  "results": [
    {
      "status": "Ongoing",
      "city": "Lakewood Ranch",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81546",
      "recalling_firm": "Ascom (US) Inc.",
      "address_1": "9024 Town Center Parkway",
      "address_2": "N/A",
      "postal_code": "34202-4159",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX UT, VA, WI",
      "recall_number": "Z-0486-2019",
      "product_description": "Ascom Telligence Nurse Call System",
      "product_quantity": "57 units",
      "reason_for_recall": "There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it.  When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.",
      "recall_initiation_date": "20181102",
      "center_classification_date": "20181120",
      "report_date": "20181128",
      "code_info": "Version 5.0 (Build 5.0.4)",
      "more_code_info": ""
    }
  ]
}