  "results": [
    {
      "status": "Terminated",
      "city": "Chicago",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69631",
      "recalling_firm": "Summit Industries Inc.",
      "address_1": "2901 W Lawrence Ave",
      "address_2": "N/A",
      "postal_code": "60625-3621",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution including the states of:  AZ, CO, IL, KY, LA, NE, NY, OH, PA, TN, TX, VA, WA and WV.",
      "recall_number": "Z-0468-2015",
      "product_description": "The J700 Floor Mounted Tube Stand (FMTS) is intended to support and position the diagnostic x-ray tube housing assembly for a medical radiographic procedure. The product consists of a vertical column (with counterbalanced vertical carriage, counterweight and pulley system), horizontal floor car and a floor mounted track to ride on.",
      "product_quantity": "32",
      "reason_for_recall": "The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury.",
      "recall_initiation_date": "20141028",
      "center_classification_date": "20141126",
      "termination_date": "20150521",
      "report_date": "20141203",
      "code_info": "Model Numbers: J700-01, J700-02, J700-03, J700-04",
      "more_code_info": ""
    }
  ]
}