  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84125",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, IL, IN, KS, LA, MA, MD, Ml, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and PR. Countries of AU, BE, CA, CN, CZ, DK, FR, DE, GT, HK, IT, KR, LU, MX, NL, PK, PL, ZA, ES, SE, CH, TW, and GB.",
      "recall_number": "Z-0484-2020",
      "product_description": "Torcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascular system, including the carotid arteries, in giographic procedures by physicians trained and experienced in angiographic techniques.",
      "product_quantity": "100",
      "reason_for_recall": "The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.",
      "recall_initiation_date": "20191021",
      "center_classification_date": "20191118",
      "termination_date": "20200630",
      "report_date": "20191127",
      "code_info": "REF (RPN): HNB5.0-38-110-P-12S-PED  REF (GPN): G13077  Lot Numbers: 9700036, 9707670  UDIs: (01)00827002130777(17)220509(10)9700036, (01)00827002130777(17)220515(10)9707670"
    }
  ]
}