      "country": "United States",
      "city": "Grove City",
      "address_1": "3600 Gantz Rd",
      "reason_for_recall": "This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.",
      "address_2": "",
      "product_quantity": "565 total devices U.S.",
      "code_info": "All Photo Interrupter sensors that are current in AIA-900 analyzers with serial number(S/N) 10000103 to S/N 11147003.",
      "center_classification_date": "20180130",
      "distribution_pattern": "Worldwide Distribution - U.S. Nationwide - Latin America",
      "state": "OH",
      "product_description": "Tosoh¿Automated Enzyme Immunoassay Analyzer AIA-900, Product Catalog No. 022930.  Product Usage:  The AIA-900 Automated Enzyme Immunoassay System is an automated analyzer for in vitro diagnostic use that measures a variety of substances in serum, plasma, and or urine using the AIA-PACK reagent series.",
      "report_date": "20180207",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Tosoh Bioscience Inc",
      "recall_number": "Z-0491-2018",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "78776",
      "termination_date": "20191231",
      "more_code_info": "",
      "recall_initiation_date": "20170611",
      "postal_code": "43123-1895",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}