{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Merrimack",
      "state": "NH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93343",
      "recalling_firm": "Atrium Medical Corporation",
      "address_1": "40 Continental Blvd",
      "address_2": "",
      "postal_code": "03054-4332",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: VA",
      "recall_number": "Z-0486-2024",
      "product_description": "Atrium Oasis Dry Suction Water Seal Chest Drain, Part Number: 3650-100, DRAIN, OASIS BRU W/AC, 3650 ATS BLOOD RECOVERY",
      "product_quantity": "",
      "reason_for_recall": "Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.",
      "recall_initiation_date": "20230905",
      "center_classification_date": "20231201",
      "report_date": "20231213",
      "code_info": "UDI-DI: 20650862113017.  Lots: 461817, 467572, 467572, 468617, 468855, 467191, 467849, 469401, 479253.",
      "more_code_info": ""
    }
  ]
}