{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91200",
      "recalling_firm": "Leica Microsystems, Inc.",
      "address_1": "10 Parkway North blvd ste 300",
      "address_2": "N/A",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0475-2023",
      "product_description": "Leica MICROSYSTEMS Proveo 8, Part Number 10449058, surgical microscope system",
      "product_quantity": "43 units",
      "reason_for_recall": "There is a potential issue resulting in the Power Supply potentially becoming overheated due to internal components overstressing. Consequently, due to the overstressing of the internal components, one specific component can fail which will cause the device to unexpectedly shutdown.",
      "recall_initiation_date": "20221121",
      "center_classification_date": "20221209",
      "report_date": "20221221",
      "code_info": "Serial Numbers:  20522001, 20722001, 21221001, 40222001, 40722001, 70722001, 80522002, 110722001, 150422001, 180322001, 211021001, 260722001, 290422001, 300122001, 270522001.    ***EXPANSION 09/08/2023, Serial Numbers: 271222001, 100423002, 31022001, 180623001, 71222001, 210523002, 270922002, 120623002, 100123001, 130623001, 290922002, 300922002, 210223001, 220223001, 251122001, 281122001, 20623002, 60922001, 200822001, 40523001, 40523002, 270922001, 61122002, 20423001, 300523002, 190822002, 300922001, 280922001."
    }
  ]
}