  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Wayne",
      "address_1": "45 Barbour Pond Dr",
      "reason_for_recall": "Maquet has received an increase in the number of complaints for \"self-activated/remains activated/overheats\" for the Hemopro VH-30000, \"Hemopro 1\"",
      "address_2": "",
      "product_quantity": "12,259 units",
      "code_info": "Catalog number  VH-3000 (US)  Lot numbers   25081686, 25081595, 25081460, 25081204, 25080939, 25080744, 25080418, 25080351, 25080264, 25079982, 25079913, 25073724, 25079449, 25079363, 25079086, 25078931, 25078713, 25078463, 25078291, 25078104, 25077927,25077643, 25077563, 25077123, 25076989, 25076372, 25076371, 25076133,     Catalog  number   VH-3000W (Foreign)  Lot numbers  25080938, 25080661, 25080311, 25078777,  25078289, 25078103 and 25077051",
      "center_classification_date": "20131206",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of Brazil, Czech Republic, Denmark, France, Hong Kong, Italy, Middle East, Netherlands, Singapore, Switzerland, and the United Kingdom.",
      "state": "NJ",
      "product_description": "Vasoview Hemopro Endoscopic Vessel Harvesting System Maquet Cardiovascular LLC 45 Barbour Pond Drive Wayne, NJ 07470     Indicated for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels oft he extremities, dissection of ducts and other structures int he extraperitoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection/vessel harvesting along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or the radial artery for use in coronary artery bypass grafting. Thorascopic procedures include exposure and dissection of structures external to the parietal pleura including nerves blood vessels and other tissues of the chest wall.",
      "report_date": "20131218",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Maquet Cardiovascular, LLC",
      "recall_number": "Z-0466-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66948",
      "termination_date": "20150424",
      "more_code_info": "",
      "recall_initiation_date": "20131030",
      "postal_code": "07470-2094",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}