  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "County Cork",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81122",
      "recalling_firm": "Merck Millipore Ltd.",
      "address_1": "Tullagreen Carrigtwohill",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of:  AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK,OR, PA, RI, TN, TX, VA, VT, WA and WI; and countries of: Canada and Australia.",
      "recall_number": "Z-0489-2019",
      "product_description": "Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15)    Minicon concentrators are non-sterile, disposable, multiwall ultrafiltration devices. Minicon¿ clinical sample concentrators are for in vitro diagnostic use and are intended for concentrating serum, urine, cerebrospinal fluid, and other body fluids prior to analysis.",
      "product_quantity": "1360 units",
      "reason_for_recall": "In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.",
      "recall_initiation_date": "20180514",
      "center_classification_date": "20181120",
      "termination_date": "20210805",
      "report_date": "20181128",
      "code_info": "Lot/Unit Numbers 9031 (Model: Minicon B-15) & 9051 (Model: Minicon CS-15)     9031 (Model: Minicon B-15):  R7HA21782, R7HA25141, R7HA25142, R7HA25143, R7JA25144, R7JA29966, R7KA29967, R7KA37204, R7KA47996, R7KA47998, R7MA62669, R7MA62670, R7MA62671, R7MA70413, R7MA73041, R7NA73042, R7NA76185, R7NA76186, R7NA84720, R7PA90587, R7PA90588, R7SA03064, R8AA13174, R8AA13175, R8AA13176, R8AA17729     9051 (Model: Minicon CS-15):  R7HA32882, R7HA33739, R7HA33741, R7MA53165, R7MA53166, R8AA13177",
      "more_code_info": ""
    }
  ]
}