{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81224",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US Nationwide, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Honduras, India, Indonesia, Ireland, Italy, Japan, Kazakhstan, Korea, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.",
      "recall_number": "Z-0481-2019",
      "product_description": "Giraffe Warmer. Device sold under the following product names: Giraffe Warmer and Giraffe Bedded Warmer.   infant radiant warmer.",
      "product_quantity": "1,724 devices total",
      "reason_for_recall": "Potential electric leakage affecting the user.",
      "recall_initiation_date": "20180925",
      "center_classification_date": "20181119",
      "termination_date": "20200713",
      "report_date": "20181128",
      "code_info": "Model Number, UDI (Serial Number):  M1118179-018104, Not Applicable (HDJM50056);    M1118179-018118, Not Applicable (HDJM50178);    M1118179-024117, Not Applicable (HDJM50583);    M1118179-018118, Not Applicable (HDJM50094);    M1118179-022861, Not Applicable (HDJR54165);    M1118179-020841, Not Applicable (HDJM50142);    M1118179-024117, Not Applicable (HDJQ50967);    M1118179-024117, Not Applicable (HDJQ50971);    M1118179-024117, Not Applicable (HDJQ50973);    M1118179-024117, Not Applicable (HDJQ50972);    M1118179-024117, Not Applicable (HDJQ50974);    M1118179-022643, Not Applicable (HDJM52091);    M1118179-018104, Not Applicable (HDJM50060);    M1118179-150421, Not Applicable (HDJT51438);    M1118179-150421, Not Applicable (HDJT51439);    M1118179-022643, Not Applicable (HDJM50696);    M1118179-150421, Not Applicable (HDJT51652);    M1118179-022643, Not Applicable (HDJQ50385);    M1118179-161255, Not Applicable (HDJT60578);    M1118179-161255, Not Applicable (HDJT60576);    M1118179-161255, Not Applicable (HDJT60572);    M1118179-161255, Not Applicable (HDJT60584);    M1118179-161255, Not Applicable (HDJT60568);    M1118179-026229, Not Applicable (HDJM51309);    M1118179-022643, Not Applicable (HDJM50420);    M1118179-022643, Not Applicable (HDJM50419);    M1118179-022643, Not Applicable (HDJM50693);    M1118179-022643, Not Applicable (HDJM51718);    M1118179-026229, Not Applicable (HDJM51567);    M1118179-226273, Not Applicable (HDJT51359);    M1118179-022393, Not Applicable (HDJM50289);    M1118179-022393, Not Applicable (HDJM50288);    M1118179-022861, Not Applicable (HDJM50957);    M1118179-022861, Not Applicable (HDJM50956);    M1118179-160267, Not Applicable (HDJT61386);    M1118179-526626, 010084068210392321GBWX7019911180619 (GBWX70199);    M1118179-023046, Not Applicable (HDJP52075);    M1118179-023046, Not Applicable (HDJM50955);    M1118179-021393, Not Applicable (HDJM50046);    M1118179-021393, Not Applicable (HDJM50527);    M1118179-024117, Not Applicable (HDJM50584);    M1118179-559396, 010084068210392321GBWX7007211180312 (GBWX70072);    M1118179-023046, Not Applicable (HDJR50514);    M1118179-023046, Not Applicable (HDJM50387);    M1118179-023046, Not Applicable (HDJM50386);    M1118179-150421, Not Applicable (HDJT50842);    M1118179-150421, Not Applicable (HDJT61375);    M1118179-025977, Not Applicable (HDJQ50585);    M1118179-025977, Not Applicable (HDJQ50581);    M1118179-527794, 010084068210392321GBWX7017111180523 (GBWX70171);    M1118179-527794, 010084068210392321GBWX7017011180523 (GBWX70170);    M1118179-022861, Not Applicable (HDJM50343);    M1118179-022643, Not Applicable (HDJM50695);    M1118179-161255, Not Applicable (HDJT51954);    M1118179-174787, Not Applicable (HDJT51358);    M1118179-174787, Not Applicable (HDJT51357);    M1118179-019026, Not Applicable (HDJL50356);    M1118179-022643, Not Applicable (HDJP52630);    M1118179-024044, Not Applicable (HDJM50573);    M1118179-024044, Not Applicable (HDJM50574);    M1118179-024044, Not Applicable (HDJM50572);    M1118179-526477, 010084068210392321GBWW6032811170511 (GBWW60328);    M1118179-019593, Not Applicable (HDJL50433);    M1118179-019593, Not Applicable (HDJL50434);    M1118179-019593, Not Applicable (HDJL50435);    M1118179-019593, Not Applicable (HDJL50432);    M1118179-018104, Not Applicable (HDJM50059);    M1118179-018114, Not Applicable (HDJM50097);    M1118179-026007, Not Applicable (HDJM51296);    M1118179-529707, 010084068210392321GBWX7023811180627 (GBWX70238);    M1118179-529707, 010084068210392321GBWX7023911180627 (GBWX70239);    M1118179-022643, Not Applicable (HDJM50694);    M1118179-022643, Not Applicable (HDJN51744);    M1118179-022861, Not Applicable (HDJM51320);    M1118179-022861, Not Applicable (HDJM51321);    M1118179-526477, 010084068210392321GBWX7024211180627 (GBWX70242);    M1118179-526477, 010084068210392321GBWW6032711170511 (GBWW60327);    M1118179-024117, Not Applicable (HDJN52275);    M1118179-018118, Not Applicable (HDJM50188);    M1118179-150421, Not Applicable (HDJT50988);    M1118179-254785, Not Applicable (HDJT51864);    M1118179-023046, Not Applicable (HDJM50536);    M1118179-023046, Not Applicable (HDJM50537);    M1118179-150421, Not Applicable (HDJT50843);    M1118179-018118, Not Applicable (HDJM50045);    M1118179-018118, Not Applicable (HDJM50186);    M1118179-018114, Not Applicable (HDJM50005);    M1118179-018114, Not Applicable (HDJM50006);    M1118179-018114, Not Applicable (HDJM50007);    M1118179-018104, Not Applicable (HDJM50061);    M1118179-018104, Not Applicable (HDJM50055);    M1118179-022643, Not Applicable (HDJM50698);    M1118179-022643, Not Applicable (HDJM50697);    M1118179-027451, Not Applicable (HDJM51565);    M1118179-027451, Not Applicable (HDJM51564);    M1118179-027451, Not Applicable (HDJM51566);    M1118179-242231, Not Applicable (HDJT50732);    M1118179-150421, Not Applicable (HDJT50839);    M1118179-150421, Not Applicable (HDJT50702);    M1118179-150421, Not Applicable (HDJT50722);    M1118179-023046, Not Applicable (HDJN50207);    M1118179-018118, Not Applicable (HDJM50179);    M1118179-018104, Not Applicable (HDJM50058);    M1118179-018118, Not Applicable (HDJM50062);    M1118179-018118, Not Applicable (HDJM50093);    M1118179-023046, Not Applicable (HDJM50654);    M1118179-025977, Not Applicable (HDJM51225);    M1118179-022125, Not Applicable (HDJM50470);    M1118179-018118, Not Applicable (HDJM50187);    M1118179-023046, Not Applicable (HDJM50388);    M1118179-023046, Not Applicable (HDJM50538);    M1118179-023046, Not Applicable (HDJP51904);    M1118179-150421, Not Applicable (HDJT50989);    M1118179-023046, Not Applicable (HDJM51553);    M1118179-023046, Not Applicable (HDJM51554);    M1118179-150421, Not Applicable (HDJT50838);    M1118179-018104, Not Applicable (HDJM50057);    M1118179-022256, Not Applicable (HDJM50180);    M1118179-018114, Not Applicable (HDJL50487);    M1118179-025977, Not Applicable (HDJM51217);    M1118179-025977, Not Applicable (HDJM51219);    M1118179-025977, Not Applicable (HDJM51224);    M1118179-025977, Not Applicable (HDJM51220);    M1118179-025977, Not Applicable (HDJM51216);    M1118179-025977, Not Applicable (HDJM51221);    M1118179-025977, Not Applicable (HDJM51218);    M1118179-025977, Not Applicable (HDJM51223);    M1118179-025977, Not Applicable (HDJM51222);    M1118179-044164, Not Applicable (HDJR50298);    M1118179-044164, Not Applicable (HDJR56108);    M1118179-026395, Not Applicable (HDJM51460);    M1118179-026395, Not Applicable (HDJM51461);    M1118179-022439, Not Applicable (HDJM50237);    M1118179-021533, Not Applicable (HDJM50147);    M1118179-027863, Not Applicable (HDJM51724);    M1118179-044164, Not Applicable (HDJR50297);    M1118179-044164, Not Applicable (HDJR52094);    M1118179-044164, Not Applicable (HDJR52095);    M1118179-044164, Not Applicable (HDJR52096);    M1118179-496293, 010084068210392321GBWW6030811170510 (GBWW60308);    M1118179-072924, Not Applicable (HDJQ50843);    M1118179-022807, Not Applicable (HDJM50309);    M1118179-235294, Not Applicable (HDJT51537);    M1118179-019903, Not Applicable (HDJL50330);    M1118179-019903, Not Applicable (HDJL50329);    M1118179-026183, Not Applicable (HDJM51322);    M1118179-067995, Not Applicable (HDJQ50334);    M1118179-067995, Not Applicable (HDJQ50333);    M1118179-276837, Not Applicable (HDJT50987);    M1118179-100772, Not Applicable (HDJR50394);    M1118179-026877, Not Applicable (HDJM51456);    M1118179-026788, Not Applicable (HDJM51462);    M1118179-278167, Not Applicable (HDJT51000);    M1118179-246044, Not Applicable (HDJT51230);    M1118179-024645, Not Applicable (HDJM50753);    M1118179-024645, Not Applicable (HDJM50754);    M1118179-024645, Not Applicable (HDJM50755);    M1118179-023078, Not Applicable (HDJM50305);    M1118179-023078, Not Applicable (HDJM50661);    M1118179-023078, Not Applicable (HDJM50660);    M1118179-024264, Not Applicable (HDJM50663);    M1118179-024264, Not Applicable (HDJM50662);    M1118179-023361, Not Applicable (HDJM50385);    M1118179-025604, Not Applicable (HDJM51050);    M1118179-044906, Not Applicable (HDJR50453);    M1118179-022468, Not Applicable (HDJM50503);    M1118179-186200, Not Applicable (HDJT50777);    M1118179-186200, Not Applicable (HDJT50781);    M1118179-160971, Not Applicable (HDJT50664);    M1118179-160971, Not Applicable (HDJT51008);    M1118179-022439, Not Applicable (HDJM50236);    M1118179-343873, Not Applicable (HDJT61094);    M1118179-024668, Not Applicable (HDJM50689);    M1118179-525532, 010084068210392321GBWW6032011170510 (GBWW60320);    M1118179-525532, 010084068210392321GBWW6033011170515 (GBWW60330);    M1118179-525532, 010084068210392321GBWW6033111170515 (GBWW60331);    M1118179-158967, Not Applicable (HDJT51411);    M1118179-024625, Not Applicable (HDJM50688);    M1118179-063038, Not Applicable (HDJQ50701);    M1118179-052375, Not Applicable (HDJQ50342);    M1118179-052375, Not Applicable (HDJQ51188);    M1118179-066171, Not Applicable (HDJQ50128);    M1118179-066171, Not Applicable (HDJQ50126);    M1118179-285625, Not Applicable (HDJT51152);    M1118179-030272, Not Applicable (HDJQ50874);    M1118179-030272, Not Applicable (HDJQ50875);    M1118179-025788, Not Applicable (HDJM51192);    M1118179-024130, Not Applicable (HDJM50752);    M1118179-024130, Not Applicable (HDJM51505);    M1118179-024130, Not Applicable (HDJM50585);    M1118179-028157, Not Applicable (HDJM51755);    M1118179-025086, Not Applicable (HDJM50888);    M1118179-022828, Not Applicable (HDJM50306);    M1118179-181417, Not Applicable (HDJT50497);    M1118179-052773, Not Applicable (HDJQ51128);    M1118179-157969, Not Applicable (HDJT61086);    M1118179-021221, Not Applicable (HDJM50000);    M1118179-300415, Not Applicable (HDJT51840);    M1118179-021221, Not Applicable (HDJM50001);    M1118179-021221, Not Applicable (HDJM50002);    M1118179-023156, Not Applicable (HDJM50655);    M1118179-021576, Not Applicable (HDJM50043);    M1118179-021576, Not Applicable (HDJM50044);    M1118179-030504, Not Applicable (HDJN50129);    M1118179-030504, Not Applicable (HDJN50126);    M1118179-023145, Not Applicable (HDJM50323);    M1118179-023156, Not Applicable (HDJM50336);    M1118179-533902, 010084068210392321GBWW60349170515 (GBWW60349);",
      "more_code_info": ""
    }
  ]
}