  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89174",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the state of WA and the countries of  France, Germany, Japan, Poland, United Kingdom.",
      "recall_number": "Z-0471-2022",
      "product_description": "Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2 emergency stop recovery issue  Model Numbers: 722221, 722228, 722225, 722068, 722226, 722078, 722224",
      "product_quantity": "10 units (1 US and 9 OUS)",
      "reason_for_recall": "(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available",
      "recall_initiation_date": "20211112",
      "center_classification_date": "20220110",
      "termination_date": "20240205",
      "report_date": "20220119",
      "code_info": "US: Model 722068 Azurion 7 B20 S/N: 109 UDI: (01)00884838085367(21)109.  OUS: Model:                                     S/N        UDI: 722226 Azurion 7 B20 176 (01)00884838099272(21)176; 722228 Azurion 5 M20 3 (01)00884838099234(21)3; 722221 Azurion 3 M12 8 (01)00884838099203(21)8; 722225 Azurion 7 B12 16 (01)00884838099265(21)16; 722225 Azurion 7 B12 73 (01)00884838099265(21)73; 722226 Azurion 7 B20 42   (01)00884838099272(21)42; 722226 Azurion 7 B20 157  (01)00884838099272(21)157; 722224 Azurion 7 M20 293 (01)00884838099258(21)293; 722078 Azurion 7 M12 220 (01)00884838085251(21)220"
    }
  ]
}