{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Louvain-la-neuve",
      "state": "N/A",
      "country": "Belgium",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89267",
      "recalling_firm": "Ion Beam Applications S.A.",
      "address_1": "Chemin Du Cyclotron 3",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "IN USA: VA, OK, NJ, TN  OUS: Czech Republic.",
      "recall_number": "Z-0466-2022",
      "product_description": "Proteus 235; Version: PTS-8 versions before PTS-8.7.2",
      "product_quantity": "5 devices",
      "reason_for_recall": "Proton Therapy System (PTS) software can be  started in the clinical environment (clinical user) while some processes are still running in the test environment (tcs user). Initiating the startup of the clinical processes in the environment is not prevented even if some processes of the test environment (tcs) are still running. It may lead to situations where the system is using  test  processes without notifying the clinical  user. If the versions of the processes are different between both environments, this could  have an impact on patient treatments.",
      "recall_initiation_date": "20211129",
      "center_classification_date": "20220107",
      "termination_date": "20240220",
      "report_date": "20220119",
      "code_info": "PAT.108 (US), PAT.110 (US), PAT.113 (US), PAT.114 (EU), SAT.119 (US)."
    }
  ]
}