  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Alameda",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66886",
      "recalling_firm": "Abbott Diabetes Care, Inc.",
      "address_1": "1360 S Loop Rd",
      "address_2": "N/A",
      "postal_code": "94502-7000",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Australia, Austria, Belgium, Denmark, Egypt, Finland, France, Germany, Jamaica, Ireland, Israel, United Kingdom, Luxembourg, Netherlands, Norway, Sweden, Switzerland, UAE, and Canada.    **Center Recommended Depth - Consumers/User**",
      "recall_number": "Z-0485-2014",
      "product_description": "FreeStyle Blood Glucose Test Strips;  For in vitro diagnostic testing.  50 count  UPC 6 99073 12050 2;  Freestyle Blood Glucose Test Strips; for Institutional Use only  50 count:  UPC  6 99073 70792 5    Manufactured by Abbott Diabetes Care Inc. Alameda, CA.    The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.",
      "product_quantity": "141,219 cartons for human use - all lot numbers",
      "reason_for_recall": "Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood  Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood  glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.",
      "recall_initiation_date": "20131118",
      "center_classification_date": "20131217",
      "termination_date": "20150803",
      "report_date": "20131225",
      "code_info": "Part number 12050-21:  Lot number: 1285007 Exp. date: 2014/06;  Part number 70792-20 (instuitional use)  Lot number: 1373262, Exp. Date 2015/02"
    }
  ]
}