  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72742",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : IL, WI, MA, FL, AL, TN, ND, OH,  TX, CA, NJ and MI.",
      "recall_number": "Z-0481-2016",
      "product_description": "Artis systems with large display from Siemens Medical Solutions USA, Inc.  Interventional Fluoroscopic X-ray for angiographic procedures.",
      "product_quantity": "17",
      "reason_for_recall": "Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in factory screening.",
      "recall_initiation_date": "20151030",
      "center_classification_date": "20151224",
      "termination_date": "20170328",
      "report_date": "20151230",
      "code_info": "Model numbers 10094135, 10094137, 10094141, 10848280, 10848281, 10848282, 10848355. Serial Numbers 123026, 147909, 123031, 103030, 147912, 147910, 147921, 147919, 131005, 121096, 125004, 137551, 121088, 109104, 109105, 109107, and 121085."
    }
  ]
}