  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Newport Beach",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69737",
      "recalling_firm": "Cure Medical LLC",
      "address_1": "3700 Newport Blvd Ste 301",
      "address_2": "N/A",
      "postal_code": "92663-3946",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-0496-2015",
      "product_description": "Pediatric, 10Fr, Intermittent Urinary Catheters  Model: P10  Pouch Label UPC (P10) *+M405P100C*  Intermediate Box Label UPC (P10) *+M405P103F*  Carton Label UPC (P10) *+M405P105H*    The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches).    There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case.    Intended to be used by pediatric males and females for the purpose of bladder drainage",
      "product_quantity": "192 cases = 1,920 boxes = 57,600 each",
      "reason_for_recall": "Some of the intermediate boxes within lot 140417-3, were labeled as P8, a French Size 8 pediatric catheter but actually contain P10, French Size 10 pediatric catheters.  Trying to use a P10 catheter for bladder drainage, where P8 catheters have been prescribed, may result in acute urinary retention; and, forced attempts to insert the French Size 10 catheter lead to urethral trauma.",
      "recall_initiation_date": "20141111",
      "center_classification_date": "20141202",
      "termination_date": "20150410",
      "report_date": "20141210",
      "code_info": "Lot Number: 140417-3"
    }
  ]
}