      "status": "Terminated",
      "city": "Montreal",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69630",
      "recalling_firm": "Zimmer CAS",
      "address_1": "75 Queen St #3300",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution in states of: AZ, CO, IL, OR, and WA.",
      "recall_number": "Z-0495-2015",
      "product_description": "PSI Knee System- Prosthesis, knee patellofemorotibial, semi-constrained, cemented, polymer/meal/polymer    The Zimmer PSI Knee System is indicated as an orthopedic instrument system to assist in the positioning of knee replacement components.",
      "product_quantity": "15",
      "reason_for_recall": "Internal discovery found that the MRI scans in some PSI knee guides involved an additional rotation adjustments that could potentially lead to misalignment due to a design flaw.",
      "recall_initiation_date": "20141103",
      "center_classification_date": "20141202",
      "termination_date": "20150213",
      "report_date": "20141210",
      "code_info": "BKL0398L 20-8070-002-01;  DHA0645L 20-8070-002-01;  DHA0646R 20-8070-002-02;  ECO0516L 20-8070-002-01;  GHE0651R 20-8070-002-02;  KAD0595L 20-8070-002-01;  KWI0523R 20-8070-002-02;  MHA0483L 20-8070-002-01;  MHA0484R 20-8070-002-02;  MKE0659R 20-8070-002-02;  MMA0728R 20-8070-002-02;  SSC0782L 20-8070-001-01;  TCA0582L 20-8070-002-01;  THO0803L 20-8070-002-01;  TMA0760L 20-8070-002-01;   Expiration dates 5/8/14-12/31/15"
    }
  ]
}