      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72861",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide including :  DC, PR, Virgin Islands and Bermuda",
      "recall_number": "Z-0467-2016",
      "product_description": "MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).",
      "product_quantity": "in US 3752 units",
      "reason_for_recall": "Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into the magnet room or other areas. This could lead to displacement of oxygen, cause cold burns, and/or pose a risk to users, patient or others.",
      "recall_initiation_date": "20151207",
      "center_classification_date": "20151222",
      "termination_date": "20170621",
      "report_date": "20151230",
      "code_info": "all units distributed since 1991"
    }
  ]
}