  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Suresnes",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81508",
      "recalling_firm": "Voluntis",
      "address_1": "22 Quai Gallieni",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).",
      "recall_number": "Z-0478-2019",
      "product_description": "Insulia Diabetes Management Companion",
      "product_quantity": "14",
      "reason_for_recall": "The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.",
      "recall_initiation_date": "20181008",
      "center_classification_date": "20181116",
      "termination_date": "20190312",
      "report_date": "20181128",
      "code_info": "Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40",
      "more_code_info": ""
    }
  ]
}