330, 200402331, 200402333, 200402334, 200402335, 200402336, 200402337, 200403976, 200404287, 200500991, 200500992, 200501790;      038-990-070U/ 201000813, 201100558;      038-990-065U/ 201000808, 202002462           *Note: product and lot number list updated per 12/23/2020 letter to customers"
    }
  ]
}