      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81224",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US Nationwide, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Honduras, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.",
      "recall_number": "Z-0482-2019",
      "product_description": "Panda iRes Infant Radiant Warmer. Device sold under the following product names: Panda iRes Warmer and Panda iRes Bedded Warmer.",
      "product_quantity": "1,724 devices total",
      "reason_for_recall": "Potential electric leakage affecting the user.",
      "recall_initiation_date": "20180925",
      "center_classification_date": "20181119",
      "termination_date": "20200713",
      "report_date": "20181128",
      "code_info": "Model Number, UDI (Serial Number):  M1112198-526395, 010084068210389321PBWX7040911180413 (PBWX70409);    M1112198-164657, Not Applicable (HDJT50683);    M1112198-018125, Not Applicable (HDJM50048);    M1112198-023789, Not Applicable (HDJM50529);    M1112198-023787, Not Applicable (HDJM50528);    M1112198-027124, Not Applicable (HDJP50065);    M1112198-018125, Not Applicable (HDJM50047);    M1112198-022592, Not Applicable (HDJM50596);    M1112198-023488, Not Applicable (HDJQ50744);    M1112198-024139, Not Applicable (HDJQ50819);    M1112198-023359, Not Applicable (HDJR50385);    M1112198-023359, Not Applicable (HDJR50389);    M1112198-023359, Not Applicable (HDJR50382);    M1112198-171607, Not Applicable (HDJS50605);    M1112198-023391, Not Applicable (HDJM50439);    M1112198-052019, Not Applicable (HDJR50487);    M1112198-018105, Not Applicable (HDJM50140);    M1112198-018105, Not Applicable (HDJM50139);    M1112198-019900, Not Applicable (HDJM50471);    M1112198-018105, Not Applicable (HDJM50138);    M1112198-527790, 010084068210389321PBWX7065611180521 (PBWX70656);    M1112198-164657, Not Applicable (HDJT51749);    M1112198-022124, Not Applicable (HDJM50440);    M1112198-018102, Not Applicable (HDJM50183);    M1112198-023608, Not Applicable (HDJQ50986);    M1112198-023608, Not Applicable (HDJQ50988);    M1112198-023608, Not Applicable (HDJQ50992);    M1112198-023608, Not Applicable (HDJQ50993);    M1112198-023608, Not Applicable (HDJQ50994);    M1112198-023608, Not Applicable (HDJQ50996);    M1112198-023608, Not Applicable (HDJQ50998);    M1112198-023608, Not Applicable (HDJQ50999);    M1112198-023608, Not Applicable (HDJQ51000);    M1112198-023608, Not Applicable (HDJQ50991);    M1112198-526395, 010084068210389321PBWX7050111180501 (PBWX70501);    M1112198-526773, 010084068210389321PBWW6086211170410 (PBWW60862);    M1112198-024139, Not Applicable (HDJM50629);    M1112198-024139, Not Applicable (HDJM50628);    M1112198-164256, Not Applicable (HDJT51641);    M1112198-184659, Not Applicable (HDJT60033);    M1112198-022592, Not Applicable (HDJM51338);    M1112198-022592, Not Applicable (HDJM51337);    M1112198-164657, Not Applicable (HDJT61440);    M1112198-164657, Not Applicable (HDJT61452);    M1112198-023848, Not Applicable (HDJQ52172);    M1112198-164256, Not Applicable (HDJT50739);    M1112198-018697, Not Applicable (HDJM50290);    M1112198-018697, Not Applicable (HDJM50291);    M1112198-024139, Not Applicable (HDJP52571);    M1112198-023488, Not Applicable (HDJM51404);    M1112198-023488, Not Applicable (HDJM51607);    M1112198-026336, Not Applicable (HDJM51242);    M1112198-022697, Not Applicable (HDJN50090);    M1112198-022822, Not Applicable (HDJM50492);    M1112198-022822, Not Applicable (HDJM50493);    M1112198-022822, Not Applicable (HDJM50495);    M1112198-022822, Not Applicable (HDJM50494);    M1112198-497258, 010084068210389321PBWV6064511160725 (PBWV60645);    M1112198-164256, Not Applicable (HDJT50740);    M1112198-023848, Not Applicable (HDJM50541);    M1112198-023848, Not Applicable (HDJM50540);    M1112198-023849, Not Applicable (HDJM50539);    M1112198-022592, Not Applicable (HDJM50602);    M1112198-022592, Not Applicable (HDJM50853);    M1112198-022592, Not Applicable (HDJM50618);    M1112198-022592, Not Applicable (HDJM50604);    M1112198-022592, Not Applicable (HDJM50594);    M1112198-022592, Not Applicable (HDJM50851);    M1112198-022592, Not Applicable (HDJM50852);    M1112198-022592, Not Applicable (HDJM50849);    M1112198-022592, Not Applicable (HDJM50603);    M1112198-022592, Not Applicable (HDJM50601);    M1112198-022592, Not Applicable (HDJM50587);    M1112198-022592, Not Applicable (HDJM50635);    M1112198-022592, Not Applicable (HDJM50610);    M1112198-022592, Not Applicable (HDJM50843);    M1112198-022592, Not Applicable (HDJM50856);    M1112198-019213, Not Applicable (HDJL50321);    M1112198-019213, Not Applicable (HDJL50320);    M1112198-019213, Not Applicable (HDJL50322);    M1112198-019213, Not Applicable (HDJL50323);    M1112198-024139, Not Applicable (HDJQ50738);    M1112198-526773, 010084068210389321PBWX7019811180306 (PBWX70198);    M1112198-526773, 010084068210389321PBWX7021611180308 (PBWX70216);    M1112198-526773, 010084068210389321PBWX7073611180531 (PBWX70736);    M1112198-526773, 010084068210389321PBWX7073811180601 (PBWX70738);    M1112198-526773, 010084068210389321PBWX7074911180603 (PBWX70749);    M1112198-175579, Not Applicable (HDJT51602);    M1112198-023848, Not Applicable (HDJM50700);    M1112198-025718, Not Applicable (HDJM51237);    M1112198-024139, Not Applicable (HDJM50703);    M1112198-024139, Not Applicable (HDJM50702);    M1112198-022592, Not Applicable (HDJM50634);    M1112198-022592, Not Applicable (HDJM51766);    M1112198-023608, Not Applicable (HDJM50497);    M1112198-023608, Not Applicable (HDJM50499);    M1112198-022592, Not Applicable (HDJM51768);    M1112198-022592, Not Applicable (HDJM51767);    M1112198-022592, Not Applicable (HDJM51770);    M1112198-022592, Not Applicable (HDJM51771);    M1112198-023608, Not Applicable (HDJM50500);    M1112198-023608, Not Applicable (HDJM50498);    M1112198-023608, Not Applicable (HDJM50501);    M1112198-023608, Not Applicable (HDJM50502);    M1112198-022592, Not Applicable (HDJM50855);    M1112198-018125, Not Applicable (HDJL50327);    M1112198-022592, Not Applicable (HDJQ50074);    M1112198-022592, Not Applicable (HDJQ50069);    M1112198-022592, Not Applicable (HDJQ50071);    M1112198-179805, Not Applicable (HDJU66024);    M1112198-023848, Not Applicable (HDJQ50658);    M1112198-023848, Not Applicable (HDJR50416);    M1112198-526395, 010084068210389321PBWX7051411180502 (PBWX70514);    M1112198-526395, 010084068210389321PBWX7087711180617 (PBWX70877);    M1112198-024139, Not Applicable (HDJQ50367);    M1112198-165361, Not Applicable (HDJT61403);    M1112198-165361, Not Applicable (HDJT61404);    M1112198-024139, Not Applicable (HDJM51922);    M1112198-526773, 010084068210389321PBWW6086611170410 (PBWW60866);    M1112198-526773, 010084068210389321PBWW6086311170410 (PBWW60863);    M1112198-022070, Not Applicable (HDJM50181);    M1112198-022592, Not Applicable (HDJM50611);    M1112198-022592, Not Applicable (HDJM50854);    M1112198-022592, Not Applicable (HDJM50847);    M1112198-022592, Not Applicable (HDJM50842);    M1112198-022592, Not Applicable (HDJM50595);    M1112198-022592, Not Applicable (HDJM50590);    M1112198-022592, Not Applicable (HDJM50591);    M1112198-022592, Not Applicable (HDJM50657);    M1112198-022592, Not Applicable (HDJM50636);    M1112198-022592, Not Applicable (HDJM50848);    M1112198-022592, Not Applicable (HDJM50844);    M1112198-022592, Not Applicable (HDJM50837);    M1112198-022592, Not Applicable (HDJM50608);    M1112198-022592, Not Applicable (HDJM50600);    M1112198-022592, Not Applicable (HDJM50593);    M1112198-022592, Not Applicable (HDJM50597);    M1112198-022592, Not Applicable (HDJM50586);    M1112198-022592, Not Applicable (HDJM50846);    M1112198-022592, Not Applicable (HDJM50839);    M1112198-022592, Not Applicable (HDJM50836);    M1112198-022592, Not Applicable (HDJM50588);    M1112198-022592, Not Applicable (HDJM50614);    M1112198-022592, Not Applicable (HDJM50613);    M1112198-022592, Not Applicable (HDJM50589);    M1112198-022592, Not Applicable (HDJM50612);    M1112198-022592, Not Applicable (HDJM50658);    M1112198-022592, Not Applicable (HDJM50642);    M1112198-022592, Not Applicable (HDJM50643);    M1112198-022592, Not Applicable (HDJM50841);    M1112198-022592, Not Applicable (HDJM50489);    M1112198-022592, Not Applicable (HDJM50617);    M1112198-022592, Not Applicable (HDJM50641);    M1112198-022592, Not Applicable (HDJM50656);    M1112198-526395, 010084068210389321PBWX7057011180510 (PBWX70570);    M1112198-526395, 010084068210389321PBWX7057511180510 (PBWX70575);    M1112198-526395, 010084068210389321PBWX7064411180518 (PBWX70644);    M1112198-526395, 010084068210389321PBWX7063611180517 (PBWX70636);    M1112198-164657, Not Applicable (HDJT60402);    M1112198-164657, Not Applicable (HDJT60407);    M1112198-024139, Not Applicable (HDJM51179);    M1112198-024139, Not Applicable (HDJM51180);    M1112198-526773, 010084068210389321PBWW6123611170515 (PBWW61236);    M1112198-164657, Not Applicable (HDJT51040);    M1112198-164256, Not Applicable (HDJT60431);    M1112198-164256, Not Applicable (HDJT60450);    M1112198-175579, Not Applicable (HDJT50476);    M1112198-526395, 010084068210389321PBWX6058411180601 (PBWX60584);    M1112198-175579, Not Applicable (HDJT50620);    M1112198-175579, Not Applicable (HDJT51433);    M1112198-175579, Not Applicable (HDJT51432);    M1112198-175579, Not Applicable (HDJT50622);    M1112198-023848, Not Applicable (HDJM50699);    M1112198-022592, Not Applicable (HDJM50605);    M1112198-022592, Not Applicable (HDJM50609);    M1112198-022592, Not Applicable (HDJM50616);    M1112198-022592, Not Applicable (HDJM50598);    M1112198-022592, Not Applicable (HDJM50592);    M1112198-022592, Not Applicable (HDJM50599);    M1112198-022592, Not Applicable (HDJM50644);    M1112198-022592, Not Applicable (HDJM50640);    M1112198-022592, Not Applicable (HDJM50607);    M1112198-022592, Not Applicable (HDJM50639);    M1112198-022592, Not Applicable (HDJM50606);    M1112198-025718, Not Applicable (HDJM51454);    M1112198-025718, Not Applicable (HDJM51455);    M1112198-022697, Not Applicable (HDJM50300);    M1112198-022697, Not Applicable (HDJM50297);    M1112198-022697, Not Applicable (HDJM50295);    M1112198-022697, Not Applicable (HDJM50301);    M1112198-022697, Not Applicable (HDJM50299);    M1112198-022697, Not Applicable (HDJM50298);    M1112198-022697, Not Applicable (HDJM50293);    M1112198-022697, Not Applicable (HDJM50296);    M1112198-022697, Not Applicable (HDJM50294);    M1112198-526773, 010084068210389321PBWW6086411170410 (PBWW60864);    M1112198-023359, Not Applicable (HDJM50415);    M1112198-023359, Not Applicable (HDJM50416);    M1112198-023359, Not Applicable (HDJM50418);    M1112198-023359, Not Applicable (HDJM50417);    M1112198-025987, Not Applicable (HDJN50470);    M1112198-018505, Not Applicable (HDJM50039);    M1112198-018505, Not Applicable (HDJM50031);    M1112198-018505, Not Applicable (HDJM50038);    M1112198-018505, Not Applicable (HDJM50042);    M1112198-018505, Not Applicable (HDJM50030);    M1112198-018505, Not Applicable (HDJM50040);    M1112198-018505, Not Applicable (HDJM50041);    M1112198-018505, Not Applicable (HDJM50033);    M1112198-018505, Not Applicable (HDJM50032);    M1112198-018505, Not Applicable (HDJM50034);    M1112198-018505, Not Applicable (HDJM50035);    M1112198-018505, Not Applicable (HDJM50037);    M1112198-018505, Not Applicable (HDJM50036);    M1112198-025849, Not Applicable (HDJQ50490);    M1112198-527554, 010084068210389321PBWX7000611180104 (PBWX70006);    M1112198-164256, Not Applicable (HDJT60421);    M1112198-164256, Not Applicable (HDJT51859);    M1112198-165121, Not Applicable (HDJT60292);    M1112198-165121, Not Applicable (HDJT60827);    M1112198-195382, Not Applicable (HDJT51837);    M1112198-018125, Not Applicable (HDJM50100);    M1112198-024139, Not Applicable (HDJM50778);    M1112198-024139, Not Applicable (HDJM50776);    M1112198-024139, Not Applicable (HDJM50787);    M1112198-526395, 010084068210389321PBWX7006911180202 (PBWX70069);    M1112198-526395, 010084068210389321PBWX7007211180205 (PBWX70072);    M1112198-025336, Not Applicable (HDJM50968);    M1112198-157601, Not Applicable (HDJT51512);    M1112198-526395, 010084068210389321PBWW6125011170516 (PBWW61250);    M1112198-526395, 010084068210389321PBWW6124911170516 (PBWW61249);    M1112198-023488, Not Applicable (HDJM50477);    M1112198-018116, Not Applicable (HDJM50280);    M1112198-018116, Not Applicable (HDJM50278);    M1112198-018116, Not Applicable (HDJM50279);    M1112198-027124, Not Applicable (HDJQ50478);    M1112198-305577, Not Applicable (HDJT51784);    M1112198-176923, Not Applicable (HDJT51790);    M1112198-305577, Not Applicable (HDJT51781);    M1112198-305577, Not Applicable (HDJT51785);    M1112198-305577, Not Applicable (HDJT51786);    M1112198-176923, Not Applicable (HDJT51791);    M1112198-305577, Not Applicable (HDJT51782);    M1112198-305577, Not Applicable (HDJT51787);    M1112198-184659, Not Applicable (HDJT51866);    M1112198-184659, Not Applicable (HDJT51913);    M1112198-184659, Not Applicable (HDJT60043);    M1112198-528031, 010084068210389321PBWX7070111180529 (PBWX70701);    M1112198-528031, 010084068210389321PBWX7070611180530 (PBWX70706);    M1112198-528031, 010084068210389321PBWX7070911180530 (PBWX70709);    M1112198-528031, 010084068210389321PBWX7071411180530 (PBWX70714);    M1112198-023848, Not Applicable (HDJP51927);    M1112198-171343, Not Applicable (HDJT51283);    M1112198-018102, Not Applicable (HDJM50096);    M1112198-018125, Not Applicable (HDJM50049);    M1112198-532026, 010084068210389321PBWX7025811180315 (PBWX70258);    M1112198-025875, Not Applicable (HDJM51188);    M1112198-025875, Not Applicable (HDJM51189);    M1112198-025399, Not Applicable (HDJM51043);    M1112198-025399, Not Applicable (HDJM51044);    M1112198-057055, Not Applicable (HDJQ50092);    M1112198-526395, 010084068210389321PBWW6124711170516 (PBWW61247);    M1112198-024585, Not Applicable (HDJQ50672);    M1112198-184659, Not Applicable (HDJT51870);    M1112198-184659, Not Applicable (HDJT51369);    M1112198-024139, Not Applicable (HDJM51612);    M1112198-164657, Not Applicable (HDJT51455);    M1112198-164657, Not Applicable (HDJT51980);    M1112198-022822, Not Applicable (HDJQ50733);    M1112198-191061, Not Applicable (HDJT61575);    M1112198-022592, Not Applicable (HDJR50548);    M1112198-527554, 010084068210389321PBWW6122011170510 (PBWW61220);    M1112198-023528, Not Applicable (HDJM50485);    M1112198-023528, Not Applicable (HDJM50484);    M1112198-023528, Not Applicable (HDJM50487);    M1112198-023528, Not Applicable (HDJM50486);    M1112198-514004, 010084068210389321PBWW6040811170216 (PBWW60408);    M1112198-184659, Not Applicable (HDJT51298);    M1112198-025987, Not Applicable (HDJM51300);    M1112198-196063, Not Applicable (HDJT60628);    M1112198-022592, Not Applicable (HDJM50637);    M1112198-030895, Not Applicable (HDJQ50065);    M1112198-023848, Not Applicable (HDJQ50825);    M1112198-098077, Not Applicable (HDJR50519);    M1112198-184659, Not Applicable (HDJT51299);    M1112198-023848, Not Applicable (HDJQ50106);    M1112198-054539, Not Applicable (HDJP51915);    M1112198-022592, Not Applicable (HDJM50473);    M1112198-022592, Not Applicable (HDJM50615);    M1112198-179222, Not Applicable (HDJT51776);    M1112198-179222, Not Applicable (HDJT51753);    M1112198-179222, Not Applicable (HDJT51777);    M1112198-179222, Not Applicable (HDJT51775);    M1112198-163959, Not Applicable (HDJT51645);    M1112198-018115, Not Applicable (HDJM50029);    M1112198-043775, Not Applicable (HDJQ50283);    M1112198-043775, Not Applicable (HDJQ50285);    M1112198-043775, Not Applicable (HDJQ50286);    M1112198-025718, Not Applicable (HDJM51056);    M1112198-526395, 010084068210389321PBWX7078811180607 (PBWX70788);    M1112198-018125, Not Applicable (HDJM50051);    M1112198-022592, Not Applicable (HDJQ50722);    M1112198-022592, Not Applicable (HDJQ50723);    M1112198-022592, Not Applicable (HDJQ50724);    M1112198-022592, Not Applicable (HDJQ50726);    M1112198-022592, Not Applicable (HDJQ50725);    M1112198-022592, Not Applicable (HDJQ50730);    M1112198-022070, Not Applicable (HDJM50242);    M1112198-064618, Not Applicable (HDJP52845);    M1112198-022059, Not Applicable (HDJM50355);    M1112198-022059, Not Applicable (HDJM50356);    M1112198-022059, Not Applicable (HDJM50357);    M1112198-022059, Not Applicable (HDJM50358);    M1112198-023608, Not Applicable (HDJQ50518);    M1112198-023359, Not Applicable (HDJM50731);    M1112198-023848, Not Applicable (HDJQ50657);    M1112198-023848, Not Applicable (HDJQ50737);    M1112198-023848, Not Applicable (HDJQ50505);    M1112198-164256, Not Applicable (HDJT51858);    M1112198-025849, Not Applicable (HDJM51399);    M1112198-032131, Not Applicable (HDJN50085);    M1112198-021534, Not Applicable (HDJM50345);    M1112198-025849, Not Applicable (HDJM51398);    M1112198-184659, Not Applicable (HDJT51562);    M1112198-184659, Not Applicable (HDJT51372);    M1112198-021233, Not Applicable (HDJM50353);    M1112198-021233, Not Applicable (HDJM50354);    M1112198-024139, Not Applicable (HDJQ50372);    M1112198-022592, Not Applicable (HDJM50638);    M1112198-026790, Not Applicable (HDJM51453);    M1112198-163959, Not Applicable (HDJT51343);    M1112198-163959, Not Applicable (HDJT51335);    M1112198-023532, Not Applicable (HDJM50474);    M1112198-022592, Not Applicable (HDJN51787);    M1112198-023848, Not Applicable (HDJM51395);    M1112198-022068, Not Applicable (HDJM50310);    M1112198-022789, Not Applicable (HDJM50317);    M1112198-295935, Not Applicable (HDJT51596);    M1112198-025584, Not Applicable (HDJM51055);    M1112198-526773, 010084068210389321PBWX7073511180531 (PBWX70735);    M1112198-030895, Not Applicable (HDJR50371);    M1112198-021233, Not Applicable (HDJM50129);    M1112198-021233, Not Applicable (HDJM50131);    M1112198-021233, Not Applicable (HDJM50132);    M1112198-021233, Not Applicable (HDJM50133);    M1112198-021233, Not Applicable (HDJM50135);    M1112198-021233, Not Applicable (HDJM50134);    M1112198-021233, Not Applicable (HDJM50127);    M1112198-021233, Not Applicable (HDJM50126);    M1112198-021233, Not Applicable (HDJM50128);    M1112198-021233, Not Applicable (HDJM50130);    M1112198-021233, Not Applicable (HDJM50137);    M1112198-021233, Not Applicable (HDJM50136);    M1112198-191821, Not Applicable (HDJT51383);    M1112198-191821, Not Applicable (HDJT51384);    M1112198-023431, Not Applicable (HDJM50453);    M1112198-023432, Not Applicable (HDJM50455);    M1112198-023430, Not Applicable (HDJM50447);    M1112198-023431, Not Applicable (HDJM50452);    M1112198-023433, Not Applicable (HDJM50458);    M1112198-023433, Not Applicable (HDJM50461);    M1112198-023433, Not Applicable (HDJM50463);    M1112198-023433, Not Applicable (HDJM50462);    M1112198-023431, Not Applicable (HDJM50450);    M1112198-023430, Not Applicable (HDJM50448);    M1112198-023432, Not Applicable (HDJM50454);    M1112198-023433, Not Applicable (HDJM50456);    M1112198-023433, Not Applicable (HDJM50459);    M1112198-023433, Not Applicable (HDJM50460);    M1112198-023430, Not Applicable (HDJM50449);    M1112198-023433, Not Applicable (HDJM50457);    M1112198-286123, Not Applicable (HDJT61566);    M1112198-286123, Not Applicable (HDJT61565);    M1112198-526773, 010084068210389321PBWX7088111180618 (PBWX70881);    M1112198-175579, Not Applicable (HDJT50955);    M1112198-175579, Not Applicable (HDJT50948);    M1112198-175579, Not Applicable (HDJT50956);    M1112198-175579, Not Applicable (HDJT50958);    M1112198-175579, Not Applicable (HDJT50951);    M1112198-022697, Not Applicable (HDJM51516);    M1112198-022697, Not Applicable (HDJM51520);    M1112198-022697, Not Applicable (HDJM51513);    M1112198-022697, Not Applicable (HDJM51518);    M1112198-022697, Not Applicable (HDJM51517);    M1112198-020611, Not Applicable (HDJL50401);    M1112198-020611, Not Applicable (HDJL50402);    M1112198-018505, Not Applicable (HDJL50466);    M1112198-018505, Not Applicable (HDJL50471);    M1112198-018505, Not Applicable (HDJL50467);    M1112198-018505, Not Applicable (HDJL50470);    M1112198-527554, 010084068210389321PBWX7038011180409 (PBWX70380);    M1112198-022592, Not Applicable (HDJR56417);    M1112198-526395, 010084068210389321PBWW7000511171101 (PBWW70005);    M1112198-175579, Not Applicable (HDJT50953);    M1112198-023848, Not Applicable (HDJR50410);    M1112198-023848, Not Applicable (HDJR50414);    M1112198-022592, Not Applicable (HDJM50840);    M1112198-023179, Not Applicable (HDJM50413);    M1112198-170328, Not Applicable (HDJT51944);    M1112198-025849, Not Applicable (HDJM51147);    M1112198-025849, Not Applicable (HDJM51144);    M1112198-025849, Not Applicable (HDJM51145);    M1112198-025849, Not Applicable (HDJM51146);    M1112198-025849, Not Applicable (HDJM51148);    M1112198-025849, Not Applicable (HDJR50344);    M1112198-025849, Not Applicable (HDJR50347);    M1112198-025849, Not Applicable (HDJR50343);    M1112198-022562, Not Applicable (HDJM50315);    M1112198-295602, Not Applicable (HDJT51434);    M1112198-178214, Not Applicable (HDJT51464);    M1112198-026336, Not Applicable (HDJM51396);    M1112198-026336, Not Applicable (HDJM51330);    M1112198-040913, Not Applicable (HDJQ50757);    M1112198-526395, 010084068210389321PBWX7041111180413 (PBWX70411);    M1112198-022563, Not Applicable (HDJM50312);    M1112198-022563, Not Applicable (HDJM50313);    M1112198-026336, Not Applicable (HDJM51332);    M1112198-024139, Not Applicable (HDJM51610);    M1112198-165121, Not Applicable (HDJT61530);    M1112198-179222, Not Applicable (HDJT51752);    M1112198-179222, Not Applicable (HDJT51774);    M1112198-024139, Not Applicable (HDJM51615);    M1112198-024139, Not Applicable (HDJM51608);    M1112198-024139, Not Applicable (HDJM51614);    M1112198-179222, Not Applicable (HDJT51395);    M1112198-021211, Not Applicable (HDJM50284);    M1112198-021211, Not Applicable (HDJM50281);    M1112198-021211, Not Applicable (HDJM50283);    M1112198-021211, Not Applicable (HDJM50287);    M1112198-021211, Not Applicable (HDJM50285);    M1112198-021211, Not Applicable (HDJM50282);    M1112198-021211, Not Applicable (HDJM50286);    M1112198-528361, 010084068210389321PBWX7023111180309 (PBWX70231);    M1112198-018505, Not Applicable (HDJL50468);    M1112198-018505, Not Applicable (HDJL50509);    M1112198-018505, Not Applicable (HDJL50510);    M1112198-527554, 010084068210389321PBWX7071511180530 (PBWX70715);    M1112198-025620, Not Applicable (HDJM51318);    M1112198-025620, Not Applicable (HDJM51317);    M1112198-025620, Not Applicable (HDJM51313);    M1112198-025620, Not Applicable (HDJM51312);    M1112198-025620, Not Applicable (HDJM51319);    M1112198-025620, Not Applicable (HDJM51316);    M1112198-025620, Not Applicable (HDJM51314);    M1112198-025620, Not Applicable (HDJM51315);    M1112198-289845, Not Applicable (HDJT51469);    M1112198-022697, Not Applicable (HDJN52089);    M1112198-022697, Not Applicable (HDJN52090);    M1112198-019563, Not Applicable (HDJL50428);    M1112198-187072, Not Applicable (HDJT61335);    M1112198-024139, Not Applicable (HDJM51020);    M1112198-018125, Not Applicable (HDJM50110);    M1112198-024139, Not Applicable (HDJM51022);    M1112198-024139, Not Applicable (HDJM51021);    M1112198-024139, Not Applicable (HDJM51019);    M1112198-018125, Not Applicable (HDJM50113);    M1112198-018125, Not Applicable (HDJM50114);    M1112198-018125, Not Applicable (HDJM50117);    M1112198-018125, Not Applicable (HDJM50118);    M1112198-018125, Not Applicable (HDJM50121);    M1112198-018125, Not Applicable (HDJM50119);    M1112198-018125, Not Applicable (HDJM50122);    M1112198-018125, Not Applicable (HDJM50109);    M1112198-018125, Not Applicable (HDJM50108);    M1112198-018125, Not Applicable (HDJM50111);    M1112198-018125, Not Applicable (HDJM50116);    M1112198-018125, Not Applicable (HDJM50115);    M1112198-018505, Not Applicable (HDJL50469);    M1112198-018926, Not Applicable (HDJL50420);    M1112198-030803, Not Applicable (HDJR56340);    M1112198-023359, Not Applicable (HDJN50677);    M1112198-018505, Not Applicable (HDJM50302);    M1112198-023359, Not Applicable (HDJM51098);    M1112198-164256, Not Applicable (HDJT50810);    M1112198-173399, Not Applicable (HDJT51365);    M1112198-033981, Not Applicable (HDJQ50821);    M1112198-033981, Not Applicable (HDJQ50822);    M1112198-022066, Not Applicable (HDJM50346);    M1112198-022632, Not Applicable (HDJM50348);    M1112198-025875, Not Applicable (HDJM51302);    M1112198-025875, Not Applicable (HDJM51301);    M1112198-203578, Not Applicable (HDJU60478);    M1112198-276838, Not Applicable (HDJT51325);    M1112198-276838, Not Applicable (HDJT51324);    M1112198-184659, Not Applicable (HDJT51869);    M1112198-539650, 010084068210389321PBWX7032711180327 (PBWX70327);    M1112198-179222, Not Applicable (HDJT51296);    M1112198-022697, Not Applicable (HDJN50627);    M1112198-183037, Not Applicable (HDJT51346);    M1112198-183037, Not Applicable (HDJT51348);    M1112198-526773, 010084068210389321PBWW6129911170518 (PBWW61299);    M1112198-526773, 010084068210389321PBWW6130411170518 (PBWW61304);    M1112198-025620, Not Applicable (HDJM51060);    M1112198-526395, 010084068210389321PBWW6124611170516 (PBWW61246);    M1112198-526395, 010084068210389321PBWW6125211170516 (PBWW61252);    M1112198-018926, Not Applicable (HDJL50408);    M1112198-018926, Not Applicable (HDJL50417);    M1112198-018926, Not Applicable (HDJL50414);    M1112198-018926, Not Applicable (HDJL50415);    M1112198-018926, Not Applicable (HDJL50419);    M1112198-018926, Not Applicable (HDJL50412);    M1112198-018926, Not Applicable (HDJL50416);    M1112198-018926, Not Applicable (HDJL50413);    M1112198-018926, Not Applicable (HDJL50421);    M1112198-066022, Not Applicable (HDJQ50098);    M1112198-526395, 010084068210389321PBWX7058211180511 (PBWX70582);    M1112198-164256, Not Applicable (HDJT51560);    M1112198-164256, Not Applicable (HDJT51632);    M1112198-164256, Not Applicable (HDJT51637);    M1112198-059620, Not Applicable (HDJR50421);    M1112198-187203, Not Applicable (HDJT60469);    M1112198-025399, Not Applicable (HDJM51039);    M1112198-025399, Not Applicable (HDJM51040);    M1112198-025399, Not Applicable (HDJM51041);    M1112198-025399, Not Applicable (HDJM51045);    M1112198-025399, Not Applicable (HDJM51038);    M1112198-025399, Not Applicable (HDJM51047);    M1112198-025399, Not Applicable (HDJM51042);    M1112198-025399, Not Applicable (HDJM51178);    M1112198-025399, Not Applicable (HDJM51037);    M1112198-025399, Not Applicable (HDJM51046);    M1112198-025399, Not Applicable (HDJM51048);    M1112198-025399, Not Applicable (HDJM51177);    M1112198-163959, Not Applicable (HDJT51247);    M1112198-184659, Not Applicable (HDJT51370);    M1112198-184659, Not Applicable (HDJT51375);    M1112198-024139, Not Applicable (HDJM51403);    M1112198-022465, Not Applicable (HDJM50316);    M1112198-022592, Not Applicable (HDJM50850);    M1112198-184659, Not Applicable (HDJT51867);    M1112198-184659, Not Applicable (HDJT51865);    M1112198-184659, Not Applicable (HDJT51914);    M1112198-024139, Not Applicable (HDJM50784);    M1112198-024139, Not Applicable (HDJM50775);    M1112198-024139, Not Applicable (HDJM50780);    M1112198-024139, Not Applicable (HDJM50783);    M1112198-024139, Not Applicable (HDJM50786);    M1112198-024139, Not Applicable (HDJM50773);    M1112198-024139, Not Applicable (HDJM50802);    M1112198-024139, Not Applicable (HDJM50782);    M1112198-024139, Not Applicable (HDJM50777);    M1112198-024139, Not Applicable (HDJM50779);    M1112198-024139, Not Applicable (HDJM50774);    M1112198-024139, Not Applicable (HDJM50785);    M1112198-024139, Not Applicable (HDJM50804);    M1112198-024139, Not Applicable (HDJM51018);    M1112198-024139, Not Applicable (HDJM51016);    M1112198-024139, Not Applicable (HDJM50772);    M1112198-024139, Not Applicable (HDJM51017);    M1112198-024139, Not Applicable (HDJQ50804);    M1112198-526438, 010084068210389321PBWX7143511180808 (PBWX71435);    M1112198-164256, Not Applicable (HDJT60283);    M1112198-164256, Not Applicable (HDJT60426);    M1112198-023848, Not Applicable (HDJQ50914);    M1112198-566956, 010084068210389321PBWX7091111180622 (PBWX70911);    M1112198-024139, Not Applicable (HDJQ50820);    M1112198-018102, Not Applicable (HDJM50182);    M1112198-043525, Not Applicable (HDJP50011);    M1112198-025718, Not Applicable (HDJM51238);    M1112198-025718, Not Applicable (HDJM51239);    M1112198-527554, 010084068210389321PBWX7082611180612 (PBWX70826);    M1112198-527554, 010084068210389321PBWX7083611180613 (PBWX70836);    M1112198-287486, Not Applicable (HDJU66023);    M1112198-287486, Not Applicable (HDJT51379);    M1112198-024139, Not Applicable (HDJM51447);    M1112198-175579, Not Applicable (HDJT60803);    M1112198-529431, 010084068210389321PBWX7045711180420 (PBWX70457);    M1112198-023848, Not Applicable (HDJN50006);    M1112198-018125, Not Applicable (HDJM50106);    M1112198-018125, Not Applicable (HDJM50107);    M1112198-526773, 010084068210389321PBWW6129711170518 (PBWW61297);    M1112198-171343, Not Applicable (HDJT50821);    M1112198-171343, Not Applicable (HDJT50805);    M1112198-171343, Not Applicable (HDJT50797);    M1112198-022592, Not Applicable (HDJM50390);    M1112198-024047, Not Applicable (HDJM51287);    M1112198-024047, Not Applicable (HDJM51289);    M1112198-024047, Not Applicable (HDJM51285);    M1112198-024047, Not Applicable (HDJM51284);    M1112198-024047, Not Applicable (HDJM51293);    M1112198-024047, Not Applicable (HDJM51286);    M1112198-024047, Not Applicable (HDJM51291);    M1112198-024047, Not Applicable (HDJM51292);    M1112198-024047, Not Applicable (HDJM51282);    M1112198-024047, Not Applicable (HDJM51294);    M1112198-024047, Not Applicable (HDJM51288);    M1112198-024047, Not Applicable (HDJM51290);    M1112198-024047, Not Applicable (HDJM51283);    M1112198-024047, Not Applicable (HDJM51295);    M1112198-018125, Not Applicable (HDJM50102);    M1112198-163959, Not Applicable (HDJT51713);    M1112198-163959, Not Applicable (HDJT51793);    M1112198-163959, Not Applicable (HDJT51794);    M1112198-499665, 010084068210389321PBWW6036811170209 (PBWW60368);    M1112198-023848, Not Applicable (HDJP52622);    M1112198-023848, Not Applicable (HDJP52618);    M1112198-164657, Not Applicable (HDJT50887);    M1112198-163959, Not Applicable (HDJT51336);    M1112198-165361, Not Applicable (HDJT60229);    M1112198-020820, Not Applicable (HDJM50276);    M1112198-033971, Not Applicable (HDJN50672);    M1112198-284422, Not Applicable (HDJT50875);    M1112198-526438, 010084068210389321PBWX7042811180416 (PBWX70428);    M1112198-022697, Not Applicable (HDJN50202);    M1112198-022697, Not Applicable (HDJM50953);    M1112198-023848, Not Applicable (HDJQ50827);    M1112198-023848, Not Applicable (HDJQ50836);    M1112198-023848, Not Applicable (HDJQ50831);    M1112198-023848, Not Applicable (HDJQ50830);    M1112198-023848, Not Applicable (HDJQ50834);    M1112198-023848, Not Applicable (HDJM51435);    M1112198-023848, Not Applicable (HDJM51394);    M1112198-177204, Not Applicable (HDJT51034);    M1112198-020611, Not Applicable (HDJM50350);    M1112198-018125, Not Applicable (HDJL50461);    M1112198-022337, Not Applicable (HDJM50375);    M1112198-022337, Not Applicable (HDJM50368);    M1112198-022337, Not Applicable (HDJM50373);    M1112198-022337, Not Applicable (HDJM50365);    M1112198-022337, Not Applicable (HDJM50383);    M1112198-022337, Not Applicable (HDJM50380);    M1112198-022339, Not Applicable (HDJM50351);    M1112198-022337, Not Applicable (HDJM50374);    M1112198-022337, Not Applicable (HDJM50376);    M1112198-022337, Not Applicable (HDJM50362);    M1112198-022337, Not Applicable (HDJM50364);    M1112198-022337, Not Applicable (HDJM50378);    M1112198-022337, Not Applicable (HDJM50361);    M1112198-022337, Not Applicable (HDJM50363);    M1112198-022337, Not Applicable (HDJM50370);    M1112198-022337, Not Applicable (HDJM50369);    M1112198-022337, Not Applicable (HDJM50366);    M1112198-022337, Not Applicable (HDJM50384);    M1112198-022337, Not Applicable (HDJM50372);    M1112198-022339, Not Applicable (HDJM50352);    M1112198-022337, Not Applicable (HDJM50371);    M1112198-022337, Not Applicable (HDJM50377);    M1112198-022337, Not Applicable (HDJM50379);    M1112198-022337, Not Applicable (HDJM50382);    M1112198-022337, Not Applicable (HDJM50381);    M1112198-022337, Not Applicable (HDJM50367);    M1112198-022592, Not Applicable (HDJM50838);    M1112198-261732, Not Applicable (HDJT61421);    M1112198-024139, Not Applicable (HDJM50803);    M1112198-024139, Not Applicable (HDJM50788);    M1112198-024139, Not Applicable (HDJM50801);    M1112198-022697, Not Applicable (HDJM51519);    M1112198-024139, Not Applicable (HDJM50781);    M1112198-023076, Not Applicable (HDJM50876);    M1112198-024139, Not Applicable (HDJQ50719);    M1112198-164256, Not Applicable (HDJT60438);    M1112198-257872, Not Applicable (HDJT50491);    M1112198-025821, Not Applicable (HDJM51120);    M1112198-164256, Not Applicable (HDJT61604);",
      "more_code_info": "M1112198-024139, Not Applicable (HDJN50240);    M1112198-024139, Not Applicable (HDJN50241);    M1112198-184659, Not Applicable (HDJT51189);    M1112198-022697, Not Applicable (HDJP50226);    M1112198-022697, Not Applicable (HDJQ50005);    M1112198-022697, Not Applicable (HDJQ50010);    M1112198-163959, Not Applicable (HDJT51342);    M1112198-163959, Not Applicable (HDJT51345);    M1112198-163959, Not Applicable (HDJT51243);    M1112198-164657, Not Applicable (HDJT50818);    M1112198-164657, Not Applicable (HDJT50752);    M1112198-020896, Not Applicable (HDJL50473);    M1112198-020896, Not Applicable (HDJL50474);    M1112198-285246, Not Applicable (HDJT51577);    M1112198-285246, Not Applicable (HDJT51673);    M1112198-284704, Not Applicable (HDJT51362);    M1112198-284704, Not Applicable (HDJT51361);    M1112198-024045, Not Applicable (HDJM50571);    M1112198-187561, Not Applicable (HDJS52743);    M1112198-018102, Not Applicable (HDJM50143);    M1112198-286082, Not Applicable (HDJT51386);    M1112198-286082, Not Applicable (HDJT51388);    M1112198-286082, Not Applicable (HDJT51389);    M1112198-286082, Not Applicable (HDJT51390);    M1112198-286082, Not Applicable (HDJT51387);    M1112198-027124, Not Applicable (HDJN50590);    M1112198-027124, Not Applicable (HDJM51574);    M1112198-027124, Not Applicable (HDJM51573);    M1112198-027124, Not Applicable (HDJM51575);    M1112198-027124, Not Applicable (HDJM51572);    M1112198-022836, Not Applicable (HDJM50396);    M1112198-022836, Not Applicable (HDJM50395);    M1112198-114875, Not Applicable (HDJR56088);    M1112198-018125, Not Applicable (HDJM50050);    M1112198-024047, Not Applicable (HDJM50569);    M1112198-022592, Not Applicable (HDJQ50146);    M1112198-150460, Not Applicable (HDJT50832);    M1112198-529431, 010084068210389321PBWX7143711180808 (PBWX71437);    M1112198-184659, Not Applicable (HDJT51188);    M1112198-184659, Not Applicable (HDJT51190);    M1112198-184659, Not Applicable (HDJT51836);    M1112198-184659, Not Applicable (HDJT51871);    M1112198-184659, Not Applicable (HDJT51187);    M1112198-184659, Not Applicable (HDJT51608);    M1112198-025251, Not Applicable (HDJM50962);    M1112198-022697, Not Applicable (HDJM50952);    M1112198-025251, Not Applicable (HDJM50960);    M1112198-025251, Not Applicable (HDJM50961);    M1112198-025251, Not Applicable (HDJM50958);    M1112198-022697, Not Applicable (HDJM50951);    M1112198-025251, Not Applicable (HDJM50959);    M1112198-022697, Not Applicable (HDJM50954);    M1112198-026731, Not Applicable (HDJM51442);    M1112198-026731, Not Applicable (HDJM51441);    M1112198-025031, Not Applicable (HDJM50910);    M1112198-026728, Not Applicable (HDJM51446);    M1112198-025031, Not Applicable (HDJM50912);    M1112198-026728, Not Applicable (HDJM51443);    M1112198-025031, Not Applicable (HDJM50907);    M1112198-164256, Not Applicable (HDJT51561);    M1112198-023608, Not Applicable (HDJN52307);    M1112198-022592, Not Applicable (HDJM50845);    M1112198-025031, Not Applicable (HDJM50908);    M1112198-025031, Not Applicable (HDJM50911);    M1112198-025031, Not Applicable (HDJM50913);    M1112198-025031, Not Applicable (HDJM50914);    M1112198-025031, Not Applicable (HDJM50909);    M1112198-026728, Not Applicable (HDJM51444);    M1112198-024139, Not Applicable (HDJM51613);    M1112198-026728, Not Applicable (HDJM51445);    M1112198-526395, 010084068210389321PBWW6201511170803 (PBWW62015);    M1112198-018125, Not Applicable (HDJM50052);    M1112198-030895, Not Applicable (HDJM52322);    M1112198-163959, Not Applicable (HDJT51341);    M1112198-526438, 010084068210389321PBWX7091711180623 (PBWX70917);    M1112198-023402, Not Applicable (HDJM50446);    M1112198-024139, Not Applicable (HDJN50903);    M1112198-024139, Not Applicable (HDJM51609);    M1112198-024139, Not Applicable (HDJM51611);    M1112198-191821, Not Applicable (HDJT51928);    M1112198-191821, Not Applicable (HDJT51929);    M1112198-175579, Not Applicable (HDJT51696);    M1112198-024585, Not Applicable (HDJM50891);    M1112198-024585, Not Applicable (HDJM50889);    M1112198-024585, Not Applicable (HDJM50890);    M1112198-023608, Not Applicable (HDJQ50434);    M1112198-023608, Not Applicable (HDJQ50435);    M1112198-023608, Not Applicable (HDJQ50437);    M1112198-023608, Not Applicable (HDJQ50438);    M1112198-527790, 010084068210389321PBWX7011911180216 (PBWX70119);    M1112198-163959, Not Applicable (HDJS55191);    M1112198-527790, 010084068210389321PBWX7123611180723 (PBWX71236);    M1112198-023848, Not Applicable (HDJR50502);    M1112198-526395, 010084068210389321PBWW6124211170516 (PBWW61242);    M1112198-526395, 010084068210389321PBWW6069411170321 (PBWW60694);    M1112198-526438, 010084068210389321PBWX7081811180611 (PBWX70818);    M1112198-526438, 010084068210389321PBWX7082411180612 (PBWX70824);    M1112198-184659, Not Applicable (HDJT51371);    M1112198-150400, Not Applicable (HDJT50694);    M1112198-022632, Not Applicable (HDJM51576);    M1112198-023848, Not Applicable (HDJM50887);    M1112198-023848, Not Applicable (HDJM50886);    M1112198-023848, Not Applicable (HDJM50884);    M1112198-023848, Not Applicable (HDJM50885);    M1112198-023528, Not Applicable (HDJN50087);    M1112198-023528, Not Applicable (HDJN50086);    M1112198-022697, Not Applicable (HDJN50653);    M1112198-527596, 010084068210389321PBWW6126811170517 (PBWW61268);    M1112198-527596, 010084068210389321PBWW6126911170517 (PBWW61269);    M1112198-166779, Not Applicable (HDJT51293);    M1112198-166779, Not Applicable (HDJT51294);    M1112198-022592, Not Applicable (HDJM50472);    M1112198-024139, Not Applicable (HDJQ50665);    M1112198-024139, Not Applicable (HDJQ50661);    M1112198-024139, Not Applicable (HDJQ50707);    M1112198-025718, Not Applicable (HDJN50076);    M1112198-526395, 010084068210389321PBWW6124811170516 (PBWW61248);    M1112198-526773, 010084068210389321PBWX7087111180615 (PBWX70871);    M1112198-526773, 010084068210389321PBWX7087411180615 (PBWX70874);    M1112198-526773, 010084068210389321PBWW6129811170518 (PBWW61298);    M1112198-184659, Not Applicable (HDJT51301);    M1112198-164657, Not Applicable (HDJT61471);    M1112198-164657, Not Applicable (HDJT61473);    M1112198-164657, Not Applicable (HDJT50264);    M1112198-018125, Not Applicable (HDJM50053);    M1112198-018125, Not Applicable (HDJM50105);    M1112198-018125, Not Applicable (HDJM50054);    M1112198-018125, Not Applicable (HDJM50103);    M1112198-018125, Not Applicable (HDJM50124);    M1112198-018125, Not Applicable (HDJM50098);    M1112198-018125, Not Applicable (HDJM50099);    M1112198-018125, Not Applicable (HDJM50101);    M1112198-023620, Not Applicable (HDJM50521);    M1112198-023620, Not Applicable (HDJM50520);    M1112198-023620, Not Applicable (HDJM50522);    M1112198-171606, Not Applicable (HDJT61437);    M1112198-528361, 010084068210389321PBWX7072511180531 (PBWX70725);    M1112198-018697, Not Applicable (HDJL50310);    M1112198-023391, Not Applicable (HDJM50434);    M1112198-023848, Not Applicable (HDJQ51423);    M1112198-023848, Not Applicable (HDJM51606);    M1112198-023848, Not Applicable (HDJM51605);    M1112198-189220, Not Applicable (HDJT51937);    M1112198-527554, 010084068210389321PBWX7036711180405 (PBWX70367);    M1112198-526773, 010084068210389321PBWW6129511170518 (PBWW61295);    M1112198-164657, Not Applicable (HDJT51042);    M1112198-164657, Not Applicable (HDJT51037);    M1112198-164657, Not Applicable (HDJT51038);    M1112198-022697, Not Applicable (HDJN50510);    M1112198-018102, Not Applicable (HDJM50184);    M1112198-018102, Not Applicable (HDJM50185);    M1112198-023431, Not Applicable (HDJM50451);    M1112198-023848, Not Applicable (HDJR50415);    M1112198-027551, Not Applicable (HDJM51624);    M1112198-027551, Not Applicable (HDJM51625);    M1112198-061236, Not Applicable (HDJP52729);    M1112198-527554, 010084068210389321PBWX7068511180524 (PBWX70685);    M1112198-023488, Not Applicable (HDJP52752);    M1112198-163959, Not Applicable (HDJT51788);    M1112198-166065, Not Applicable (HDJT51060);    M1112198-018125, Not Applicable (HDJM50123);    M1112198-018505, Not Applicable (HDJM50526);    M1112198-018125, Not Applicable (HDJM50112);    M1112198-023488, Not Applicable (HDJQ50741);    M1112198-020996, Not Applicable (HDJM50277);    M1112198-032131, Not Applicable (HDJP52181);    M1112198-022822, Not Applicable (HDJM51064);    M1112198-024586, Not Applicable (HDJM50823);    M1112198-032131, Not Applicable (HDJP50027);    M1112198-032131, Not Applicable (HDJP50020);    M1112198-032131, Not Applicable (HDJP50018);    M1112198-032131, Not Applicable (HDJP50024);    M1112198-023488, Not Applicable (HDJP52886);    M1112198-024139, Not Applicable (HDJR50405);    M1112198-024796, Not Applicable (HDJM50821);    M1112198-024796, Not Applicable (HDJM50822);    M1112198-024796, Not Applicable (HDJM50820);    M1112198-526395, 010084068210389321PBWX7077211180606 (PBWX70772);    M1112198-042314, Not Applicable (HDJN52360);    M1112198-023848, Not Applicable (HDJM51214);    M1112198-023848, Not Applicable (HDJM51215);    M1112198-228205, Not Applicable (HDJT50565);    M1112198-034977, Not Applicable (HDJR56664);    M1112198-030407, Not Applicable (HDJN50162);    M1112198-164075, Not Applicable (HDJT50917);    M1112198-164075, Not Applicable (HDJT50918);    M1112198-164075, Not Applicable (HDJT50914);    M1112198-501885, 010084068210389321PBWV6139611161116 (PBWV61396);    M1112198-034977, Not Applicable (HDJR56132);    M1112198-034977, Not Applicable (HDJQ50782);    M1112198-034977, Not Applicable (HDJQ50374);    M1112198-164075, Not Applicable (HDJT50547);    M1112198-034977, Not Applicable (HDJQ50838);    M1112198-164075, Not Applicable (HDJT50912);    M1112198-026442, Not Applicable (HDJM51348);    M1112198-164075, Not Applicable (HDJT50916);    M1112198-034977, Not Applicable (HDJN50606);    M1112198-031731, Not Applicable (HDJN50000);    M1112198-022438, Not Applicable (HDJM50239);    M1112198-022438, Not Applicable (HDJM50240);    M1112198-034977, Not Applicable (HDJN52101);    M1112198-034977, Not Applicable (HDJN52102);    M1112198-248289, Not Applicable (HDJT51743);    M1112198-248289, Not Applicable (HDJT51744);    M1112198-248289, Not Applicable (HDJT51730);    M1112198-248289, Not Applicable (HDJT51739);    M1112198-248289, Not Applicable (HDJT51741);    M1112198-248289, Not Applicable (HDJT51742);    M1112198-248289, Not Applicable (HDJT51745);    M1112198-025419, Not Applicable (HDJM51025);    M1112198-099937, Not Applicable (HDJR50322);    M1112198-300776, Not Applicable (HDJT51579);    M1112198-035135, Not Applicable (HDJN50611);    M1112198-024460, Not Applicable (HDJM50725);    M1112198-024460, Not Applicable (HDJM50724);    M1112198-020774, Not Applicable (HDJL50441);    M1112198-513873, 010084068210389321PBWW6126311170516 (PBWW61263);    M1112198-507358, 010084068210389321PBWW6011111170117 (PBWW60111);    M1112198-279475, Not Applicable (HDJT50676);    M1112198-166161, Not Applicable (HDJT51712);    M1112198-305498, Not Applicable (HDJT51765);    M1112198-305498, Not Applicable (HDJT51770);    M1112198-305498, Not Applicable (HDJT51771);    M1112198-172880, Not Applicable (HDJT51778);    M1112198-305498, Not Applicable (HDJT51764);    M1112198-305498, Not Applicable (HDJT51766);    M1112198-305498, Not Applicable (HDJT51768);    M1112198-305498, Not Applicable (HDJT51769);    M1112198-172880, Not Applicable (HDJT51779);    M1112198-305498, Not Applicable (HDJT51767);    M1112198-305498, Not Applicable (HDJT51772);    M1112198-284393, Not Applicable (HDJT51363);    M1112198-187180, Not Applicable (HDJT50826);    M1112198-284723, Not Applicable (HDJT51014);    M1112198-167099, Not Applicable (HDJS50456);    M1112198-040593, Not Applicable (HDJP50190);    M1112198-032718, Not Applicable (HDJN50201);    M1112198-032717, Not Applicable (HDJN50236);    M1112198-027608, Not Applicable (HDJM51644);    M1112198-026680, Not Applicable (HDJM51433);    M1112198-026680, Not Applicable (HDJM51434);    M1112198-026680, Not Applicable (HDJM51430);    M1112198-027608, Not Applicable (HDJM51645);    M1112198-027608, Not Applicable (HDJM51643);    M1112198-022158, Not Applicable (HDJM50153);    M1112198-022158, Not Applicable (HDJM50157);    M1112198-022158, Not Applicable (HDJM50158);    M1112198-026239, Not Applicable (HDJM51323);    M1112198-026194, Not Applicable (HDJM51328);    M1112198-026194, Not Applicable (HDJM51327);    M1112198-026680, Not Applicable (HDJM51428);    M1112198-026680, Not Applicable (HDJM51431);    M1112198-027608, Not Applicable (HDJM51642);    M1112198-027608, Not Applicable (HDJM51641);    M1112198-027608, Not Applicable (HDJM51649);    M1112198-027608, Not Applicable (HDJM51647);    M1112198-027608, Not Applicable (HDJM51648);    M1112198-027608, Not Applicable (HDJM51646);    M1112198-022158, Not Applicable (HDJM50148);    M1112198-022158, Not Applicable (HDJM50149);    M1112198-022158, Not Applicable (HDJM50150);    M1112198-022158, Not Applicable (HDJM50151);    M1112198-022158, Not Applicable (HDJM50152);    M1112198-022158, Not Applicable (HDJM50166);    M1112198-022158, Not Applicable (HDJM50155);    M1112198-022158, Not Applicable (HDJM50165);    M1112198-022158, Not Applicable (HDJM50159);    M1112198-022158, Not Applicable (HDJM50154);    M1112198-022158, Not Applicable (HDJM50156);    M1112198-022158, Not Applicable (HDJM50163);    M1112198-022158, Not Applicable (HDJM50160);    M1112198-022158, Not Applicable (HDJM50167);    M1112198-022158, Not Applicable (HDJM50172);    M1112198-022158, Not Applicable (HDJM50173);    M1112198-022158, Not Applicable (HDJM50171);    M1112198-022158, Not Applicable (HDJM50161);    M1112198-022158, Not Applicable (HDJM50174);    M1112198-022158, Not Applicable (HDJM50168);    M1112198-022158, Not Applicable (HDJM50177);    M1112198-022158, Not Applicable (HDJM50164);    M1112198-022158, Not Applicable (HDJM50176);    M1112198-022158, Not Applicable (HDJM50175);    M1112198-023766, Not Applicable (HDJM50523);    M1112198-023766, Not Applicable (HDJM50524);    M1112198-026196, Not Applicable (HDJM51324);    M1112198-026194, Not Applicable (HDJM51326);    M1112198-026194, Not Applicable (HDJM51329);    M1112198-026681, Not Applicable (HDJM51426);    M1112198-026680, Not Applicable (HDJM51429);    M1112198-026196, Not Applicable (HDJM51688);    M1112198-019901, Not Applicable (HDJL50325);    M1112198-022158, Not Applicable (HDJM50162);    M1112198-022158, Not Applicable (HDJM50170);    M1112198-026681, Not Applicable (HDJM51425);    M1112198-026680, Not Applicable (HDJM51432);    M1112198-023911, Not Applicable (HDJM50552);    M1112198-026680, Not Applicable (HDJM51427);    M1112198-023911, Not Applicable (HDJM50551);    M1112198-072444, Not Applicable (HDJQ50784);    M1112198-082904, Not Applicable (HDJQ51699);    M1112198-033651, Not Applicable (HDJQ50235);    M1112198-026194, Not Applicable (HDJQ50856);    M1112198-026194, Not Applicable (HDJQ50857);    M1112198-026194, Not Applicable (HDJQ50858);    M1112198-093690, Not Applicable (HDJR50350);    M1112198-062757, Not Applicable (HDJR50441);    M1112198-026785, Not Applicable (HDJM51463);    M1112198-534499, 010084068210389321PBWW61351170523 (PBWW61351);    M1112198-534499, 010084068210389321PBWW61349170523 (PBWW61349);    M1112198-025132, Not Applicable (HDJM50830);    M1112198-028111, Not Applicable (HDJM51743);    M1112198-023077, Not Applicable (HDJM51499);    M1112198-033375, Not Applicable (HDJN50318);    M1112198-023077, Not Applicable (HDJM51498);    M1112198-023077, Not Applicable (HDJM51496);    M1112198-023077, Not Applicable (HDJM51500);    M1112198-023077, Not Applicable (HDJM51497);    M1112198-023077, Not Applicable (HDJM51546);    M1112198-023077, Not Applicable (HDJM50319);    M1112198-023077, Not Applicable (HDJM50633);    M1112198-033375, Not Applicable (HDJN51580);    M1112198-023077, Not Applicable (HDJM51299);    M1112198-279379, Not Applicable (HDJT51015);    M1112198-024265, Not Applicable (HDJM50671);    M1112198-304512, Not Applicable (HDJT51746);    M1112198-304512, Not Applicable (HDJT51747);    M1112198-024265, Not Applicable (HDJM50670);    M1112198-024265, Not Applicable (HDJN50577);    M1112198-024265, Not Applicable (HDJM50672);    M1112198-024265, Not Applicable (HDJM50667);    M1112198-024265, Not Applicable (HDJM50669);    M1112198-024265, Not Applicable (HDJN50578);    M1112198-024265, Not Applicable (HDJM50668);    M1112198-156841, Not Applicable (HDJT61351);    M1112198-171262, Not Applicable (HDJT50961);    M1112198-171262, Not Applicable (HDJT50962);    M1112198-171995, Not Applicable (HDJT51832);    M1112198-021879, Not Applicable (HDJM50144);    M1112198-021879, Not Applicable (HDJM50238);    M1112198-027431, Not Applicable (HDJM51603);    M1112198-025605, Not Applicable (HDJM51053);    M1112198-025605, Not Applicable (HDJM51052);    M1112198-072984, Not Applicable (HDJQ50840);    M1112198-025605, Not Applicable (HDJM51051);    M1112198-024385, Not Applicable (HDJM50684);    M1112198-027521, Not Applicable (HDJM51562);    M1112198-070446, Not Applicable (HDJQ50588);    M1112198-028738, Not Applicable (HDJM51877);    M1112198-027431, Not Applicable (HDJM51550);    M1112198-027431, Not Applicable (HDJM51602);    M1112198-527554, 010084068210389321PBWW61219170510 (PBWW61219);    M1112198-055158, Not Applicable (HDJP51971);    M1112198-048339, Not Applicable (HDJQ51867);    M1112198-022469, Not Applicable (HDJM50488);    M1112198-062439, Not Applicable (HDJQ51856);    M1112198-263514, Not Applicable (HDJT50628);    M1112198-520122, 010084068210389321PBWW60086170113 (PBWW60086);    M1112198-287085, Not Applicable (HDJT51195);    M1112198-287085, Not Applicable (HDJT51209);    M1112198-287085, Not Applicable (HDJT51210);    M1112198-287085, Not Applicable (HDJT51193);    M1112198-287085, Not Applicable (HDJT51196);    M1112198-287085, Not Applicable (HDJT51198);    M1112198-287085, Not Applicable (HDJT51199);    M1112198-287085, Not Applicable (HDJT51206);    M1112198-166381, Not Applicable (HDJT51408);    M1112198-565983, 010084068210389321PBWX7142711180807 (PBWX71427);    M1112198-565983, 010084068210389321PBWX7109711180709 (PBWX71097);    M1112198-565983, 010084068210389321PBWX7112511180711 (PBWX71125);    M1112198-565983, 010084068210389321PBWX7112711180711 (PBWX71127);    M1112198-565983, 010084068210389321PBWX7113311180711 (PBWX71133);    M1112198-565983, 010084068210389321PBWX7117011180717 (PBWX71170);    M1112198-565983, 010084068210389321PBWX7118411180718 (PBWX71184);    M1112198-565983, 010084068210389321PBWX7108611180709 (PBWX71086);    M1112198-565983, 010084068210389321PBWX7101011180703 (PBWX71010);    M1112198-565983, 010084068210389321PBWX7107811180706 (PBWX71078);    M1112198-565983, 010084068210389321PBWX7116511180717 (PBWX71165);    M1112198-565983, 010084068210389321PBWX7118311180718 (PBWX71183);    M1112198-565983, 010084068210389321PBWX7118611180718 (PBWX71186);    M1112198-334395, Not Applicable (HDJT60554);    M1112198-334395, Not Applicable (HDJT60555);    M1112198-025032, Not Applicable (HDJM51075);    M1112198-245509, Not Applicable (HDJT51120);    M1112198-245509, Not Applicable (HDJT51124);    M1112198-232001, Not Applicable (HDJS55072);    M1112198-233603, Not Applicable (HDJS53484);    M1112198-056034, Not Applicable (HDJP52034);    M1112198-056034, Not Applicable (HDJP52038);    M1112198-056034, Not Applicable (HDJP52031);    M1112198-035237, Not Applicable (HDJN50633);    M1112198-035237, Not Applicable (HDJN50635);    M1112198-035237, Not Applicable (HDJN50637);    M1112198-057254, Not Applicable (HDJQ52429);    M1112198-056034, Not Applicable (HDJQ51028);    M1112198-057254, Not Applicable (HDJQ51719);    M1112198-171995, Not Applicable (HDJT51819);    M1112198-171995, Not Applicable (HDJT51820);    M1112198-171995, Not Applicable (HDJT51821);    M1112198-171995, Not Applicable (HDJT51826);    M1112198-171995, Not Applicable (HDJT51830);    M1112198-171995, Not Applicable (HDJT51833);    M1112198-171995, Not Applicable (HDJT51831);    M1112198-171995, Not Applicable (HDJT51822);    M1112198-171995, Not Applicable (HDJT51825);    M1112198-171995, Not Applicable (HDJT51824);    M1112198-171995, Not Applicable (HDJT51827);    M1112198-171995, Not Applicable (HDJT51828);    M1112198-171995, Not Applicable (HDJT51829);    M1112198-192823, Not Applicable (HDJS51940);    M1112198-043587, Not Applicable (HDJN52168);    M1112198-043587, Not Applicable (HDJN52155);    M1112198-025535, Not Applicable (HDJM50996);    M1112198-025535, Not Applicable (HDJM50998);    M1112198-025535, Not Applicable (HDJM50994);    M1112198-025535, Not Applicable (HDJM50995);    M1112198-025535, Not Applicable (HDJM50990);    M1112198-025535, Not Applicable (HDJM50997);    M1112198-025535, Not Applicable (HDJM51003);    M1112198-025535, Not Applicable (HDJM50999);    M1112198-025535, Not Applicable (HDJM50992);    M1112198-025535, Not Applicable (HDJM51000);    M1112198-025535, Not Applicable (HDJM50993);    M1112198-025535, Not Applicable (HDJM51004);    M1112198-025535, Not Applicable (HDJM51005);    M1112198-025535, Not Applicable (HDJM51001);    M1112198-025535, Not Applicable (HDJM51002);    M1112198-025535, Not Applicable (HDJM50991);    M1112198-043587, Not Applicable (HDJQ50116);    M1112198-041192, Not Applicable (HDJQ50461);    M1112198-024049, Not Applicable (HDJM50568);    M1112198-024049, Not Applicable (HDJM50564);    M1112198-024049, Not Applicable (HDJM50567);    M1112198-024049, Not Applicable (HDJM51026);    M1112198-024049, Not Applicable (HDJM50565);    M1112198-024049, Not Applicable (HDJM50566);    M1112198-026575, Not Applicable (HDJM51531);    M1112198-025338, Not Applicable (HDJM50969);    M1112198-175561, Not Applicable (HDJS50794);    M1112198-027715, Not Applicable (HDJM51726);    M1112198-027541, Not Applicable (HDJM51545);    M1112198-027543, Not Applicable (HDJM51542);    M1112198-027543, Not Applicable (HDJM51543);    M1112198-027541, Not Applicable (HDJM51544);    M1112198-027543, Not Applicable (HDJM51541);    M1112198-027543, Not Applicable (HDJM51604);    M1112198-171262, Not Applicable (HDJT50579);    M1112198-036543, Not Applicable (HDJQ50523);    M1112198-036543, Not Applicable (HDJQ50522);    M1112198-021394, Not Applicable (HDJM50014);    M1112198-505101, 010084068210389321PBWW6168111170629 (PBWW61681);    M1112198-020821, Not Applicable (HDJL50438);    M1112198-020821, Not Applicable (HDJL50440);    M1112198-020821, Not Applicable (HDJL50439);    M1112198-020821, Not Applicable (HDJL50437);    M1112198-020821, Not Applicable (HDJL50436);    M1112198-036543, Not Applicable (HDJQ50544);    M1112198-516623, 010084068210389321PBWW6034311170207 (PBWW60343);    M1112198-034073, Not Applicable (HDJQ51209);    M1112198-171262, Not Applicable (HDJT50960);    M1112198-021394, Not Applicable (HDJM50013);    M1112198-171262, Not Applicable (HDJT50571);    M1112198-171262, Not Applicable (HDJT50581);"
    }
  ]
}