      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63612",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) and countries of: Japan and Netherlands.",
      "recall_number": "Z-0472-2013",
      "product_description": "OSS Reamer Sleeve REF132-472662   KNEE REAMER SLEEVE 13.0 thru 20.0  STAINLESS STEEL    The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.",
      "product_quantity": "36",
      "reason_for_recall": "Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve.",
      "recall_initiation_date": "20121101",
      "center_classification_date": "20121204",
      "termination_date": "20130429",
      "report_date": "20121212",
      "code_info": "Part Numbers 32-472662, 13.0 thru 20.0    Lot 154590, 156980, 519470, 519480, 542290, 628890, 755530"
    }
  ]
}