  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianola",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86619",
      "recalling_firm": "Bayer Medical Care, Inc.",
      "address_1": "1 Bayer Dr",
      "address_2": "N/A",
      "postal_code": "15051-9702",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed to healthcare facilities nationwide throughout the US.",
      "recall_number": "Z-0483-2021",
      "product_description": "Brand Name: MEDRAD Stellant FLEX Syringe Kits    Common Name: Angiographic Syringe    Catalog/REF Number: FLEXD-150-SCS  Description:  2 - 150 mL FLEX syringes, 1 large saline spike, 1 small contrast spike, patient t-connector tubing & prime tube.    Catalog/REF Number: FLEXD-150-SPK  Description: 2 - 150 mL FLEX syringes, 2 small spikes, patient t-connector tubing & prime tube.",
      "product_quantity": "92,420 units",
      "reason_for_recall": "Specific lots of MEDRAD Stellant FLEX  150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.",
      "recall_initiation_date": "20201005",
      "center_classification_date": "20201123",
      "termination_date": "20210319",
      "report_date": "20201202",
      "code_info": "Catalog/REF Number:   FLEXD-150-SCS   Material Number:           87078817   Lot Numbers:                 8584324, 8586383, 8586384, 8586385,                                          8586386, 8586388     UDI Numbers:                            (01)00616258022888(11)200620(17)240620(10)8584324                          (01)00616258022888(11)200620(17)240620(10)8584324                          (01)00616258022888(11)200620(17)240620(10)8584324                          (01)00616258022888(11)200620(17)240620(10)8584324                          (01)00616258022888(11)200620(17)240620(10)8584324                          (01)00616258022888(11)200620(17)240620(10)8584324    Catalog/REF Number:  FLEXD-150-SPK  Material Number:          85020218  Lot Number:                  8585234    UDI Numbers: (01)00616258010656(11)200720(17)240720(10)8585234"
    }
  ]
}