      "status": "Terminated",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63518",
      "recalling_firm": "Hospira Inc.",
      "address_1": "275 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of Australia, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Denmark, France, Greece, Hong Kong, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Libya, Malaysia, Mexico, Netherlands, New Zealand, Oman, Pakistan, Philippines, Poland, Portugal, Saudi Arabia, Singapore, Lavenia, South Africa, Spain, Switzerland, Taiwan, Turkey, United Kingdom, United Arab Emirates and Vietnam.",
      "recall_number": "Z-0482-2013",
      "product_description": "Plum A+ Hyperbaric Single Channel Infusion Pumps with Hospira MedNet Software and a Connectivity Engine peripheral module that provides wired Ethernet and wireless local networking, allowing the MedNet software to download drug libraries to the infuser and enable the auto-programming feature; Hospira, Inc., Lake Forest, IL 60045; pump list number 11005, module list number 11006      Product Usage: The Plum A+ Hyperbaric Infusion System is exclusively designed for hyperbaric therapy in monospace and multiplace chambers.",
      "product_quantity": "N/A",
      "reason_for_recall": "The volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed.  The manual indicates a clockwise rotation of the knob decreases the volume; however, on some of the devices the alarm volume increases when the knob is turned clockwise.",
      "recall_initiation_date": "20121026",
      "center_classification_date": "20121207",
      "termination_date": "20161114",
      "report_date": "20121219",
      "code_info": "pump list number 11005 and module list number 11006; all serial numbers"
    }
  ]
}