  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63619",
      "recalling_firm": "Philips Healthcare Inc.",
      "address_1": "3000 Minuteman Road",
      "address_2": "N/A",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, India, Indonesia, Iraq, Italy, Japan, Latvia, Lithuania, Libya, Malta, Mexico, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE and UK.",
      "recall_number": "Z-0475-2013",
      "product_description": "Philips Digital Diagnost Software Version R2.0.2    Product Usage:  This system is used for making X-ray exposures for diagnostics",
      "product_quantity": "623 units",
      "reason_for_recall": "With Digital Diagnost R2.0.2 a mirrored \"R in a circle\" is burnt into the image (if mirrored) in the lower right corner of the image.  When such an image is mirrored back inside the PACS system, an unmirrored \"R in a circle\" appears in the lower left corner.  This can be mistaken for a \"Right patient side\" marker, although this can appear on the left patient side.",
      "recall_initiation_date": "20121030",
      "center_classification_date": "20121204",
      "termination_date": "20170117",
      "report_date": "20121212",
      "code_info": "All Digital Diagnost systems with release 2.0.2 and 2.0.2 SP1  Models 712020, 712022"
    }
  ]
}