{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63612",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "address_2": "N/A",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) and countries of: Japan and Netherlands.",
      "recall_number": "Z-0471-2013",
      "product_description": "OSS Reamer Sleeve REF132-472661 KNEE REAMER SLEEVE 8.0 THRU 12.5 STAINLESS STEEL    The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.",
      "product_quantity": "46",
      "reason_for_recall": "Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve.",
      "recall_initiation_date": "20121101",
      "center_classification_date": "20121204",
      "termination_date": "20130429",
      "report_date": "20121212",
      "code_info": "Part Numbers 32-472661 , 8.0 thru 12.5   Lot 013020, 050310, 056280, 154670, 418380, 691220, 755510, 860171, 866460"
    }
  ]
}