{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Shenyang, Liaoning",
      "state": "N/A",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63553",
      "recalling_firm": "Philips And Neusoft Medical Systems Co., Ltd.",
      "address_1": "16 Century Road, Neusoft Park,",
      "address_2": "Hun Nan Industrial Area",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.",
      "recall_number": "Z-0466-2013",
      "product_description": "NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501.  NeuViz Dual series CT Scanner System, part number 989605651321    The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.",
      "product_quantity": "20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series",
      "reason_for_recall": "The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach.  This issue may affect NeuViz Dual and NeuViz 16 scanners.",
      "recall_initiation_date": "20121018",
      "center_classification_date": "20121204",
      "termination_date": "20130708",
      "report_date": "20121212",
      "code_info": "NeuViz 16 system Serial numbers:    N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024.     Neuviz Dual Serial numbers:    400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013."
    }
  ]
}